Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180201

    Validate with FDA (Live)

    Date Cleared
    2018-05-23

    (119 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Access System is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical Access System is comprised of an Access Catheter and a Delivery Catheter. The distal portion of the Access Catheter is a single-lumen, variable stiffness catheter. Like the predicate device, the proximal portion is a stainless-steel control wire. The Delivery Catheter is a hubbed, singlelumen variable stiffness catheters are hydrophilically coated. The Route 92 Medical Access System is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    The provided text describes the acceptance criteria and the studies conducted for the Route 92 Medical Access System.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for all tests was "PASS". The reported device performance for all tests was also "PASS". This is explicitly stated for each test category.

    Test CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityDemonstrates non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic, minimal complement activation, non-activator of intrinsic coagulation, non-hemolytic, similar thromboresistance, non-mutagenic, and non-clastogenic.All tests passed, concluding the device meets biocompatibility requirements.
    Dimensional VerificationConformance to specificationsPASS
    Luer IntegrityPer ISO 80369-7:2016PASS
    RHV SealingSealing around catheter shaftsPASS
    Tensile StrengthStrength of catheter sections and bondsPASS
    Kink ResistanceResistance to kinking in a defined bend diameterPASS
    Torsion ResistanceIntegrity after rotationPASS
    Tip FlexibilityFlexibility of the tipPASS
    Air LeakagePer ISO 10555-1:2013 Annex DPASS
    Liquid Leakage / Static BurstPer ISO 10555-1:2013 Annex CPASS
    Dynamic BurstMechanical integrity up to specified pressuresPASS
    FlowContrast media flow rates equivalent to predicate devicesPASS
    Hydrophilic Coating IntegrityIntegrity after multiple insertion/withdrawal cyclesPASS
    Particulate RecoveryEffluent water tested per USP <788> after multiple cyclesPASS
    Corrosion ResistancePer ISO 10555-1:2013 Annex APASS
    Simulated Use TestingDeliverability and compatibility in a neurovascular modelPASS
    Packaging IntegrityPer ISO 11607-1 Part 1 & ISO 11607-2 Part 2PASS
    RadiopacityRadiopacity in an animal model under fluoroscopyPASS

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the specific "sample size" for each individual test within the performance and biocompatibility sections. Instead, it generally refers to "All samples" or "test animals" for biocompatibility.

    • For Biocompatibility: The tests used "animals" (Guinea Pigs, Rabbits, Mice, potentially others for in-vitro tests). The number of animals per test is mentioned for some (e.g., "None of the animals", "rabbits administered").
    • For Performance Testing: The results consistently state "All samples met the pre-determined acceptance criteria." without specifying the number of samples.
    • Data Provenance: The studies are non-clinical, meaning they are laboratory or animal-based tests. The document does not specify the country of origin of the data, nor whether it is retrospective or prospective, though non-clinical tests are generally prospective laboratory studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The studies described are non-clinical in nature, focusing on the physical and biological characteristics of the device. There is no mention of "experts" establishing a "ground truth" for a test set in the context of human evaluation or diagnostic accuracy, as this device is a medical instrument, not a diagnostic algorithm.

    4. Adjudication Method for the Test Set

    This information is not provided and is not applicable given the nature of the non-clinical testing performed. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies evaluating human interpretation (e.g., image reading) where disagreement among experts needs resolution.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic AI systems or those involving human interpretation of data, which is not the function of the Route 92 Medical Access System.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied to diagnostic AI is not directly applicable here. For the non-clinical tests:

    • Biocompatibility: Ground truth is established by standardized biological evaluation methods and safety endpoints (e.g., no signs of toxicity, irritation, sensitization).
    • Performance Testing: Ground truth is established by engineering specifications, recognized industry standards (e.g., ISO standards), and functional requirements (e.g., flow rates, mechanical integrity).

    8. The Sample Size for the Training Set

    This information is not applicable. This document describes the testing of a physical medical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1