    K Number
    K173871

    Validate with FDA (Live)

    Date Cleared
    2018-09-17

    (271 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Callispheres Embolic Microspheres and 8Spheres Embolic Microspheres are intended to be used for the embolization of arteriovenous malformations (AVMs) and hypervascular tumors, including uterine fibroids.

    Device Description

    CalliSpheres and 8Spheres Embolic Microspheres are compressible hydrogel microspheres with a regular shape, smooth surface, and calibrated size, which are formed as a result of chemical modification on polyvinyl alcohol (PVA) materials. CalliSpheres and 8Spheres Embolic Microspheres consist of a macromer derived from polyvinyl alcohol (PVA), and are hydrophilic, non-resorbable, and are available in a range of sizes. The preservation solution is 0.9% sodium chloride solution. The water content of fully polymerized microsphere is 91% ~ 94%. Microspheres can tolerate compression of 30%. CalliSpheres Embolic Microspheres are dyed blue to aid in the visualization of the microspheres in the delivery syringe. 8Spheres Embolic Microspheres are undyed and with natural color. CalliSpheres and 8Spheres Embolic Microspheres are supplied sterile and packaged in sealed glass vials.

    CalliSpheres Embolic Microspheres and 8Spheres Embolic Microspheres are intended to be used for the embolization of arteriovenous malformations (AVMs) and hypervascular tumors, including uterine fibroids. By blocking the blood supply to the target area, the tumor or malformation is starved of nutrients and shrinks in size.

    CalliSpheres and 8Spheres Embolic Microspheres can be delivered through typical microcatheters in the 1.7- 4 Fr range. At the time of use, CalliSpheres and 8Spheres Embolic Microspheres are mixed with a nonionic contrast agent to form a suspension solution. CalliSpheres and 8Spheres Embolic Microspheres are intended for single use and are supplied sterile and non-pyrogenic.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study validating the performance of CalliSpheres Embolic Microspheres and 8Spheres Embolic Microspheres, which are vascular embolization devices.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    The document primarily focuses on non-clinical performance testing (in-vitro bench testing, packaging, shelf-life, sterilization, chemical characterization, and biocompatibility) and an animal study. The "acceptance criteria" are generally implied to be "met predefined acceptance criteria" or "met specifications" for the bench tests, and comparability to a predicate device for the animal study.

    Test ItemAcceptance Criteria (Implied)Reported Device Performance
    In-Vitro Bench Testing
    AppearanceMet predefined acceptance criteria (e.g., USP <1> & <790>)Met predefined acceptance criteria
    pH testMet predefined acceptance criteria (e.g., USP <791>)Met predefined acceptance criteria
    Size range confirmationMet predefined acceptance criteria (e.g., Method adapted from ISO 13322-1:2014)Met predefined acceptance criteria
    Compressibility TestMet predefined acceptance criteria (e.g., retained shape/no breakage after 30% compression for 10s)Met predefined acceptance criteria
    Catheter deliverability testMet predefined acceptance criteria (e.g., no clogging, no breakage/shape change)Met predefined acceptance criteria
    Water content testMet predefined acceptance criteria (e.g., USP <731>)Met predefined acceptance criteria
    Impurities & Residual Solvents TestMet predefined acceptance criteria (e.g., USP <1086>, <467>, <621>, <857>)Met predefined acceptance criteria
    SuspensionMet predefined acceptance criteria (e.g., time to achieve <10 min, time maintaining >4 mins)Met predefined acceptance criteria
    Sterility TestMet predefined acceptance criteria (e.g., USP <71>)Met predefined acceptance criteria
    Bacterial EndotoxinsMet predefined acceptance criteria (e.g., USP <85>, not more than 0.5 EU/mL)Met predefined acceptance criteria
    Packaging & Shelf Life
    Packaging IntegrityMaintain integrity as per ASTM F1980-16 for 2 years shelf lifeAll samples met specifications, supported 2-year shelf life
    Shelf LifeMaintain specifications for 2 yearsAll samples met specifications after accelerated aging equivalent to 2 years
    Sterilization
    Sterility Assurance Level (SAL)10^-6 (per ANSI/AAMI/ISO 17665-1:2006(R)2013)Achieved 10^-6 SAL
    Chemical Characterization
    Risk of Chronic Toxicity & CarcinogenicityLow risk (based on margins of safety from exhaustive extraction)Determined to have low risk
    Biocompatibility
    CytotoxicityNo evidence of cell lysis or toxicity (grade < 2)Met requirements (grade < 2)
    Intracutaneous StudyMet requirements for irritationMet requirements
    Maximization Sensitization StudyNo delayed dermal contact sensitizationNot considered a sensitizer
    Systemic ToxicityNo mortality or evidence of systemic toxicityMet requirements
    HemolysisNon-hemolytic (hemolytic index for direct contact < 2%)Hemolytic index 0.6%, non-hemolytic
    Partial Thromboplastin TimeNot statistically different from negative reference materialMet requirements
    SC5b-9 Complement Activation AssayNot considered a potential activatorNot considered a potential activator
    USP Rabbit Pyrogen StudyTotal rise of rabbit temperatures within acceptable USP requirements, absence of pyrogensMet requirements for absence of pyrogens
    Genotoxicity: Bacterial Reverse MutationNon-mutagenicNon-mutagenic for tested strains
    Genotoxicity: Mouse Lymphoma AssayNot mutagenic (no >2-fold increase in mean mutant frequency)Not mutagenic
    ISO Muscle Implantation StudyMacroscopic reaction not significant compared to negative control; microscopic reaction minimal to slightMacroscopic reaction not significant; microscopic reaction slight
    Systemic Toxicity (13 weeks)No changes in hematology/clinical chemistry related to implantation; no evidence of systemic response; local tissue response comparable to controlNo systemic toxicity; minimal to no reaction in tissue
    Chronic Systemic Toxicity & Carcinogenicity (Biological Risk Assessment)Not considered to have chronic systemic toxicity and carcinogenicity riskNot considered to have chronic systemic toxicity and carcinogenicity risk
    Animal Performance Testing
    Recanalization/Durability of OcclusionComparable to predicate device (Embosphere)Complete embolization effectively achieved
    Local/Systemic Foreign Body ReactionsMild and comparable to predicate deviceMild and comparable
    Embolization EffectivenessComparable to predicate deviceComplete embolization effectively achieved
    Ease of DeliveryEasily delivered to target vessels, comparable to predicateEasily delivered to target vessels
    Rupture/Puncture of Blood VesselsNo rupture or perforationNo vascular rupture or perforation
    Non-target Embolization/Device MigrationComparable to predicate deviceComparable non-target embolization
    Clinicopathological ExaminationNo significant difference in adverse reactions compared to predicateNo significant difference

    2. Sample sized used for the test set and the data provenance

    • In-vitro Bench Testing: The specific sample sizes for each bench test are not explicitly provided, but it's stated that for Packaging Integrity and Shelf Life, "3 batches of each tested for all device specifications."
    • Animal Performance Testing:
      • Sample Size: 24 swine (12 for CalliSpheres Embolic Microspheres, 12 for Embosphere microspheres).
      • Data Provenance: The study was an "animal study... intended to simulate the clinical application... A pig model was chosen." This indicates prospective animal data, conducted to support the device's substantial equivalence. The country of origin for the study is not explicitly stated in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This document describes a pre-market notification (510(k)) for a medical device (embolic microspheres). It does not describe a study involving human experts establishing ground truth for AI model testing. The ground truth for the non-clinical tests is based on established laboratory standards, and for the animal study, it's based on clinical observations, histopathological results, and standard measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not a human expert-based ground truth adjudication for an AI device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a medical device (embolic microspheres), not an AI-assisted diagnostic or triaging tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document is for a medical device, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • In-vitro Bench Testing: Ground truth is based on established international and national standards (e.g., USP, ISO, ASTM guidelines) and predefined acceptance criteria for physical, chemical, and biological properties.
    • Animal Performance Testing: Ground truth was established through:
      • Direct observation and measurement: Ease of delivery, presence of rupture/perforation, non-target embolization.
      • Clinical tests: Blood routine, coagulation function, renal function, liver function.
      • Imaging: DSA angiography.
      • Pathology/Histopathology: Evaluation of local and systemic foreign body reactions, and assessment of embolization effectiveness after necropsy.

    8. The sample size for the training set

    Not applicable. This document describes the validation of a physical medical device (embolic microspheres), not an AI model requiring a training data set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI model.