    Manufacturer
    Date Cleared
    2018-02-23

    (102 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco Evolution Series Steam Sterilizer Models HC-3000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

    • Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. .
    • . Steam Flush Pressure-Pulse (SFPP) - is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

    The Amso Evolution Series Prevacum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 5-1):
    Table 5-1. Amso Evolution Series Prevacuum Steam Sterilizer factory-programmed sterilization cycles and cycle values

    CyclesSterilizeTemperatureSterilizeTimeDry TimeRecommended Load
    Prevac270°F (132°C)4 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each andFabric Packs. Refer toTable 5-3 for recommendedquantities.
    Prevac270°F (132°C)4 minutes5 minutesSingle Fabric Pack
    Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer toTable 5-3 for recommendedquantities.
    Liquid*250°F (121°C)45 minutesN/ARefer to Table 5-4 forrecommended quantities.
    Prevac275°F (135°C)3 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 5-3 for recommendedquantities.
    Gravity270°F (132°C)15 minutes45 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 5-3 for recommendedquantities.
    Gravity250°F (121°C)30 minutes45 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 5-3 for recommendedquantities.
    Gravity270°F (132°C)25 minutes15 minutesFabric Packs. Refer toTable 5-3 for recommendedquantities.
    DARTWarm-Up270°F (132°C)3 minutes1 minuteN/A
    DART270°F (132°C)3 ½ minutes1 minuteBowie-Dick Test Pack,DART Test Pack
    Leak TestN/AN/AN/AN/A
    • The liquid cycle is for non-patient contact use only.

    The Amsco Evolution Series Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 5-2):

    CyclesSterilizeTemperatureSterilizeTimeDry TimeRecommended Load
    SFPP270°F (132°C)4 minutes45 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each and FabricPacks. Refer to Table 5-3 forrecommended quantities.
    SFPP275°F (135°C)3 minutes45 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 5-3 for recommendedquantities.
    Prevac270°F (132°C)4 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each and FabricPacks. Refer to Table 5-3 forrecommended quantities.
    Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer to Table5-3 for recommendedquantities.
    SFPP270°F (132°C)4 minutes5 minutesSingle Fabric Pack
    Prevac275°F (135°C)3 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 4-3 for recommendedquantities.
    Gravity270°F (132°C)15 minutes45 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 5-3 for recommendedquantities.
    Liquid*250°F (121°C)45 minutesN/ARefer to Table 5-4 forrecommended quantities.
    DARTWarm-Up270°F (132°C)3 minutes1 minuteN/A
    DART270°F (132°C)3½ minutes1 minuteBowie-Dick Test Pack,DART Test Pack
    Leak TestN/AN/AN/AN/A
    Table 5-2. Amso Evolution Series Steam Flush Pressure-Pulse (SFPP) Sterilizer factory-programmed sterilization cycles and ycle values
    *The liquid cycle is for non-patient contact use only.
    Device Description

    The Amsco Evolution Floorloader Steam Sterilizer models HC-2000 and HC-3000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

    • . Prevacuum - is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
    • Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, ● Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
      The chamber sizes of the proposed Amsco Evolution Floorloader Steam Sterilizer Models are as follows:
    • 26" x 61" x 49" (Model HC 2000) .
    • 26'' x 61'' x 72'' (Model HC 3000)
    AI/ML Overview

    The document describes the Amsco Evolution Floorloader Steam Sterilizer (HC-2000 and HC-3000) and its substantial equivalence to a predicate device. The information provided primarily focuses on the device's technical specifications, intended use, and the verification activities conducted for specific modifications. It does not contain a typical "study" proving device performance against acceptance criteria in the sense of a clinical trial or an AI algorithm's performance study with expert ground truth.

    Instead, the "acceptance criteria" and "device performance" relate to engineering verification and validation activities for modifications made to an existing device.

    Here's an analysis based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    Device ModificationVerification / Validation ActivityAcceptance CriteriaResults of Verification / Validation
    Control replacementPerformance testingPer ST8Pass
    Electrical safetyPer IEC 61010-1Pass
    Door gasket replacementDoor operation checkProper function and no alarms after multiple cyclesPass
    Software modificationsSoftware validationSoftware shall be appropriately verified and validatedPass
    Vacuum pump replacementPerformance testingPer ST8Pass
    Printer replacementSoftware validationSoftware shall be appropriately verified and validatedPass
    Valve modificationsPerformance testingPer ST8Pass
    Switch modificationActuationProper actuation, no alarmsPass
    Door mounting bracket modificationDoor operation checkProper function after multiple cyclesPass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The verification activities described are typical engineering tests, not studies with "test sets" of data in the context of an AI/algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. The acceptance criteria primarily refer to engineering standards (e.g., ST8, IEC 61010-1) and functional tests, not expert-derived ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. As these are engineering verification tests, an adjudication method for a test set (as typically used in clinical studies) would not be applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided. This document pertains to a steam sterilizer, a medical device, but it is not an AI-powered diagnostic tool, so an MRMC study comparing human readers with and without AI assistance is irrelevant and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not provided. This device is not an algorithm or AI system, so the concept of standalone performance for an algorithm does not apply.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for the verification activities appears to be based on established engineering standards and functional requirements. For example, "Per ST8" or "Per IEC 61010-1" explicitly states the standard utilized. For functional checks like "Door operation check," the ground truth is "Proper function and no alarms after multiple cycles." This is an objective functional outcome, not based on expert consensus, pathology, or outcomes data in a medical diagnostic sense.

    8. The sample size for the training set

    This information is not provided. This device is a steam sterilizer, not an AI or machine learning system that would have a "training set."

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable to the device described.