    K Number
    K173806

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-01

    (76 days)

    Product Code
    Regulation Number
    870.1290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CorPath GRX System is intended for use in the remote delivery and manipulation of guidewires and rapid exchange balloon/stent catheters, and remote manipulation of guide catheters during percutaneous coronary intervention (PCI) procedures.

    Device Description

    The CorPath GRX System is intended to allow physicians to deliver and manipulate commercially available guidewires, rapid exchange balloon/stent catheters and guide catheters during percutaneous coronary intervention procedures. During the use of the CorPath GRX System, the physician maneuvers the devices using intuitive controls under independent angiographic fluoroscopy visual quidance using computer controlled movements while in a seated position away from the radiation source.

    The CorPath GRX System is composed of the following two functional sub-units:

      1. Bedside Unit Which consists of the Extended Reach Arm, Robotic Drive and Single-use Cassette
      1. Remote Workspace Which consists of the Control Console, angiographic monitor(s), hemodynamic monitors, X-ray foot pedal, and optional Interventional Cockpit.

    Commercially available guidewires, rapid exchange balloon/stent catheters, and guide catheters are loaded into the Single-use Cassette. By using the joysticks or the Control Console touch screen, the physician can control the Robotic Drive to advance, retract, and rotate the guidewire, advance and retract the rapid exchange catheter, and advance, retract, and rotate the guide catheter. The Robotic Drive and Control Console communicate via a single communication cable.

    In addition, the CorPath GRX System Software contains a functionality for an automated movement of the guidewire, known as "Rotate on Retract." This feature, when enabled by the physician will rotate the quidewire a set amount upon retraction of the quidewire joystick to facilitate redirection of the guidewire which it is being directed to the lesion location.

    AI/ML Overview

    The medical device in question is the CorPath GRX System, a steerable catheter control system used for remote delivery and manipulation of guidewires and rapid exchange balloon/stent catheters, and remote manipulation of guide catheters during percutaneous coronary intervention (PCI) procedures.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for performance in a quantitative manner. Instead, it focuses on demonstrating substantial equivalence to a predicate device (CorPath GRX System, K160121). The "acceptance criteria" are implied to be that the модифицированный CorPath GRX System maintains the same performance, safety, and functionality as its predicate.

    Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
    Functional Equivalence: Device performs the same intended functions as the predicate device (remote delivery and manipulation of guidewires, rapid exchange balloon/stent catheters, and guide catheters during PCI).The proposed CorPath GRX System and the predicate CorPath GRX System have the same technological characteristics. There have been no changes to the CorPath GRX System with respect to hardware design, materials, packaging, sterilization, or indication. The system allows physicians to deliver and manipulate commercially available guidewires, rapid exchange balloon/stent catheters and guide catheters using intuitive controls under independent angiographic fluoroscopy visual guidance, with computer-controlled movements while in a seated position away from the radiation source. An automated "Rotate on Retract" feature is also present.
    Performance Equivalence: Device performance (e.g., precision, speed, reliability of manipulation) is comparable to the predicate device.Performance Testing was conducted. All testing has demonstrated that the device is substantially equivalent to the predicate device. (Specific quantitative performance metrics are not provided in the summary).
    Safety Equivalence: Device is as safe as the predicate device and does not introduce new or increased safety risks.The determination of substantial equivalence for this device was based on a detailed device description and non-clinical laboratory testing. The testing demonstrated that the device is safe for its intended use and can be considered substantially equivalent to the predicate device. (Specific details of safety testing are not provided in the summary beyond general mention of non-clinical tests).
    Software Equivalence: Software functions identically or acceptably different with no adverse impact.Software Verification and Validation testing was conducted. All testing has demonstrated that the device is substantially equivalent to the predicate device.
    Functional Testing Equivalence: All components and sub-units operate as intended.Functional Testing was conducted. All testing has demonstrated that the device is substantially equivalent to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes non-clinical laboratory tests (Performance Testing, Functional Testing, Software Verification and Validation testing). It does not mention a "test set" in the context of clinical data, human subjects, or images from a specific country. This is a submission for a robotic system, not an AI diagnostic algorithm. Therefore, the testing refers to bench testing and software validation.
    The provenance of these non-clinical tests is internal to the manufacturer (Corindus, Inc.).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is a robotic system for performing PCI, not an AI diagnostic device that requires expert-established ground truth for image interpretation. The "ground truth" for this device's performance would be its ability to correctly manipulate catheters and guidewires as designed and safely.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is a non-clinical device performance and verification study, not a study involving human interpretation of data requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The document assesses the substantial equivalence of an updated robotic system to its predicate, not the impact of AI assistance on human readers for a diagnostic task.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable in the context of typical AI diagnostic algorithms. The CorPath GRX System itself is a system that operates with "human-in-the-loop" (the physician controls it). The "Rotate on Retract" feature is an automated movement (an algorithm within the system), but its standalone performance as a diagnostic tool is not relevant here. The overall system's performance is intrinsically linked to the physician's operation.

    7. Type of Ground Truth Used:

    The "ground truth" for this device is based on engineering specifications, functional requirements, and safety standards demonstrated through non-clinical laboratory testing. This includes verifying that the device correctly performs the mechanical movements for guidewire and catheter manipulation, and that its software functions as intended, adhering to safety protocols. It is not based on expert consensus, pathology, or outcomes data in the way a diagnostic AI would be.

    8. Sample Size for the Training Set:

    Not applicable. This document describes the validation of a robotic system, not the training of an AI algorithm from a dataset of cases. The "training" of the system refers to its design, engineering, and programming according to established principles, not machine learning from a data set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As noted above, there isn't a "training set" in the machine learning sense. The "ground truth" for the development and testing of such a system is established through a rigorous engineering design process, functional specifications, and adherence to relevant standards for medical device safety and performance.