    Device Name
    VITEK MS
    Manufacturer
    Date Cleared
    2018-12-21

    (205 days)

    Product Code
    Regulation Number
    866.3378
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VITEK® MS is a mass spectrometry system using matrix-assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) for the identification of microorganisms cultured from human specimens.

    The VITEK® MS system is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial, yeast and mould infections.

    Device Description

    The VITEK® MS v3.0 system is a system consisting of kit reagents (VITEK® MS-CHCA, VITEK® MS-FA, VITEK® MS Mycobacterium/Nocardia Kit, VITEK® MS Mould Kit), VITEK® MS-DS target slides, and the VITEK® MS (original equipment VITEK® MS Prep Station, Knowledge Base v3.2.0, software, and the VITEK" MS (original eq

    AI/ML Overview

    This document focuses on the VITEK® MS system, a mass spectrometry system for microorganism identification, and its performance characteristics. It outlines the acceptance criteria and details of the study conducted to prove the device meets these criteria.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the reported performance targets which show very high agreement rates for identification. The primary performance metric is the "Total Correct Genus ID (One Choice and Low Discrimination)".

    Here's a table summarizing criteria derived from the reported performance, reflecting typical expectations for such devices, and the observed performance for different organism groups.

    Organism GroupAcceptance Criteria (Implied) - Total Correct Genus IDReported Device Performance (Total Correct Genus ID)
    Gram-positive bacteria>95%95.8% (3594/3750)
    Gram-negative bacteria>95%95.2% (5788/6079)
    Gram-negative bacteria (Brucella only)>90%91.7% (220/240)
    Yeasts>95%96.8% (1316/1360)
    Moulds>90%92.7% (1398/1508)
    Mycobacterium>95%96.5% (777/805)
    Nocardia>95%97.9% (374/382)
    Overall (All Organisms Total)>95%95.4% (13247/13884)

    Note: The reported performance also details "One Choice Correct" and "Low Discrimination Correct" breakouts, as well as discordant and no-ID rates, providing a comprehensive view of the device's accuracy. For example, for "All" clinical bacteria and yeast isolates tested, the agreement with reference identification was 98.8% (4189/4241).

    2. Sample Size Used for the Test Set and Data Provenance

    Test Set Sample Size:

    • Overall Performance (Routine fresh and stock strains): 13,884 samples across various organism groups.
    • Clinical Studies (Clinical isolates): 4,241 test results.
    • Brucella specific testing (ATCC): 240 strains for clinical trial, 180 for reproducibility, 45 for challenge testing.

    Data Provenance:
    The data was collected from:

    • Routine fresh and stock strains from patient cultures in clinical microbiology laboratories and bioMérieux laboratories in the United States and France. This indicates a mix of prospective and retrospective (stock strains) data and geographic diversity.
    • Clinical isolates were tested at "one external clinical trial site and one internal site" for bacteria (other than Brucella) and yeasts.
    • For Brucella, clinical trial, reproducibility, and challenge testing were performed at "one external clinical trial site" (specifically, ATCC is mentioned for Brucella testing). This suggests a focus on specific, well-characterized strains for Brucella.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts used or their specific qualifications (e.g., "radiologist with 10 years of experience"). However, it mentions that "Samples were sequenced by the appropriate reference laboratory, and if needed, additional analysis was performed internally to obtain a reference identification (GenBank, dendrogram analysis)."

    This implies that the ground truth was established by:

    • Personnel at "appropriate reference laboratories" (implying qualified microbiologists/scientists).
    • Internal personnel at bioMérieux.
    • The use of "proficiency panels tested by all participating technologists at the external sites" for the clinical trial suggests that the individuals performing the tests were trained and qualified.

    While specific numbers and detailed qualifications of individual experts are not provided, the methodology points to reliance on established laboratory practices and reference techniques for accurate identification.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe a formal adjudication method (like 2+1 or 3+1 for human readers of images). For microbiological identification, the "adjudication" is inherent in the ground truth establishment process:

    • Reference methods (DNA sequencing analysis, supplemental testing) are considered the gold standard.
    • If sequencing was insufficient, "additional analysis was performed internally to obtain a reference identification (GenBank, dendrogram analysis)." This suggests a process where discrepancies or difficult identifications are resolved by deeper, more definitive, and comparative analysis rather than a simple consensus among multiple readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done in the traditional sense of comparing human readers with and without AI assistance for interpretation tasks. This device is an automated system for identifying microorganisms, not an AI for image interpretation that assists human diagnosticians. The study focuses on the accuracy of the automated system against a definitive reference method.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, the study primarily evaluates the standalone performance of the VITEK® MS system. The reported performance metrics (e.g., "Total Correct Genus ID") reflect the accuracy of the device's algorithm in identifying microorganisms when presented with samples.

    While the "Interpretation of results and use of the VITEK® MS system require a competent laboratorian" is mentioned in the device description, the performance tables quantify the output of the machine itself compared to the ground truth, effectively reporting its standalone accuracy.

    7. The Type of Ground Truth Used

    The ground truth for microorganism identification was established using:

    • DNA sequencing analysis: This is the primary reference method.
    • Supplemental testing (when necessary): This could include biochemical tests, serology, or other definitive microbiological techniques.
    • Previously well-characterized strains (ATCC or equivalent): For certain challenge or quality control tests, established strains were used, for which the identity is already known and highly reliable.
    • GenBank and dendrogram analysis: Used for additional internal analysis to obtain reference identification when sequencing was not fully conclusive, reinforcing the ground truth.

    This indicates a robust, multi-faceted approach to ground truth establishment, relying on genetic and phenotypic gold standards.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size used for the training set for the VITEK® MS Knowledge Base (KB) v3.2.0. It mentions that the "Knowledge Base includes data representing 1316 species and 1158 taxa displayed" and "1095 species of bacteria" and "232 species of fungi." This describes the content of the database used for identification, which is analogous to the "training data" that built the system's ability to recognize these species. However, the exact number of samples or spectra used to build this database is not provided in this excerpt.

    9. How the Ground Truth for the Training Set Was Established

    The document describes the "Knowledge Base" as having been "developed from spectra of a number of microbial species." While it doesn't detail the training process or the ground truth establishment for the training set specifically, it's highly implied that the same rigorous methods used for the test set (DNA sequencing, supplemental testing, well-characterized strains) would have been used to curate the reference spectra within the Knowledge Base. This ensures that the foundational data used by the device for identification is accurate and reliable. The continuous updates to the "Knowledge Base v3.2.0" suggest an ongoing process of incorporating new and validated spectral data.