Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180292

    Validate with FDA (Live)

    Date Cleared
    2018-10-12

    (253 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

    Device Description

    The PENTAX Medical Video Upper GI Scope EG34-i10 is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the upper gastrointestinal tract. The PENTAX Medical Video Upper GI Scope EG34-i10 is compatible with the PENTAX Medical EPK-i7010 Video Processor (K150618) and PENTAX Medical EPK-i5010 Video Processor (K143727).

    The PENTAX Medical Video Upper GI Scope EG34-i10 is composed of three main components: an Insertion Portion, Control Body and PVE Connector. The Insertion is inserted into the body cavity of patient. The Insertion Portion includes the Distal End and Bending Section. The Distal End of the Insertion Portion includes the Objective Lens, Light Guide, Instrument Channel, Air/Water Nozzle, and Water Jet Nozzle.

    The Air/Water Feeding System is used to deliver the air and water to the Objective Lens from the Air/Water Nozzle. The Air/Water Feeding Valve is attached to the Air/Water Cylinder. When the hole at the top of Air/Water Feeding Valve is covered, the air is delivered. When the Air/Water Feeding Valve is pushed, the water is delivered. A Water Jet System is used to stream forward the sterile water from the Water Jet Nozzle.

    The Control Body is held by the user's hand. The Control Body includes the Angulation Control Knob, Angulation Lock Knob/Lever, Air/Water Cylinder, Suction Cylinder, Remote Control Button, and Instrument Channel Inlet. The Bending Section is bent by the Angulation Control Knob to operate the endoscope angulation Lock Knob/Lever is used to adjust the rotation torque of the Angulation Control Knob. The Remote Button is used to operate the function of video processor and external device from the control body, as necessary.

    The Inlet Seal is attached to the Instrument Channel Inlet. Endotherapy Device such as Biopsy Forceps may be inserted from the Instrument Channel Inlet into the body cavity through the instrument channel.

    The PVE Connector is connected to the Video Processor via an Electrical Contacts. The Light Guide of the Distal End is used to illuminate the body cavity by light that is carried through the Light Carrying Bundle. The Light Carrying Bundle guides the light from Light Guide Plug that is connected to the Light Source inside the video processor. The CCD built into the Distal End receives reflected light (image data) from the body cavity, and sends the image data to the Video Processor through the video cable. The image data is converted into the image signal by the Video Processor, and the image inside the body cavity is displayed on the Monitor.

    The PENTAX Medical Video Upper Gl Scope EG34-i10 is a reusable semi-critical device. Since it is packaged non-sterile, it must be high-level disinfected or sterilized before initial use. Prior to each subsequent procedure, it must be subjected to an appropriate cleaning and either highlevel disinfection or sterilization processes.

    AI/ML Overview

    This FDA 510(k) summary describes the PENTAX Medical Video Upper GI Scope EG34-i10, a reusable semi-critical device intended for endoscopy and endoscopic surgery within the upper digestive tract. The submission primarily focuses on demonstrating substantial equivalence to its predicate device, the PENTAX Video Upper GI Scopes EG29-i10 (K131902), by showing that minor differences in dimensional specifications and materials do not adversely impact safety or effectiveness.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria and reported device performance in the format usually seen for AI/software-based medical devices with specific performance metrics like sensitivity, specificity, or AUC. Instead, it details various engineering and performance tests conducted to establish substantial equivalence. The "acceptance criteria" are implied by the successful completion and "passing results" of these engineering tests, confirming that the device meets relevant standards and performs comparably to the predicate.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Reprocessing Validation:All acceptance criteria must be satisfied for each test.All acceptance criteria were satisfied for:
    - Soil Accumulation Study(e.g., demonstrably low soil retention)Performed with passing results.
    - Cleaning Efficacy Study(e.g., effective removal of soil/microorganisms)Performed with passing results.
    - HLD Efficacy Study(e.g., effective high-level disinfection)Performed with passing results.
    - Rinsing Validation(e.g., effective removal of disinfectants)Performed with passing results.
    - Sterilization Validation(e.g., effective sterilization where applicable)Performed with passing results.
    BiocompatibilityCompliance with ISO 10993-1.Conducted with passing results.
    Electrical SafetyCompliance with IEC 60601-1, IEC 60601-2-18, ANSI/AAMI ES60601-1.Complies with all listed standards.
    EMCCompliance with IEC 60601-1-2 / EN 60601-1-2.Complies with listed standard.
    Software V&VSoftware is identical to the predicate device.Utilizes the same software as the predicate device.
    Optical Testing:
    - Light DistributionComparable to predicate device.Comparable to predicate device (EG29-i10).
    - Spectral DistributionComparable to predicate device.Comparable to predicate device (EG29-i10).
    - Total Luminous FluxComparable to predicate device.Comparable to predicate device (EG29-i10).
    - Photobiological Safety (IEC 62471)Risk group 2 or lower; safety equivalent to predicate.All subject/predicate combinations are risk group 2 or lower and considered equivalent.
    Usability TestingNo difference in operations; usability equivalent to predicate.No difference in all operations, usability is equivalent.

    2. Sample size used for the test set and the data provenance

    The document describes device performance testing and validation, not a study with a traditional "test set" of clinical data for a diagnostic algorithm. The tests conducted (reprocessing, biocompatibility, electrical safety, EMC, optical, usability) involve physical samples of the device and its components, or measurements under controlled lab conditions.

    • Sample Size: Not specified in terms of number of devices or units used for each test, but it is implied that a sufficient number was tested to demonstrate compliance and comparability. These are engineering tests, not statistical clinical trials with patient populations.
    • Data Provenance: The data comes from internal testing conducted by PENTAX Medical, as implied by the "PENTAX Medical completed optical testing" and "Testing revealed..." statements. These are laboratory-based, prospective engineering tests rather than retrospective or prospective clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is not an AI/ML algorithm requiring expert ground truth or adjudication for diagnostic performance. Its validation is based on engineering standards and comparisons to a physical predicate device.

    4. Adjudication method for the test set

    Not applicable. As this is not a study requiring expert clinical adjudication for diagnostic outcomes, no such method was employed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an endoscope, a hardware instrument, not an AI-assisted diagnostic tool. No MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not involve a standalone algorithm for performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the context of clinical diagnostic ground truth. The "ground truth" for the engineering tests would be defined by the specifications, standards (e.g., IEC, ISO), and the performance characteristics of the predicate device. For example:

    • For reprocessing: Standardized microbiological assays or chemical residue tests demonstrating inactivation/removal targets.
    • For optical testing: Direct measurement of light properties and comparison to predicate's measured values, and adherence to photobiological safety standards.

    8. The sample size for the training set

    Not applicable. This device does not involve a training set as it is not an AI/ML system.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a training set.