    K Number
    K180404

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-16

    (30 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    3M™ Clarity™ Ultra Brackets and Adhesive Coated Appliance System are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth.

    Device Description

    3M™ Clarity™ Ultra Brackets and Adhesive Coated Appliance System are intended for use in orthodontic treatment. Clarity Ultra Bracket is an esthetic bracket with a sliding door. The brackets are composed of aluminum oxide, which gives them a clear appearance. A nickel titanium wire serves as part of the door mechanism and is attached to the bracket using a cured adhesive. The brackets incorporate a water-soluble color placement indicator system. The archwire slot, vertical slot, and door are marked to facilitate bracket positioning. Tie wing(s) and/or door are color coded with dots to facilitate bracket identification. The brackets are available with or without APC™ Flash-Free pre-coated adhesive. The pre-coated adhesive ensures a consistent amount of adhesive on the bracket.

    AI/ML Overview

    The provided document is a 510(k) Summary for the 3M™ Clarity™ Ultra Self-Ligating Brackets and Adhesive Coated Appliance System. It details the device's technical specifications and a comparison to predicate devices, but it does not contain information typically found in a study proving a device meets acceptance criteria, particularly for AI/machine learning devices.

    The document discusses "Performance Testing" which refers to bench testing of mechanical properties, not clinical performance or AI algorithm performance. Therefore, I cannot extract information related to AI acceptance criteria or studies from this document.

    For completeness, I will extract what is available regarding the device's performance testing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states: "Bench testing consistent with the clinical application and intended use was used to evaluate the performance of Clarity™ Ultra Self-Ligating Brackets and Adhesive Coated Appliance System compared to the predicate devices, Clarity™ Advanced and Damon 4Clear. Properties evaluated include: rotation play angle, wire rotation strength, shear-peel bond strength, torque strength, and squeeze debond strength. 3M Health Care has tested Clarity™ Ultra Self-Ligating Brackets and Adhesive Coated Appliance System and has concluded that the performance tests results demonstrate substantial equivalence to the predicate devices."

    However, specific numerical "acceptance criteria" values or "reported device performance" values for these properties are not provided in the document. It only states that the device was found to be "substantially equivalent" to the predicate devices.

    2. Sample size used for the test set and the data provenance: Not applicable. This document describes bench testing on the device itself, not a test set of data for an AI algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI algorithm is not relevant to this bench testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to mechanical bench testing. The "ground truth" in this context would be the measured mechanical properties of the devices.

    8. The sample size for the training set: Not applicable. This device is not an AI algorithm.

    9. How the ground truth for the training set was established: Not applicable.