Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180613

    Validate with FDA (Live)

    Date Cleared
    2018-07-12

    (126 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTELITE UNIVERSAL FLOW device is indicated for use as a tooth shade resin material in dental procedures, such as:

    • Direct anterior and posterior restorations
    • Cavity base or liner
    • Blocking out cavity undercuts before fabricating indirect restorations
    • Repair of porcelain/composite
    Device Description

    The ESTELITE UNIVERSAL FLOW tooth shade resin device is a light-cured, radiopaque composite resin for use in anterior and posterior restorations. The device incorporates Radical-Amplified Photopolymerization initiator technology (RAP technology), which facilitates a shortened light curing time and ample working time.
    ESTELITE UNIVERSAL FLOW comes in three viscosities: High, Medium, and Super Low. The device is available in either a syringe or pre-loaded tip. Each type contains spherical silica-zirconia filler and composite filler. The spherical filler has mean particle size of 200 and a particle size range of 100 to 300 nm.
    The device does not come sterilized and is not intended to be sterilized prior to use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ESTELITE UNIVERSAL FLOW device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (from ISO 4049:2009)Reported Device Performance (ESTELITE UNIVERSAL FLOW)
    Sensitivity to ambient lightConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    Depth of cureConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    Flexural strengthConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    Water sorptionConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    SolubilityConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    ShadeConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    Color stabilityConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    RadiopacityConformed to ISO 4049 requirementsConformed to ISO 4049 requirements
    SterilizationNon-sterile (as per predicate)Non-sterile
    Shelf life3 years at 0-25°C (32-77°F) (as per predicate)3 years at 0-25°C (32-77°F)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the non-clinical tests (e.g., how many specimens were tested for flexural strength). It only states that the testing was conducted in accordance with ISO 4049:2009. ISO standards typically specify minimum sample sizes for such tests.

    The data provenance is non-clinical testing conducted by the manufacturer, Tokuyama Dental Corporation. The country of origin of the data is not specifically mentioned, but the manufacturer is based in Japan. This testing is prospective for the ESTELITE UNIVERSAL FLOW device, as it was performed to demonstrate conformity.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is a dental material, and the "ground truth" for its performance characteristics is established by adherence to physical and chemical property standards (ISO 4049:2009), not by expert interpretation of images or clinical outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. The "test set" involves physical and chemical property measurements against a standard, not interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a dental restorative material, not an AI-powered diagnostic or assistive technology for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a material, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance claims is the requirements specified in ISO 4049:2009, "Dentistry -- Polymer-based Restorative Materials." The device's physical and chemical properties were measured and compared against these established international standards.

    For biocompatibility, the ground truth relies on the established safety record of identical ingredients used in previously authorized devices (ESTELITE BULK FILL Flow, ESTELITE SIGMA QUICK, and ESTELITE FLOW QUICK) over many years without reportable adverse incidents. This is justification for not requiring further biocompatibility testing under Section 6 of ISO 10993-1.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1