Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173775

    Validate with FDA (Live)

    Date Cleared
    2018-04-06

    (115 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.
    • Fixation of proximal tibial metaphysis osteotomy
    • Hand and foot bone fragment and osteotomy fixation and joint arthrodesis.
    • Fixation of small bone fragments (i.e. small fragments of bone which are not comminuted to the extent to preclude staple placement). These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremities; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in flat bone such as the pelvis, scapula and sternum.
    Device Description

    The Clench® Compression Staple is an implant for use in bone fixation and reconstructive surgical procedures for the management of bone fractures. The implant is designed to provide constant compression over a bone fracture site. Implants are made from Nickel-Titanium Alloy per ASTM F2063 and range in size from 7mm to 20mm wide with two equal length legs ranging from 5mm to 20mm long.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Clench Compression Staple, formatted as requested.

    Note: The provided FDA 510(k) summary focuses on demonstrating substantial equivalence to predicate devices through technical characteristics and mechanical performance testing, rather than clinical efficacy studies often seen with AI/clinical decision support devices. Therefore, many of the requested categories related to human-in-the-loop performance, expert consensus on ground truth for clinical cases, and multi-reader multi-case studies are not applicable to this type of device submission.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Clench Compression Staple are based on demonstrating mechanical performance equivalent to or better than predicate devices, and conformance to recognized industry standards. The reported performance refers to the device meeting these established standards.

    Acceptance Criteria CategorySpecific Criteria / StandardReported Device Performance
    Mechanical PerformanceStatic axial pull-out (per ASTM F564-10)Confirmed that the Subject device performed as intended.
    Mechanical PerformanceStatic Four-Point bending (per ASTM F564-10)Confirmed that the Subject device performed as intended.
    Mechanical PerformanceDynamic Four-Point bending (per ASTM F564-10)Confirmed that the Subject device performed as intended.
    Material PropertiesCorrosion Susceptibility (per ASTM F2129)Confirmed that the Subject device performed as intended.
    Material PropertiesTransformation Temperature (per ASTM F2082)Confirmed that the Subject device performed as intended.
    Substantial EquivalenceComparison to predicate devices (K142292, K002695) in intended use, indications for use, materials, technological characteristics, and labeling.Documentation provided demonstrating substantial equivalence.

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a "test set" in terms of clinical data or patient cases. The testing conducted was mechanical (bench testing) of the device itself.
      • The sample size for each mechanical test (static axial pull-out, static four-point bending, dynamic four-point bending, corrosion susceptibility, transformation temperature) is not explicitly stated in this summary but would have been defined by the ASTM standards and validated through the testing.
      • Data provenance is from bench testing of the manufactured device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. Ground truth for mechanical testing is established by the specified ASTM standards and measurement protocols, not by expert medical opinion. Engineers and lab technicians conduct and interpret these tests according to the standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. This type of adjudication is typically used for clinical image interpretation or expert consensus on patient outcomes, not for mechanical bench testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This device is a bone staple, an orthopedic implant, not an AI or diagnostic imaging device. Therefore, an MRMC study is not relevant or performed.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device (implant), not an algorithm or software. Standalone performance is evaluated through the mechanical and material tests described.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the mechanical and material testing, the "ground truth" is defined by the established performance criteria and methodologies within the relevant ASTM standards (ASTM F564-10, ASTM F2129, ASTM F2082). These standards define how the device should behave under specific conditions.
    7. The sample size for the training set:

      • Not applicable. There is no "training set" in the context of mechanical testing for a physical implant. The design and manufacturing processes are refined through engineering principles, not machine learning training.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1