510(k) Data Aggregation

    K Number
    K180073

    Validate with FDA (Live)

    Date Cleared
    2018-08-29

    (232 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VitaVitro Vitrification Kit is intended for the vitrification of human blastocysts for assisted reproduction technologies (ART). This kit is designed for use with VitaVitro Warming Kit.

    VitaVitro Warming Kit is intended for the warming of human blastocysts that have undergone vitrification using VitaVitro Vitrification Kit for ART procedures.

    VitaVitro Straw Set is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human blastocyst stage embryos.

    Device Description

    VitaVitro Vitrification and Warming Kits, and the VitaVitro Straw Set are used for vitrification, storage, and warming of blastocysts as part of human assisted reproduction technology (ART) procedures.

    The VitaVitro Vitrification Kit includes three media components, Human Holding Medium (HHM), Human Vitrification Solution 1 (HV1) and Human Vitrification Solution 2 (HV2). All three media have the same base formulation, but HV1 and HV2 contain the cryoprotectants ethylene glycol and dimethyl sulfoxide in increasing concentrations. HV2 also includes sucrose as a non-permeating cryoprotectant. During the vitrification process, blastocysts are first equilibrated in HHM, and then exposed sequentially to HV1 and HV2. Using this methodology, the permeating cryoprotectants replace water in the blastocyst prior to vitrification and storage in liquid nitrogen. The VitaVitro Vitrification Kit includes single 1.0 ml vials of HHM, HV1, and HV2.

    The VitaVitro Warming Kit includes three media used stepwise for warming and removing cryoprotectants from vitrified blastocysts. The VitaVitro Warming Kit includes Human Warming Solution 1 (HW1), Human Warming Solution 2 (HW2) and HHM. These media products are the same with the exception that HW1 and HW2 also include sucrose in decreasing concentrations to aid in the rehydration of the blastocysts. The VitaVitro Warming Kit includes two 1.5 ml vials of HW1, one 1.0 ml vial of HW2, and one 1.8 ml vial of HHM.

    The VitaVitro Straw Set consists of two components, a straw that holds the blastocysts, and a container that is used to protect the straw. Both components are made from copolyester. The closed end of the container also includes a weight to aid in maintaining the orientation of the device in liquid nitrogen and prevent floating. During use, the container is pre-cooled by placing the closed end in liquid nitrogen with the open end extending above the level of the liquid nitrogen. Straw loading is done by placing the narrow tip of the straw in a 1 µl drop of vitrification media, which draws the media and blastocysts into the storage device. The straw component is then inserted into the pre-cooled container component to effect vitrification. The container component is then sealed to prevent contact between the samples and liquid nitrogen.

    All of the products are provided sterile with a six-month shelf-life. The media in the vitrification and warming kits undergo aseptic filtration, while the storage devices are ethylene oxide sterilized.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the VitaVitro Vitrification Kit, VitaVitro Warming Kit, and VitaVitro Straw Set, comparing it to a predicate device (Cryotop® Vitrification Kit and Cryotop® Thawing Kit, K160864).

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (Subject Device: VitaVitro)Reported Device Performance (VitaVitro)
    Vitrification & Warming Kit Media
    pH7.2 - 7.6pH: 7.2-7.6 for all solutions (met)
    Osmolality (HHM)295-315Met (specific value not given, but stated to meet acceptance specifications)
    Osmolality (HV1)N/A (not specified in comparison table)N/A (not specified in comparison table)
    Osmolality (HV2)N/A (not specified in comparison table)N/A (not specified in comparison table)
    Osmolality (HW1, diluted 1:1)600-850Met (specific value not given, but stated to meet acceptance specifications)
    Osmolality (HW2)850-1000Met (specific value not given, but stated to meet acceptance specifications)
    Endotoxin< 0.25 EU/mL<0.25 EU/mL (met)
    Mouse Embryo Assay (2-Cell MEA, blastocysts 72h)≥ 80% development to blastocyst at 72h (in comparison to control)≥80% development to blastocyst at 72h (met)
    SterilityAseptic processing and sterilityMet ISO 13408-1:2008 and ISO 13408-2:2003, and USP<71> Sterility Testing (met)
    Shelf-lifeMaintain product specifications for pH, osmolality, MEA, endotoxin, and sterility over six monthsMet (product specifications maintained over six-month shelf-life)
    Straw Set (Cryopreservation Storage Device)
    Cooling RateNot an acceptance criteria but a measured characteristic5,127 °C/min
    Warming RateNot an acceptance criteria but a measured characteristic17,899 °C/min
    Endotoxin< 0.5 EU/device<0.5 EU/device (met)
    Mouse Embryo Assay (2-Cell MEA, blastocysts 72h)≥ 80% development to blastocyst at 72h (in comparison to control)≥80% development to blastocyst at 72h (met)
    Sealed Device DurabilityNo burst/breakage after 30 second immersion in liquid nitrogenNo burst/breakage after 30 second immersion in liquid nitrogen (met)
    Simulated Distribution TestingMet ASTM D4169-16Met (testing performed per ASTM D4169-16)
    Package IntegrityMet ASTM F1929:2012 (dye penetration) and ASTM F88:2009 (seal strength)Met (testing performed per specified ASTM standards)
    Shelf-lifeMaintain product specifications for endotoxin, MEA, cooling/warming rate, and sealed device durability over six monthsMet (product specifications maintained over six-month shelf-life)

    2. Sample Size and Data Provenance for Test Set

    • Sample Size for Test Set: The document does not explicitly state the sample size (number of embryos or tests) used for the Mouse Embryo Assay (MEA) or other performance tests. It refers to "Two-cell mouse embryos" without specifying the quantity.
    • Data Provenance: The document does not specify the country of origin of the data. The manufacturer is Shenzhen VitaVitro Bio-tech Co., Ltd. in China. The study appears to be retrospective, as it involves non-clinical performance testing of the device for submission.

    3. Number of Experts and Qualifications for Ground Truth

    • This information is not provided in the document. The studies described are non-clinical, primarily laboratory-based performance tests (e.g., pH, osmolality, endotoxin, mouse embryo assay, physical durability), rather than clinical studies requiring expert interpretation of results.

    4. Adjudication Method for Test Set

    • This information is not applicable as the studies are non-clinical performance tests, not clinical studies with human subject data requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data with and without AI assistance. The VitaVitro kits are reproductive media and storage devices, not diagnostic software.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone (algorithm only) performance study was not done. This device is a physical product (media and a straw set), not a software algorithm.

    7. Type of Ground Truth Used

    • For the Vitrification Kit and Warming Kit media, the "ground truth" or reference standard for performance was established through:
      • Defined Chemical/Physical Specifications: pH, osmolality, endotoxin levels.
      • Biological Assay Performance: Mouse Embryo Assay (MEA) results (percent development to blastocyst) compared against an internal control group.
      • Sterility Standards: Adherence to ISO and USP standards.
      • Shelf-life stability: Retention of the above characteristics over time.
    • For the Straw Set, the "ground truth" or reference standard was established through:
      • Physical Characteristics: Cooling/warming rates, sealed device durability (no burst/breakage).
      • Biological Assay Performance: Mouse Embryo Assay (MEA) results from extracts compared against an internal control group.
      • Sterility and Packaging Integrity Standards: Endotoxin levels, simulated distribution testing, and package integrity testing per ASTM standards.
      • Shelf-life stability: Retention of the above characteristics over time.

    8. Sample Size for Training Set

    • This information is not applicable as the document describes non-clinical performance testing for a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for Training Set was Established

    • This information is not applicable as there is no training set for an AI/ML algorithm.