Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181374

    Validate with FDA (Live)

    Device Name
    NRFit Syringe
    Date Cleared
    2018-12-14

    (204 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for local administration, and regional administration, specifically, wound infiltrations, peripheral nerve block, and epidural block. The device is not indicated for subarachnoid/spinal blocks.

    Device Description

    The proposed device is a disposable neuraxial syringe consist of three components: piston, plunger and barrel; and provided in a variety of volumes: 1ml, 3ml, 6ml, 10ml, 30ml, 35ml, 50ml, 30ml and 60 ml. The barrel of the syringe is printed with graduated markings indicating the volume of liquid inside the barrel.

    The proposed device is used to inject anesthesia medicine with the cooperation of the spinal anesthesia needles and epidural anesthesia needles which has an NRFit connectors in compliance with ISO 80369-6. The NRFit tip could help to reduce the residual risk of misconnections.

    The proposed device is provided in two statuses: sterile. The non-sterile NRFit syringe is intended to be sterilized prior to use to repackagers/medical device manufacturers, and the sterile NRFit syringe is supplied to the end user. The sterilized product is sterilized by Ethylene Oxide Gas to achieve a SAL of 10th and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

    AI/ML Overview

    The acceptance criteria and device performance for the NRFit Syringe are detailed in Section 8 of the provided 510(k) summary (pages 5 and 6 of the document). The study demonstrating the device meets these criteria is referred to as "Non-Clinical Test Conclusion."

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is reported as conformity to the specified standards and their respective clauses/annexes.

    Individual Test Defined in ISO 80369-6 (for connectors)Acceptance Criteria (Requirement Defined in ISO 80369-6)Reported Device Performance
    Fluid LeakageClause 6.1Conforms
    Stress CrackingClause 6.3Conforms
    Resistance to separation from axial loadClause 6.4Conforms
    Resistance to separation from unscrewingClause 6.5Conforms
    Resistance to overridingClause 6.6Conforms
    Individual Test Defined in ISO 7886-1:1993 (for syringes)Acceptance Criteria (Requirement Defined in ISO 7886-1:1993)Reported Device Performance
    CleanlinessClause 5Conforms
    Limits for acidity or alkalinityClause 6Conforms
    Limits for extractable metalsClause 7Conforms
    LubricantClause 8Conforms
    Tolerance on graduated capacityClause 9Conforms
    Graduated scaleClause 10Conforms
    BarrelClause 11Conforms
    Piston/plunger assemblyClause 12Conforms
    Dead spaceClause 14.1Conforms
    Freedom from air and liquid leakage past pistonClause 14.2Conforms

    Additionally, the device was tested against and found to comply with the following standards:

    • ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals
    • ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
    • ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
    • ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
    • ASTM F88/F88M-15 standard method for seal strength of flexible barrier materials
    • USP 39-NF34 <85> Bacterial Endotoxins Test
    • ISO 80369-20:2015 Small bore connectors for liquids and gases in healthcare applications- Part 20: Common test methods
    • USP <788> Particulate Matter in Injections (Method 1)

    2. Sample Size Used for the Test Set and Data Provenance
    The document does not explicitly state the sample size used for each individual test. It indicates that "Non-clinical tests were conducted to verify that the proposed device met all design specifications" and that the "test results demonstrated that the proposed device complies with the following standards." The provenance of the data (country of origin, retrospective or prospective) is not specified, but the submission is from Jiangsu Caina Medical Co., Ltd in China, suggesting the tests were likely conducted by or for them.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
    Not applicable. The ground truth for these non-clinical tests is established by the specifications of the referenced ISO and ASTM standards, not by expert consensus from medical professionals.

    4. Adjudication Method for the Test Set
    Not applicable. The tests are objective measurements against established standard requirements, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
    Not applicable. This is a medical device (syringe) and not an AI or diagnostic imaging device. Therefore, MRMC studies are not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
    Not applicable. This is a medical device (syringe) and not an AI algorithm.

    7. The Type of Ground Truth Used
    The ground truth used for the non-clinical tests is based on engineering and performance specifications defined within internationally recognized standards (ISO, ASTM, USP). This includes measurable physical properties and biological safety limits.

    8. The Sample Size for the Training Set
    Not applicable. There is no training set mentioned as this is not a machine learning or AI-based device.

    9. How the Ground Truth for the Training Set Was Established
    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1