    K Number
    K180864

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-07-20

    (109 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WATCHMAN™ TruSeal™ Access System is intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

    Device Description

    The Boston Scientific WATCHMAN™ TruSeal™ Access System is composed of an Access Sheath and Dilator. The TruSeal Access System is used to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

    AI/ML Overview

    The document is a 510(k) premarket notification for the WATCHMAN™ TruSeal™ Access System. It does not describe a study involving an AI/Machine Learning device or a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. Instead, it details the performance and biocompatibility testing of a medical catheter system. Therefore, I cannot provide information on acceptance criteria, reader studies, ground truth, or training set relevant to an AI/ML device based on this document.

    However, I can extract the acceptance criteria and study information related to the physical device's performance testing:

    1. Table of Acceptance Criteria and Reported Device Performance
    The document lists various performance tests conducted. While specific quantitative acceptance criteria values are not provided in this summary, the document states: "The results of the tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing." This implies that the device met pre-defined acceptance criteria for each test.

    Test CategorySpecific Performance TestReported Device Performance
    Device PerformanceDimensional CharacterizationConformed to requirements
    Distal and Proximal Marker LocationConformed to requirements
    Curve ShapeConformed to requirements
    TruSeal Access System Guidance / CrossingConformed to requirements
    Contrast Flow RateConformed to requirements
    Leak-Free ConduitsConformed to requirements
    Pressure RetentionConformed to requirements
    Corrosion ResistanceConformed to requirements
    Sheath Force TransmissionConformed to requirements
    Kink ResistanceConformed to requirements
    Tip DeflectionConformed to requirements
    TorqueabilityConformed to requirements
    Snap Fit Attachment ForceConformed to requirements
    Access Sheath Compression ForceConformed to requirements
    Dilator StiffnessConformed to requirements
    TensileConformed to requirements
    RadiopacityConformed to requirements
    Luer Lock TestingConformed to requirements
    ParticulatesConformed to requirements
    SurfaceConformed to requirements
    Package PerformanceMaster Shipper Carton IntegrityConformed to requirements
    Shelf Carton ConditionConformed to requirements
    Sterile Barrier IntegrityConformed to requirements
    Sterile Barrier Seal StrengthConformed to requirements
    Ease of Opening PouchConformed to requirements
    TruSeal Access System ContainmentConformed to requirements
    Tray ConditionConformed to requirements
    Removal From PackagingConformed to requirements
    DFU and eDFU Card Print QualityConformed to requirements
    Label Adhesion and Print QualityConformed to requirements
    BiocompatibilityCytotoxicityConformed to requirements (biocompatible)
    SensitizationConformed to requirements (biocompatible)
    Intracutaneous ReactivityConformed to requirements (biocompatible)
    Acute Systemic ToxicityConformed to requirements (biocompatible)
    Complement ActivationConformed to requirements (biocompatible)
    Materials Mediated PyrogenicityConformed to requirements (biocompatible)
    Hemolysis Direct ContactConformed to requirements (biocompatible)
    Hemolysis Extract MethodConformed to requirements (biocompatible)
    In Vitro Platelet and Leukocyte CountConformed to requirements (biocompatible)
    Partial Thromboplastin TimeConformed to requirements (biocompatible)

    2. Sample size used for the test set and the data provenance: Not applicable. This document describes physical, chemical, and biological testing of a medical device, not a data-driven test set for an algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for device performance testing. Ground truth in this context refers to established standards and test methodologies (e.g., ISO standards, internal Boston Scientific specifications).

    4. Adjudication method for the test set: Not applicable for device performance testing. Performance was determined by meeting pre-defined physical, mechanical, and biological specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For device performance and biocompatibility, the "ground truth" is defined by established engineering specifications, relevant industry standards (e.g., EN ISO 10555-1: 2013), and regulatory guidance documents (e.g., Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 2010).

    8. The sample size for the training set: Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.