    Intended Use

    The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    The GC70 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    The GU60A & GU60A-65 Digital X-ray Imaging Systems are intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    The GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    The GF50A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.

    The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

    Device Description

    GC70, GU60&GU60A-65, GF50, GF50A, GR40CW, GM85 and GC85A are used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-station, SAMSUNG digital X-ray operation software, and sent to the Picture Archiving & Communication System (PACS) sever for reading images.

    The IPE operates, from the input image, the roles of a region-of-interest extraction, tonescale mapping, noise reduction and texture restoration. The IPE employing an advanced noise reduction algorithm (hereinafter "new IPE") is shown that the image quality of PA radiograph for average adult chest, exposed at the condition of 50% lower dose at Entrance Skin Exposure (ESE) in comparison with the condition of the conventional noise reduction algorithm (hereinafter "old IPE"), is substantially equivalent.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study proving the device meets those criteria, specifically concerning dose reduction capabilities of the Image Post-processing Engine (IPE) with an advanced noise reduction algorithm in Samsung Digital X-ray Systems (GC70, GU60A, GU60A-65, GF50, GF50A, GR40CW, GM85, and GC85A).

    Here is the requested information:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criterion is the ability of the new IPE to reduce X-ray dose while maintaining image quality comparable to the old IPE for diagnostic confidence. The specific dose reduction percentages are the performance metrics.

    Acceptance CriterionReported Device Performance
    Dose Reduction for Adult Abdominal RadiographsUp to 47.5% dose reduction for abdominal radiographs of adult, compared to the old IPE while achieving similar image quality.
    Dose Reduction for Pediatric AbdomenUp to 45% dose reduction for pediatric abdomen, compared to the old IPE while achieving similar image quality.
    Dose Reduction for Pediatric Chest15.5% dose reduction for pediatric chest, compared to the old IPE while achieving similar image quality.
    Dose Reduction for Pediatric SkullUp to 27% dose reduction for pediatric skull, compared to the old IPE while achieving similar image quality.

    2. Sample Size Used for the Test Set and Data Provenance

    • Adult Abdominal Radiograph Test Set:

      • Anatomical phantom images: Number of images not specified, but taken at "various exposure condition." The study states, "the new IPE with an advanced noise reduction algorithm retained the quality of images captured at 47.5% reduced exposure in comparison with the old IPE."
      • Clinical images: Number of images not specified, but used to "confirm that it was possible to reduce the dose in clinical images as well."
      • Provenance: Not explicitly stated, but the submission is from Samsung Electronics Co., LTD. Republic of Korea. The clinical testing was conducted at "one medical site."
      • Retrospective or Prospective: Not specified.
    • Pediatric Population Test Set (Chest, Abdomen, Skull):

      • Number of images: "Series of dose-simulated images" for each body part.
      • Number of patients: Not specified explicitly, but mentioned as "each patient."
      • Provenance: Not explicitly stated, but the submission is from Samsung Electronics Co., LTD. Republic of Korea. The clinical testing was conducted at "one medical site."
      • Retrospective or Prospective: Not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Adult Abdominal Radiograph Test Set:

      • Anatomical phantom images were reviewed by three professional radiologists.
      • Clinical images were reviewed by two professional radiologists.
      • Qualifications: "Professional radiologists" (no further details on experience given).
    • Pediatric Population Test Set:

      • Three experienced pediatric radiologists.
      • Qualifications: "Experienced pediatric radiologists" (no further details on experience given).

    4. Adjudication Method for the Test Set

    The adjudication method is not explicitly detailed. However, for both adult and pediatric studies, images were "scored by the 5-point grading scale" for assessment of image quality. This implies individual scoring, and for the pediatric study, "Three experienced pediatric radiologists assessed the series of dose-simulated images to decide the optimal dose for each patient." The decision for the "optimal dose" for pediatric cases suggests a consensus or agreement among these experts, but the exact method (e.g., majority vote, discussion to reach consensus) is not specified.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was done to evaluate human readers' improvement with AI vs. without AI assistance. The study focused on the device's standalone performance in enabling dose reduction while maintaining image quality as assessed by human readers. The new IPE is a component within the imaging system, not an AI assistance tool for human readers.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies evaluated the standalone performance of the new IPE algorithm in terms of enabling dose reduction while maintaining image quality. The performance was assessed by comparing images processed by the new IPE at reduced doses against images from the old IPE or a reference, with human experts providing the assessment of image quality and diagnostic appropriateness.

    7. The Type of Ground Truth Used

    The ground truth for both adult and pediatric studies was expert consensus/assessment of image quality and diagnostic appropriateness.

    • For adult abdominal radiographs: Expert radiologists scored images based on a 5-point grading scale, considering anatomical regions, physical parameters, sharpness, and visualization.
    • For pediatric populations: Experienced pediatric radiologists assessed dose-simulated images to determine the "optimal dose" at which image quality remained appropriate for diagnosis.

    Additionally, phantom studies (TOR CDR radiography phantom and anthropomorphic phantom) were used to quantitatively assess image quality metrics like Contrast to Noise Ratio (CNR), Detail Compacted Contrast (DCC), and Modulation Transfer Functions (MTF).

    8. The Sample Size for the Training Set

    The document does not provide information about the training set size for the Image Post-processing Engine (IPE) algorithm. It focuses on the validation of the algorithm's dose reduction capabilities.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established for the IPE algorithm.