Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173648

    Validate with FDA (Live)

    Date Cleared
    2018-08-30

    (276 days)

    Product Code
    Regulation Number
    886.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for:

    • viewing the posterior segment of the eye, including two- and three-dimensional imaging
    • cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT)
    • fundus imaging
    • fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA)
    • autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak)
    • performing measurements of ocular anatomy and ocular lesions.

    The device is indicated as an aid in the detection and management of various ocular diseases, including:

    • age-related macular degeneration
    • macular edema
    • diabetic retinopathy
    • retinal and choroidal vascular diseases
    • glaucoma

    The device is indicated for viewing geographic atrophy.

    The SPECTRALIS HRA+OCT and SPECTRALIS OCT include the following reference databases:

    • a retinal nerve fiber layer thickness reference database, which is used to quantitatively compare the retinal nerve fiber layer in the human retina to values of Caucasian normal subjects – the classification result being valid only for Caucasian subjects
    • a reference database for retinal nerve fiber thickness and optic nerve head neuroretinal rim paramements, which is used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values of normal subjects of different races and ethnicities representing the population mix of the USA (Glaucoma Module Premium Edition)
    Device Description

    The Heidelberg Engineering SPECTRALIS HRA+OCT is a device used to image the anterior and posterior segments of the human eye and to aid in the assessment and management of various diseases of the posterior segment, such as age-related macular degeneration, diabetic retinopathy, and glaucoma. The device is capable of acquiring infrared and blue reflectance images (K101223), as well as a composite color image referred to as "MultiColor image" due to addition of a green laser light source (K121993). New reference data for peripapillary retinal nerve fiber layer thickness and optic nerve head parameters have been added in K152205, as well as two (2) new accessory objective lenses (Wide Field, Ultra-Widefield), and OCT Enhanced Depth Imaging mode. In K172649, an updated 55° objective lens (WFO2) with enhanced anti-reflective coating was added to allow for Widefield MultiColor imaging, as well as an updated and faster spectrometer, called OCT2 module, a replacement of the patient and data management system, named Heyex 2, restructured user documentation, and Windows 10 support.

    The modification to the SPECTRALIS is:
    Widefield OCT - In addition to OCT imaging with the standard objective (for posterior segment imaging; 30° field of view) and the Anterior Segment Objective (anterior segment imaging; 30° field of view), OCT imaging is enabled with the Wide Field Objective (WFO) and Wide Field Objective 2 (WFO2). Both objectives provide a 55° diameter circular field of view. The field of view can be reduced by software to 35° and 25°. The only difference between WFO and WFO2 is an enhanced antireflective coating to allow for MultiColor imaging on the WFO2 (cleared in K172649). Both coatings cover the OCT wavelength, so OCT imaging can be enabled for both objective lenses.

    AI/ML Overview

    This document describes the 510(k) summary for the SPECTRALIS HRA+OCT and variants, a medical device for ophthalmic diagnostic imaging.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria described for this device are primarily related to non-clinical performance (bench testing) and maintaining substantial equivalence to a predicate device. There are no explicit quantitative clinical acceptance criteria provided in the document in the format of a table with reported device performance against them.

    However, based on the "NON-CLINICAL PERFORMANCE SUMMARY" and "OPTICAL RESOLUTION TESTING" sections, we can infer some criteria and reported performance:

    Acceptance CriteriaReported Device Performance
    Compliance with FDA Recognized Consensus StandardsThe modified SPECTRALIS was evaluated according to ISO 14971, AAMI/ANSI ES60601-1:2005, IEC 60601-1-2 Edition 3: 2007-03, ISO 10993-1, AAMI/ANSI/ISO 10993-5:2009/(R) 2014, ISO 10993-12 Fourth Edition 2012-07-01, ISO 10993-18:2005, IEC 60825-1 Edition 2.0 2007-03, and AAMI/ANSI/IEC 62304:2006.
    Safety and Effectiveness"was found to meet the requirements of the applicable parts."
    Design Control ComplianceHeidelberg Engineering designed and developed the modified SPECTRALIS per the company's Design Control procedure, complying with FDA Quality System Regulations CFR Part 820 and ISO 13485:2012, and Risk Management procedures complying with ISO 14971:2007.
    Risk AssessmentRisk assessment was conducted on the modified SPECTRALIS; impact of design modifications assessed on the predicate 510(k) cleared device.
    Manufacturing and Testing EquivalenceThe modified SPECTRALIS is manufactured and tested in the exact manner as the predicate 510(k) cleared device.
    Functional EquivalenceBench testing (including light safety, image geometry, image quality assessment) and software verification/validation confirmed the modified SPECTRALIS HRA+OCT functions equivalently to the predicate SPECTRALIS HRA+OCT.
    Optical Lateral Resolution (for WFO lens)Average optical lateral resolution for the Standard Objective (SO) was 13µm. Average optical lateral resolution for the Wide Field Objective (WFO) was 18µm. (This indicates an acceptable loss of optical lateral resolution, roughly 3 resolution elements, consistent with theoretical evaluations for the wide-field lens).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a test set involving human subjects or clinical data in the traditional sense for evaluating device performance against a diagnostic endpoint. The performance evaluation is based on non-clinical bench testing.

    • Sample Size for Test Set: Not applicable, as performance was assessed through bench testing on the device itself and a resolution test chart, not a separate clinical test set.
    • Data Provenance: Not applicable, as it's not clinical data. The testing was performed by Heidelberg Engineering GmbH as part of their design control process.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not explicitly applicable to this submission, as the evaluation is primarily for technical performance and substantial equivalence based on bench testing. There is no mention of "ground truth" derived from expert consensus for clinical disease states in this context. The "ground truth" for optical resolution, for example, is established by the known properties of the 1951 USAF resolution test chart.

    4. Adjudication Method for the Test Set

    Not applicable, as there was no clinical test set requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or reported in this document. This submission focuses on the technical modifications and substantial equivalence of the device, not an AI-assisted diagnostic workflow.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This device, the SPECTRALIS HRA+OCT, is a diagnostic imaging device that produces images for human interpretation and aids in detection and management. It is not an "algorithm only" device in the sense of AI-driven interpretation. Therefore, a standalone algorithm-only performance study is not directly relevant to this device's function as described. Its performance is related to its ability to acquire high-quality images and measurements, which are then used by clinicians.

    7. The Type of Ground Truth Used

    For the bench testing described, the "ground truth" effectively refers to physical standards and theoretical evaluations:

    • Physical standards: Use of a 1951 USAF resolution test chart to objectively quantify optical lateral resolution.
    • Theoretical evaluations: The measured optical resolution values were compared to theoretical expectations.
    • Predicate device's established performance: The primary "ground truth" for this 510(k) is the performance and safety profile of the legally marketed predicate device (K172649), to which the modified device claims substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a modification to an existing imaging device, not a machine learning or AI algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI algorithm mentioned for this device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1