Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173953

    Validate with FDA (Live)

    Device Name
    uDR 780i
    Date Cleared
    2018-01-26

    (30 days)

    Product Code
    Regulation Number
    892.1680
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The uDR 780i digital medical x-ray system is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other anatomic sites. Applications can be performed with the subject sitting, standing, or lying in the prone or supine position. Not for mammography.

    Device Description

    The uDR 780i is a digital radiography (DR) system that is designed to provide radiography examinations of sitting, standing or lying patients. It consists of the following components: Tube Ceiling Suspension with tube and collimator, Bucky Wall Stand, Elevating Table, High Voltage Generator, wireless flat panel detectors and an acquisition workstation. The system generates images which can be transferred through DICOM network for printing, review and storage.

    AI/ML Overview

    Based on the provided text, the device in question (uDR 780i) is a digital medical X-ray system, not an AI-powered diagnostic device. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on demonstrating substantial equivalence to a predicate conventional X-ray system, rather than an AI/ML algorithm's performance on a diagnostic task like detecting pathology.

    The document is a 510(k) summary for a medical device. For devices like this, the "acceptance criteria" are generally compliance with recognized performance standards and demonstration that the device's technical specifications and intended use are substantially equivalent to a legally marketed predicate device. The "study" largely involves non-clinical bench testing and a review of clinical images by an expert to confirm diagnostic quality.

    Here's the breakdown of information based on the typical requirements for an AI/ML medical device, applied to this conventional X-ray system where applicable:

    Acceptance Criteria and Device Performance (for a Conventional X-ray System)

    The acceptance criteria are not explicitly laid out in a quantifiable table as they would be for an AI diagnostic device (e.g., sensitivity, specificity thresholds). Instead, the "acceptance" is based on demonstrating substantial equivalence to a predicate device (NOVA FA DR System, K133782) regarding intended use, technological characteristics, safety, and effectiveness.

    Table 1: Comparison of Technology Characteristics (Relevant for "Performance")

    ItemProposed Device uDR 780iPredicate Device NOVA FA DR SystemRemark (If applicable, implies meeting acceptance through equivalence)
    General
    Product CodeKPRKPRSame
    Regulation No.892.1680896.1680Same
    ClassIIIISame
    Intended UseDiagnostic radiographic exposures of skull, spinal column, chest, abdomen, extremities, other anatomic sites for adults and pediatrics. Not for mammography.Diagnostic radiographic exposures of skull, spinal column, chest, abdomen, extremities, other body parts for adults and pediatrics. Not for mammography.Same
    Specifications (Selected - see full Table 1 for all)
    High Voltage Generator
    Rated Power/kW65kW/80kW50kW/65kW/80kWNote 1: Does not affect safety/effectiveness as clinical applications achievable under 65kW/80kW.
    Max. tube Voltage (kV)150kV150kVSame
    Shortest exposure time1ms1msSame
    X-Ray Tube Assemble
    Focus Nominal Value0.6/1.2mm0.6/1.2mmSame
    Maximum peak voltage150kV150kVSame
    Anode Heat Content300kHU/400kHU300kHU/400kHUSame
    Anode Target Angle12°12°Same
    X-ray tube assembly Heat content900kJ (1.3MHU) / 1111kJ (1.5MHU)900kJ (1.3MHU) / 950kJ (1.33MHU)Note 2: Determines continuous exposure time, but not image quality, thus no impact on safety/effectiveness.
    Flat Panel Detector
    ConfigurationBattery or AC operatedBattery or AC operatedSame
    Digital PanelsAmorphous silicon (a-Si)Amorphous silicon (a-Si)Same
    ScintillatorCesium iodide (CsI)Cesium iodide (CsI)Same
    Specifications3320x3408 125μm3320x3408 125μmSame
    Effective radiographic size41.5cm x 42.6cm41.5cm x 42.6cmSame
    Collimator
    Inherent filtration1mm Al1mm AlSame
    Copper prefilterwithout filter, 0.1 mm, 0.2 mm, 0.3 mm;without filter, 0.1 mm, 0.2 mm, 0.3 mm;Same
    Display
    Specification24inch, 1200x192019inch, 1024 x 1280Note 3: Larger display is user-friendly, no impact on safety/effectiveness.
    DICOMDICOM3DICOM3Same
    Patient Table
    Motorized vertical travel38.2cm40cmNote 4: Small difference, affects user experience but not clinical application or safety/effectiveness.
    X-ray absorption≤0.8mmAl≤1mmAlNote 5: Lower absorption is better, optimizes image quality, no impact on safety/effectiveness.
    Max. patient weight225kg350kgNote 7: 225kg sufficient for most patients, no impact on safety/effectiveness.
    Software Function (e.g., Image Search, Image Viewing, Image measurement, Image Annotation, Raw image Data processing, Post image data processing)YesYesAll "Same" functionality
    Safety
    Electrical SafetyAAMI ANSI ES60601-1; IEC 60601-1Comply with IEC60601-1Same
    EMCComply with IEC60601-1-2Comply with IEC60601-1-2Same
    BiocompatibilityTested (ISO 10993-5, -10)Comply with ISO10993-5, ISO10993-10Same

    Study Information (as applicable for a conventional X-ray system submission)

    1. Sample size used for the test set and the data provenance:

      • The document states "Sample clinical images accompanied by dose information and information detailing acquisition protocols and parameters were reviewed." It does not specify a numerical sample size for this "test set" of images.
      • Data Provenance: Not explicitly stated, but typically for 510(k) submissions of conventional imaging devices, real-world images from a hospital or clinic would be used. The document is from Shanghai United Imaging Healthcare Co., Ltd. in CHINA, implying the images could be from China, but this is not explicitly stated. It's a retrospective review of generated images, not implied as prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • One expert was used: "a board certified radiologist".
      • Qualifications: "board certified radiologist with a statement indicating that images are of diagnostic quality." No specific years of experience are mentioned.
    3. Adjudication method for the test set:

      • "None" explicitly described for establishing ground truth from multiple readers. There was only one radiologist reviewer for the "clinical images."
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

      • No. This type of study (MRMC comparing human readers with and without AI assistance) is not applicable or mentioned because the uDR 780i is a conventional X-ray system, not an AI-powered diagnostic assist device. The "effectiveness" is demonstrated through substantial equivalence to a predicate device and diagnostic quality of images.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a conventional X-ray system. The "performance" is about image quality for human interpretation, not an algorithm's standalone diagnostic ability. The statement "images are of diagnostic quality" indicates that the images produced by the device, when viewed by a human expert, are sufficient for diagnosis.
    6. The type of ground truth used:

      • Expert consensus (from a single board-certified radiologist) on the diagnostic quality of the images generated by the device. It's about image quality, not the presence or absence of specific pathologies based on a gold standard (like pathology or outcomes data).
    7. The sample size for the training set:

      • Not applicable. As a conventional X-ray system, there isn't an AI model with a distinct "training set" in the sense of machine learning. The system's "training" refers to its design and engineering adhering to standards and producing high-quality images.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reason as point 7. The "ground truth" here is the engineering and design principles that ensure the system produces diagnostically acceptable images, confirmed through non-clinical testing and expert review of clinical images.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1