    Intended Use

    The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles:

    • · Dynamic Air Removal 270°F (132°C) 4 minutes
    • · Dynamic Air Removal 275°F (135°C) 3 minutes
    • · Gravity 250°F (121°C) 30 minutes
    • · Gravity 270°F (132°C) 15 minutes.
      When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes.
    Device Description

    The product is intended to monitor the critical parameters of steam sterilization cycles described in the indications for use by producing an optical change (signal) that is detected by the STERIS proprietary reader, Celerity Steam Incubator in 20 minutes to confirm the viability of the biological indicator at the end of a steam sterilization process. The product consists of a biological organism known to be resistant to steam (Geobacillus stearothermophilus) and a defined nutrient media. A reporter enzyme, which is produced by the native organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    AI/ML Overview

    Here's an analysis of the provided information to answer your questions about acceptance criteria and the study that proves the device meets them:

    Device: Celerity 20 Steam Biological Indicator

    1. A table of acceptance criteria and the reported device performance

    Based on the document, only one performance test is explicitly mentioned with an acceptance criterion and conclusion:

    TestAcceptance CriteriaReported Device Performance/Conclusion
    Simulated UseDemonstrate no growth when exposed to a full cycleFull cycle - no growth

    The document also lists several technical characteristics for the Celerity 20 Steam Biological Indicator (modified) and compares them to the predicate device. While these aren't explicitly stated as "acceptance criteria" in the same way as the "Simulated Use" test, they represent performance parameters that the device is expected to meet and does meet, often being "Identical" to the predicate.

    For completeness, here's a combined table including those technical characteristics as implicit acceptance criteria if one assumes they must match the predicate:

    FeatureAcceptance Criteria (typically matching predicate)Reported Device Performance
    Simulated UseDemonstrate no growth when exposed to a full cycleFull cycle - no growth
    Indicator organism> 90% similarity to ATCC 7953 Geobacillus stearothermophilus> 90% similarity to ATCC 7953 Geobacillus stearothermophilus (Identical to predicate)
    Mechanism of actionAn enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moietyAn enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety (Identical to predicate)
    AccessoriesAutomated incubator / readerAutomated incubator / reader (Identical to predicate)
    Viable spore population1.0 - 4.0 x 10^6 spore/SCBI1.0 - 4.0 x 10^6 spore/SCBI (Identical to predicate)
    ResistanceD121 ≥ 1.5 min D132 ≥ 10 s D135 ≥ 8 sD121 ≥ 1.5 min D132 ≥ 10 s D135 ≥ 8 s (Identical to predicate)
    Survival TimeMeets the longer of FDA and ISO 11138-3 requirementsMeets the longer of FDA and ISO 11138-3 requirements (Identical to predicate)
    Kill TimeMeets the shorter of FDA and ISO 11138-3 requirementsMeets the shorter of FDA and ISO 11138-3 requirements (Identical to predicate)
    Hold timeUp to 72 hUp to 72 h (Identical to predicate)
    Carrier growth inhibition / media growthPositive growth of less than 100 spores after primary packaging and media are subject to worst case steam exposurePositive growth of less than 100 spores after primary packaging and media are subject to worst case steam exposure (Identical to predicate)
    Culture Conditions55- 59 °C, media included in SCBI, 20-minute incubation time.55- 59 °C, media included in SCBI, 20-minute incubation time. (Identical to predicate)
    Primary PackagingDirect inoculum on plastic vial, cap with recovery media.Direct inoculum on plastic vial, cap with recovery media. (Identical to predicate, though the cap label changed, this refers to the fundamental structure)
    Shelf Life10 months (predicate)10 months (Identical to predicate, but the note states "Shelf-life testing is ongoing", implying this is the target or current demonstrated shelf life, not necessarily a completed test definitively proving the full 10 months for the modified device if testing is still ongoing at the time of submission)
    Cap labelSingle-ply label on cap edge, Process indicator on labelSingle-ply label on cap edge, Process indicator on label (Differs from predicate, which had a two-ply label on cap top. This is a design change, not a performance criterion in the same way, but noted for comparison.)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size used for the "Simulated Use" test or any other specific test. It also does not provide information about the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This type of information is not applicable to a biological indicator device. The "ground truth" for a biological indicator is determined by the presence or absence of microbial growth, which is an objective biological outcome rather than an expert interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, the determination of microbial growth is objective, not requiring human adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a biological indicator, not an AI-assisted diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a biological indicator, not an algorithm. Its performance is inherent to its biological and chemical properties. The Celerity™ Steam Incubator provides an automated fluorescent result, which could be considered a "standalone" reading, but it's not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for relevant tests (like viability, resistance, survival, kill time, and simulated use) is based on the presence or absence of microbial growth (Geobacillus stearothermophilus), which is an objective biological outcome. This is determined through standard microbiological techniques.

    8. The sample size for the training set

    Not applicable. This device is a biological indicator, not a machine learning model. Therefore, there is no "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established

    Not applicable. No training set is involved for this type of device.