    Intended Use

    Portable Electro-Stimulation Therapy Device, model LGT-235 is indicated for the Symptomatic relief of chronic intractable pain, Post-traumatic pain and Post-surgical pain of the arms and legs.

    Device Description

    The Portable Electro-Stimulation Therapy Device, model LGT-235 is a stimulator which sends gentle electrical current to underlying nerves via electrodes (Knee brace or self-adhesive electrode) applied on the skin with two operational modes and powered by rechargeable lithium battery; the device system is made up of mobile app MStim Arth, main unit and electrodes (Knee brace and self-adhesive electrode).

    The Mobile App MStim Arth provides access to treatment controls for Portable Electro-Stimulation Therapy Device from a compatible mobile device for selecting a pre-programmed output mode and treatment time.

    The main unit provides access to adjust the intensity up or down, put ON/OFF the main unit.

    The two modes that Portable Electro-Stimulation Therapy Device, model LGT-235 employs are transcutaneous electrical stimulation (TENS) and Microcurrent (MCR).

    TENS and MCR are both specifically targets the sensory nerves, which are responsible for sending pain signals to the brain. TENS and MCR use tiny electrical impulses sent through the skin to nerves to modify the pain perception. TENS and MCR do not cure any physiological problem; it only helps control the pain, this activates the underlying sensory nerves. Knee brace is placed on the leg or arm close to the area of pain.

    AI/ML Overview

    The provided text describes the Portable Electro-Stimulation Therapy Device, model LGT-235, and its substantial equivalence to a predicate device (Sonicator Plus 920, Model ME 920), which implies that it meets certain performance criteria. However, the document does not explicitly list quantitative acceptance criteria for the device's performance in terms of clinical efficacy or diagnostic accuracy. Instead, it focuses on demonstrating equivalence through technical specifications and compliance with recognized safety standards.

    Therefore, many of the requested fields cannot be filled as they relate to clinical studies or performance metrics that are not detailed in this 510(k) summary.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of explicit acceptance criteria for clinical performance (e.g., pain reduction percentage, accuracy). Instead, it compares the technical characteristics and output specifications of the subject device to the predicate device, with the implicit acceptance criterion being "similar" or "same," meaning the differences do not raise new questions of safety or effectiveness.

    CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Subject Device)
    ClassificationSame as Predicate21 CFR 882.5890 (Same)
    Product CodeSame as PredicateGZJ (Same)
    FDA ClassSame as Predicate2 (Same)
    Intended UseSame as PredicateSymptomatic relief of chronic intractable pain, Post-traumatic pain, Post-surgical pain (Same)
    Principle of operationSame as PredicateSends gentle electrical current to muscle group via electrodes applied on the skin (Same)
    Target PopulationSame as PredicateAdults (Same)
    Anatomical SiteSame as PredicateArm and legs (Same)
    WaveformSame as PredicateSymmetrical biphasic (Same)
    Maximum Output VoltageSimilar to Predicate, acceptable differencesTENS: 50V @ 500 Ω, 115 V @ 2 kΩ, 125 V @ 10 kΩ; MCR: 0.45 V@ 500 Ω, 1.8 V@ 2 kΩ, 7.4 V @ 10 kΩ (Similar, acceptable error in MCR)
    Maximum Output CurrentSimilar to Predicate, acceptable differencesTENS: 100 mA@ 500 Ω, 58 mA@ 2 kΩ, 13 mA@ 10 kΩ; MCR: 0.92 mA@ 500 Ω, 0.90 mA@ 2 kΩ, 0.74 mA@ 10 kΩ (Similar, acceptable error in MCR)
    Pulse Width (TENS)Similar to Predicate100 ~ 600 µs (Similar, within safe/effective range of predicate)
    Pulse Width (MCR)Similar to Predicate1.25 ms ~ 1.67 s (Similar, within safe/effective range of predicate)
    Frequency (TENS)Similar to Predicate0.5 ~ 250 Hz (Similar, predicate range overriding proposed device suggests safety)
    Frequency (MCR)Similar to Predicate0.3 ~ 400 Hz (Similar, predicate range overriding proposed device suggests safety)
    Maximum IntensitySame as Predicate100 mA (Same)
    Net Charge (µC/pulse)Same as Predicate0 µC @ 500Ω (Same)
    Maximum Phase ChargeSimilar to PredicateTENS: 60 (µC) @ 500Ω; MCR: 75 (µC) @ 500Ω (Similar, predicate range overriding proposed device suggests safety)
    Maximum Current DensitySimilar to Predicate, acceptable differencesTENS: 1.97 (mA/cm²) @ 500Ω; MCR: 0.026 mA/cm² @ 500Ω (Similar, TENS proposed device is overriding predicate; for MCR, slightly larger but proven safe by IEC 60601-2-10)
    Maximum Power DensitySimilar to Predicate, acceptable differencesTENS: 8.26 mW/cm² @ 500Ω; MCR: 0.016 W/cm² @ 500Ω (Similar, TENS proposed device is overriding predicate; MCR within acceptable limits per guidance)
    BiocompatibilityPass ISO 10993-5 and 10Knee brace demonstrated safety by passing ISO 10993-5 and -10. (Meets)
    Electrical SafetyPass ANSI AAMI ES60601-1Demonstrated electrical safety by passing ANSI AAMI ES60601-1 and meeting relevant chapters. (Meets)
    Electromagnetic Compatibility (EMC)Pass IEC 60601-1-2Demonstrated EMC by passing IEC 60601-1-2. (Meets)
    Safety of Nerve & Muscle StimulatorsPass IEC 60601-2-10Demonstrated safety by passing IEC 60601-2-10. (Meets)
    Home Healthcare Environment SafetyPass IEC 60601-1-11Demonstrated safety by passing IEC 60601-1-11. (Meets)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission focuses on non-clinical (engineering, safety, and performance) testing, and does not describe a clinical test set or its provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. The document describes non-clinical testing and equivalency to a predicate device, not a study involving expert-established ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. There is no described clinical test set with ground truth requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/not provided. The device is an electro-stimulation therapy device for pain relief, not an AI-assisted diagnostic or interpretative device that would typically involve human "readers" or MRMC studies for AI improvement.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable/not provided. The device is a therapy device, not an algorithm, and its performance is assessed against electrical and safety standards, not standalone diagnostic accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable/not provided. The "ground truth" for the non-clinical tests involved comparing the device's measured electrical outputs and safety characteristics against established international standards (ISO, ANSI, IEC) and the specifications of the predicate device.

    8. The sample size for the training set

    This information is not applicable/not provided. As this is a 510(k) for a physical medical device (electro-stimulation therapy), not an AI/ML algorithm, there is no "training set" in the conventional sense.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. See response to point 8.