Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180715

    Validate with FDA (Live)

    Date Cleared
    2018-07-04

    (107 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The React™ 68 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath. The proximal end of the React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The React™ 68 Catheter is designed with a hydrophilic coating.

    AI/ML Overview

    This is a 510(k) premarket notification for the React™ 68 Catheter. The document focuses on demonstrating substantial equivalence to predicate devices through design comparisons and extensive bench testing. Animal and clinical studies were not deemed necessary for this submission.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes numerous functional, biocompatibility, and sterilization-related tests. The "Acceptance Criteria" for each test are implicitly that the device "met the acceptance criteria" or demonstrated results within acceptable limits for safe and effective use. The "Reported Device Performance" is the stated outcome of each test.

    Test DescriptionAcceptance Criteria (Implicit)Reported Device Performance
    Biocompatibility:
    Cytotoxicity (Elution Method)No evidence of cell lysis/toxicity (grade < 2)Grade of 0 (no reactivity), considered non-cytotoxic
    Sensitization (Guinea Pig Maximization)No evidence causing delayed dermal contact sensitizationNo evidence of causing delayed dermal contact sensitization
    Irritation (Intracutaneous Reactivity)Difference between test/control scores of 0Difference of 0.0 for NaCl and sesame oil extracts, considered non-irritant
    Acute Systemic Toxicity (Systemic Toxicity)No mortality or evidence of systemic toxicityNo mortality or evidence of systemic toxicity
    Hemocompatibility (Hemolysis)Hemolytic index within acceptable limitsHemolytic index of 0.9% (direct), 0.0% (extract)
    Hemocompatibility (Complement Activation)Test article not statistically higher than controlsTest article not statistically higher than controls, not a potential activator of complement system
    Hemocompatibility (Thrombogenicity)Minimal thrombus formationMinimal thrombus formation after 4 hours
    Pyrogenicity (Material Mediated)No animal temperature rise $\ge$ 0.5°C over baselineNo animal showed a temperature rise of 0.5°C or more
    Performance Data – Bench (Microbial):
    Ethylene Oxide ResidualMet ISO 10993-7 acceptance criteriaMet the acceptance criteria for ethylene oxide residual
    Ethylene Chlorohydrin ResidualMet ISO 10993-7 acceptance criteriaMet the acceptance criteria for ethylene chlorohydrin residual
    BioburdenMet ISO 11737-1 acceptance criteriaMet the acceptance criteria for bioburden
    Bacterial EndotoxinMet ANSI/AAMI ST72 and USP <161> acceptance criteriaMet the acceptance criteria for bacterial endotoxin
    Performance Data – Bench (Performance):
    Visual InspectionMet acceptance criteriaMet the acceptance criteria for visual inspection
    Dimensional MeasurementsMet acceptance criteria for all measured dimensionsMet the acceptance criteria for dimensional measurements
    Tip BucklingMet acceptance criteria for maximum compressive forceMet the acceptance criteria for tip buckling
    Kink ResistanceMet acceptance criteria for maximum kink diameterMet the acceptance criteria for kink resistance
    ParticulateMet USP <788> acceptance criteriaMet the acceptance criteria for particulate evaluation
    Coating LubricityMet acceptance criteria for average friction forcesMet the acceptance criteria for coating lubricity
    Tensile StrengthMet ISO 10555-1 Annex B acceptance criteriaMet the acceptance criteria for tensile strength at hub and shaft
    Liquid LeakMet ISO 10555-1 Annex C acceptance criteriaMet the acceptance criteria for liquid leak
    Corrosion ResistanceMet ISO 10555-1 Annex A acceptance criteriaMet the acceptance criteria for corrosion resistance
    Hub Aspiration ResistanceMet ISO 10555-1 Annex D acceptance criteriaMet the acceptance criteria for hub air aspiration
    CompatibilityMet acceptance criteria for visual damage during useMet the acceptance criteria for delivering and retrieving interventional devices
    RadiopacityMet acceptance criteria (markerband/wall thickness, fluoroscopy)Met the acceptance criteria for radiopacity
    Luer StandardsMet ISO 594-1 and ISO 80369-7 acceptance criteriaMet the acceptance criteria for luer standards
    Torque StrengthAble to withstand typical clinical torsional forcesAble to withstand torsional forces typical of clinical use
    Dynamic PressureAble to withstand typical clinical pressuresAble to withstand pressures typical of clinical use
    Coating IntegrityCoating remained covered and lubriciousRemained coated and lubricous
    UsabilityMet acceptance criteria for maneuverability and flexibilityMet the acceptance criteria for usability

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes for each specific test within the bench testing. However, such tests would typically involve a statistically relevant number of units to ensure reliability. The data provenance is from bench testing conducted by the manufacturer, Micro Therapeutics, Inc. d/b/a ev3 Neurovascular. The data is prospective as it was generated specifically for this submission. The country of origin of the data is not explicitly stated but can be inferred as the United States given the FDA submission and the company's address in Irvine, CA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This device is a physical catheter, and its performance is evaluated through objective, standardized physical and chemical tests (e.g., ISO standards, USP, visual inspection, dimensional measurements). There is no "ground truth" derived from expert interpretation in the context of diagnostic performance testing of, for example, medical images.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the ground truth for this type of device's performance is established through objective laboratory testing against predetermined specifications and international standards, not through adjudications of expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This is a physical medical device (catheter) and not an AI or imaging diagnostic device. Therefore, MRMC studies are irrelevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the React™ 68 Catheter's performance is defined by its compliance with:

    • Industry Standards: Such as ISO 10993 (Biocompatibility), ISO 11737-1 (Bioburden), ANSI/AAMI ST72 (Bacterial Endotoxin), USP <161>/<788> (Bacterial Endotoxin/Particulate), ISO 10555-1 (Catheter testing), ISO 594-1/80369-7 (Luer standards).
    • Predicate Device Performance: Demonstrating substantial equivalence to legally marketed predicate devices in terms of materials, dimensions, and functional characteristics.
    • Engineering Specifications: Internal design requirements for dimensions, material properties, and functional performance (e.g., tip buckling, kink resistance, lubricity, tensile strength).

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI or machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1