510(k) Data Aggregation

    K Number
    K182177

    Validate with FDA (Live)

    Device Name
    Accipiolx
    Manufacturer
    Date Cleared
    2018-10-26

    (77 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Accipiolx is a software workflow tool designed to aid in prioritizing the clinical assessment of adult non-contrast head CT cases with features suggestive of acute intracranial hemorrhage in the acute care environment. Accipiolx analyzes cases using an artificial intelligence algorithm to identify suspected findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage.

    Accipiolx is not intended to direct attention to specific portions of an image or to anomalies other than acute intracranial hemorrhage. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out hemorrhage or otherwise preclude clinical assessment of CT cases.

    Device Description

    Accipiolx is a software device designed to be installed within healthcare facility radiology networks to identify and prioritize non-contrast head CT (NCCT) scans based on algorithmically-identified findings of acute intracranial hemorrhage (alCH). The device, developed using computer vision and deep learning technologies, facilitates prioritization of CT scans containing findings of alCH. There are two main components of the software device: (1) the Accipiolx Agent and (2) the MaxQ-Al Engine. The Agent serves as an active conduit which receives head CT studies from a PACS and transfers them to the Engine. After successful processing of a case via the MaxQ-Al Engine, the Accipiolx Agent receives the Engine results and returns them to the PACS or workstation for use in worklist prioritization.

    Accipiolx works in parallel to and in conjunction with the standard care of workflow. After a CT scan has been performed, a copy of the study is automatically retrieved and processed by Accipiolx. The device performs identification and classification of objects consistent with alCH, and provides a case-level indicator which facilitates prioritization of cases with potential acute hemorrhagic findings for urgent review.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study as described in the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Predefined Performance Goals)Reported Device Performance
    SensitivityLess than 92% (inferred from "exceeded" statement)92% (95% Cl: 87.29-95.68%)
    SpecificityLess than 86% (inferred from "exceeded" statement)86% (95% Cl: 80.18-90.81%)

    Notes: The document states that the reported results "exceeded the predefined performance goals." This implies the acceptance criteria were less than the achieved performance, meaning the device had to perform at least
    as well as a certain threshold. However, the exact numerical thresholds for the acceptance criteria are not explicitly stated, so I've inferred them based on the "exceeded" statement.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 360 cases
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but collected from "over 30 US sites." This suggests the data originated from the United States.
      • Retrospective/Prospective: Retrospective study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: At least two expert neuroradiologist readers.
    • Qualifications of Experts: Expert neuroradiologist readers. No specific experience in years is provided.

    4. Adjudication method for the test set

    • Adjudication Method: Concurrence of at least two expert neuroradiologist readers. This implies a 2-reader consensus model.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • A MRMC comparative effectiveness study involving human readers with vs. without AI assistance was not done. The study described focused on the standalone performance of the AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone study was done. The performance testing specifically evaluated the "device sensitivity and specificity...compared to ground truth." This describes the algorithm's performance in isolation.

    7. The type of ground truth used

    • Type of Ground Truth: Expert consensus (established by concurrence of at least two expert neuroradiologist readers).

    8. The sample size for the training set

    • Sample Size for Training Set: Not explicitly stated. The text mentions "Accipiolx was developed using a training CT cases collected from multiple institutions and CT manufacturers," but it does not provide a specific number for the training set size.

    9. How the ground truth for the training set was established

    • How Ground Truth for Training Set was Established: Not explicitly stated how the ground truth for the training specific was established. The document mentions "optimization of object and feature identification, algorithmic training and selection/optimization of thresholds," which implies a process was followed, but the method of establishing training ground truth is not detailed.