    Device Name
    SpiroSphere
    Date Cleared
    2019-02-15

    (416 days)

    Product Code
    Regulation Number
    868.1890
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SpiroSphere is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older.

    It can be used by physicians in the office or hospital.

    Device Description

    The SpiroSphere is a compact spirometry device. It's Sensor Unit is battery-powered. The Main Unit can be powered by battery or power supply. The SpiroSphere is used to measure inspiratory and expiratory lung function parameters in adults and children 4 years and older. The measured data is saved into the device and can be read out at any time.

    A printer can be connected with the SpiroSphere and all needed data (e.g. reports, Screenshots) can be printed. Moreover it is possible to transfer data by USB, Wifi, 3G, and Ethernet.

    Pulmonary functions

    • Slow Spirometry
    • Forced Spirometry
    • Flow-Volume loop and Volume-Time tracing, pre/post tests
    • Trending capabilities
    AI/ML Overview

    The provided text describes the SpiroSphere, a device for measuring lung function, and its substantial equivalence to a predicate device (SpiroPro). However, it does not contain the detailed information required to answer your specific questions about acceptance criteria and a study proving the device meets them.

    The document is a 510(k) summary for premarket notification to the FDA. It focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing the detailed clinical or performance study results you're asking for.

    Here's a breakdown of why the requested information is absent:

    • Acceptance Criteria and Reported Device Performance (Table): The document states adherence to ATS/ERS standards (2005) for spirometry accuracy and precision and lists various parameters measured (FEV1, FVC, PEF, etc.). However, it does not provide a specific table of acceptance criteria (e.g., +/- x% accuracy) or numerical reported device performance against those criteria. It only generally states that "Tests were performed to confirm that the SpiroSphere meets the recommendations for accuracy and precision for Spirometry of the American Thoracic Society (ATS) according to ATS/ERS standards 2005."
    • Sample Size, Data Provenance, Ground Truth Experts, Adjudication Method, MRMC Study, Standalone Performance, Training Set Sample Size/Ground Truth: These details are typically found in a dedicated clinical or performance study report, which is not included in this summary. The 510(k) summary only mentions general verification and validation activities and conformance to standards.

    In summary, based only on the provided text, I cannot complete the table or answer the specific questions about the study details because that level of detail is not present in this FDA 510(k) summary.

    The document discusses:

    • Device Name: SpiroSphere
    • Predicate Device: SpiroPro (K000648) and SpiroPro with BT (K092324)
    • Intended Use/Indications for Use: To measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older, usable by physicians in office or hospital.
    • Technological Characteristics: Uses a pneumotachograph (Lilly Type) for pressure to flow conversion, measures various parameters (FEV1, FVC, PEF, FEF25-75, VC, IC, ERV) conforming to ATS/ERS Spirometry Standards (2005).
    • Testing Mentioned: Software verification and validation (IEC 62304), Risk analysis (ISO 14971), electrical safety (IEC 60601-1:2012), EMC (IEC 60601-1-2:2014), wireless technology (FDA Guidance 2013), Human Factors/Usability (IEC 62366), Biocompatibility (ISO 10993-1:2009, ISO 18562:2017), Cybersecurity (FDA Guidance 2014), Reprocessing (FDA Guidance 2015).

    To answer your questions thoroughly, a complete performance study report would be required, which typically contains the specific data and methodologies.