Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181787

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-04-25

    (294 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cediflo Enteral Feeding Tubes are intended for the administration of enteral nutrition, fluids, and/or medications by the nasoenteric route into the stomach or small intestine. Indicated for patients 2 years and above which require nutritional support, are not able to meet their nutritional requirements by oral intake and have functioning and accessible gastrointestinal tract. Maximum duration of use: 42 days.

    The Cediflo Junior Enteral Feeding Tubes are intended for the administration of enteral nutrition, fluids, and/or medications by the nasoenteric route into the stomach or small intestine. Indicated for neonatal and pediatric patients which require nutritional support, are not able to meet their nutritional requirements by oral intake and have functioning and accessible gastrointestinal tract.

    Device Description

    The Cediflo and Cediflo Junior Enteral Feeding Tubes are sterile, single use devices intended for use in acute care facilities, long-term care facilities, and home. They are made from radiopaque polyurethane tubing printed with centimeter markings and bonded at the proximal end to a rigid male ENFit connector with a tethered cap. The proposed device is offered in lengths of 50 cm to 120 cm and diameters of 5 Fr (1.66 mm) to 16 Fr (5.33 mm). These tubes are supplied with or without a guidewire. The distal tip of the polyurethane tubing is closed and rounded with four side holes to allow for fluid flow.

    The male ENFit connector of the proposed device is designed to comply with ISO 80369-3:2016 and allows for connection to enteral feeding specific sets and syringes while reducing the likelihood of misconnections to devices that are not intended for enteral administration. A twist on cap is tethered to the male ENFit connect via a strap.

    The guidewire, when provided, of the proposed device is a stainless-steel braided wire with a rounded distal tip which is pre-loaded into the polyurethane feeding tube. The guidewire stiffens the polyurethane tubing to aide in insertion of the device. The guidewire has a lubricious coating that aids in removal of the guidewire from the feeding tube in situ. A flow through guidewire connector is attached to the proximal end of the guidewire which provides the connection between the wire and feeding tube.

    AI/ML Overview

    The provided document (K181787) is a 510(k) premarket notification for Cediflo Enteral Feeding Tubes & Cediflo Junior Enteral Feeding Tubes. This type of medical device is a physical tube used for nutritional support, not a software-based AI/ML device. Therefore, the questions related to AI/ML acceptance criteria, ground truth establishment, expert adjudication, MRMC studies, and training/test set data are not applicable to this document.

    The document primarily relies on non-clinical performance testing to demonstrate substantial equivalence to predicate devices, as well as biocompatibility testing. No clinical performance testing, animal testing, or AI/ML specific studies were conducted or required for this device.

    Here's a breakdown of the relevant information provided in the document regarding how the device meets acceptance criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't present a formal table of "acceptance criteria" in the way one would for an AI/ML device with specific metrics like sensitivity or specificity. Instead, it demonstrates compliance with relevant international standards and performs specific mechanical and material tests.

    Criteria/TestAcceptance Method (Implicit)Reported Device Performance (Implicit - Pass/Fail)
    BiocompatibilityISO 10993-1Pass (Successfully evaluated per ISO 10993-1)
    ISO 80369-3:2016 (Small-bore connectors)Compliance with standard, including tests for:Pass (Designed and constructed in compliance)
    - Fluid leakageMeet standard requirementsMet
    - Stress crackingMeet standard requirementsMet
    - Resistance to separation from axial loadMeet standard requirementsMet
    - Resistance to separation from unscrewingMeet standard requirementsMet
    - Resistance to overridingMeet standard requirementsMet
    - Disconnection by unscrewingMeet standard requirementsMet
    - Dimensional verificationMeet standard requirementsMet
    EN 1615:2000 (Feeding Tubes)Compliance with standard, including tests for:Pass (Designed and constructed in compliance)
    - LeakageMeet standard requirementsMet
    - Tensile StrengthMeet standard requirementsMet
    - Flow Rate (with water and commercially available enteral nutrition)Meet standard requirementsMet
    ISO 11070 (Guidewire - if included)Compliance with standard, including tests for:Pass (Testing performed)
    - Tensile StrengthMeet standard requirementsMet
    - Fracture ResistanceMeet standard requirementsMet
    - Flexion ResistanceMeet standard requirementsMet
    Tubing Kink ResistanceEN 13868:2002Pass (Testing performed)
    Shelf LifeAccelerated aging conditioningSubstantiated
    Substantial EquivalenceComparison to predicate devices based on intended use, indications, clinical settings, target population, design characteristics, materials, and performance.Determined as "Substantially Equivalent"

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated as "sample size" in the context of a "test set" for an algorithm. Instead, the tests refer to a sufficient number of manufactured units or materials to validate the design and manufacturing claims (e.g., "accelerated aging conditioning" implies a batch of devices).
    • Data Provenance: The testing was conducted by or for Cedic s.r.l., located in Peschiera Borromeo (MI), Italy. This is a manufacturer's internal testing or contracted testing to demonstrate compliance with standards. It is retrospective in the sense that it's done on manufactured products for regulatory submission, but not in the clinical data sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not applicable. This is a physical device, not an AI/ML algorithm that requires expert-established ground truth from medical images/data. The "ground truth" for this device's performance is objective measurement against engineering standards and material properties.

    4. Adjudication Method for the Test Set:

    • Not applicable. No human adjudication of clinical data is involved. Performance is determined by whether the device meets quantitative and qualitative engineering and material standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is not a study type relevant to an enteral feeding tube.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) study was done:

    • Not applicable. This is not an algorithm. The "standalone" performance here refers to the device's ability to meet its specifications independently, which is what the non-clinical tests demonstrate (e.g., flow rate, tensile strength).

    7. The Type of Ground Truth Used:

    • The "ground truth" for this device's performance is based on established engineering standards (ISO, EN), material specifications, and physical measurements (e.g., flow rate, tensile strength, leakage). It's objective, quantitative data derived from laboratory testing, not medical consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set:

    • Not applicable. There is no "training set" for a traditional medical device like an enteral feeding tube.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. No training set exists.

    In summary, this 510(k) clearance relied on demonstrating that the Cediflo Enteral Feeding Tubes meet established performance standards and are substantially equivalent to existing predicate devices through comprehensive non-clinical (laboratory and engineering) testing, rather than clinical trials or AI/ML performance studies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1