510(k) Data Aggregation

    K Number
    K173580

    Validate with FDA (Live)

    Date Cleared
    2018-02-15

    (87 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vinyl Patient Examination Glove (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Vinyl Patient Examination Gloves. The proposed device is yellow. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The proposed device is non-sterile.

    AI/ML Overview

    The provided text details the 510(k) summary for the "Vinyl Patient Examination Glove (Yellow)" (K173580) and compares it to a predicate device (K163168). The information specifically relates to the technical specifications and performance of the examination gloves, rather than an AI/ML device or a medical imaging system. Therefore, many of the requested categories in the prompt regarding AI/ML device evaluation are not applicable.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for the Vinyl Patient Examination Glove (Yellow) are primarily based on established industry standards and specific physical and biological properties. The reported device performance is presented in comparison to a predicate device and adherence to these standards.

    ITEMAcceptance Criteria (Proposed Device K173580)Reported Device Performance (Proposed Device K173580)Predicate Device (K163168) Performance
    Product CodeLYZLYZLYZ
    Regulation No.21 CFR 880.625021 CFR 880.625021 CFR 880.6250
    ClassIII
    Intended UseDisposable device for medical purposes, worn on examiner's hands to prevent contamination between patient and examiner.Meets intended use.Meets intended use.
    Powdered or Powder freePowder freePowder freePowder free
    Design FeatureAmbidextrousAmbidextrousAmbidextrous
    Labeling InformationSingle-use indication, powder free, device color, device name, glove size and quantity, Vinyl Examination Gloves, Non-Sterile.Meets labeling requirements.Meets labeling requirements.
    Length (min)230 mm (S, M, L, XL)S: 230mm, M: 230mm, L: 230mm, XL: 230mmS: 230mm, M: 235mm, L: 245mm, XL: 245mm (XS: 230mm)
    Width (tolerance)+/- 5 mmS: 85mm, M: 95mm, L: 105mm, XL: 115mmS: 85mm, M: 95mm, L: 105mm, XL: 115mm (XS: 80mm)
    Finger Thickness (min)0.05 mm0.05 mm0.05 mm
    Palm Thickness (min)0.08 mm0.08 mm0.08 mm
    Comply with ASTM D5250YesYes (Analysis 1 and 2 confirm compliance despite differences in specific values)Yes
    ColorantYellowYellowWhite, Blue, Yellow
    Tensile Strength Before Aging (min)12 MPa12 MPa15 MPa
    Ultimate Elongation Before Aging (min)300 %300 %380 %
    Tensile Strength After Aging (min)12 MPa12 MPa15 MPa
    Ultimate Elongation After Aging (min)300 %300 %380 %
    Freedom from HolesBe free from holes when tested in accordance with ASTM D5151 Level=G-I, AQL=1.5Meets this criterion.Be free from holes when tested in accordance with ASTM D5151 AQL=1.5
    Powder ContentMeets requirements of ASTM D52501.8 mg (Implied to meet requirement)Meets requirements of ASTM 5250
    MaterialVinylVinylVinyl
    Biocompatibility (Irritation)Not an irritant (Comply with ISO 10993-10)Not an irritantNot an irritant
    Biocompatibility (Sensitization)Not a sensitizer (Comply with ISO 10993-10)Not a sensitizerNot a sensitizer
    Biocompatibility (Cytotoxicity)Not show potential toxicity to L-929 cells (Comply with ISO 10993-5)Not show potential toxicity to L-929 cellsNot show potential toxicity to L-929 cells
    Label and LabelingMeet FDA's Requirements of 21 CFR 801Meets FDA's requirements.Meets FDA's requirements.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify general "test set" sample sizes or data provenance in the context of clinical studies (as it states "No clinical study is included in this submission"). For non-clinical tests like physical properties and freedom from holes, the standards (e.g., ISO 2859-1 for sampling) outline the sampling procedures, but specific sample sizes used in this submission are not explicitly stated. The manufacturer is "NATURE MEDICAL PRODUCTS CO., LTD." in Hebei, China, which suggests the tests were likely conducted in China. These non-clinical tests would be considered prospective for the device being evaluated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device is a Vinyl Patient Examination Glove, not an AI/ML device requiring expert medical interpretation for ground truth. The "ground truth" for these types of devices is based on established laboratory testing and adherence to recognized performance standards (e.g., ASTM, ISO standards) for physical and biological properties.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable since this document relates to a physical medical device (examination gloves) and not a diagnostic system requiring human interpretation or adjudication for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an examination glove, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an examination glove.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on:

    • Industry Standards: Compliance with established physical and chemical property requirements outlined in standards such as ASTM D5250, ASTM D5151, ASTM D6124.
    • Biocompatibility Testing: Results from in vitro and in vivo tests (ISO 10993-5, ISO 10993-10) determining irritation, sensitization, and cytotoxicity.
    • Manufacturing Specifications: Adherence to the manufacturer's own design and material specifications.

    8. The sample size for the training set

    This is not applicable. There is no "training set" in the context of an AI/ML model for this type of physical medical device.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for an AI/ML model.