510(k) Data Aggregation

    K Number
    K173267

    Validate with FDA (Live)

    Date Cleared
    2017-12-07

    (57 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis MasterLoc™ is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery.

    Hip replacement is indicated in the following cases:

    • · Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia;
    • · Avascular necrosis of the femoral head;
    • · Acute traumatic fracture of the femoral head or neck;
    • · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
    Device Description

    The MasterLoc™ Stem: Lateralized Plus implants are a line extension to Medacta's MasterLoc Stem product line (cleared under K151531 and K160289). The MasterLoc "" Stem: Lateralized Plus implants offer an additional caput-collum-diaphyseal (CCD) angle of 122° and neck offsets for various patient anatomies. The MasterLoc™ Stem: Lateralized Plus implants include cementless, flat, dual tapered design stems intended for total hip arthroplasty in primary or revision surgery.

    The MasterLoc™ Stem: Lateralized Plus implants are made with a titanium alloy substrate (Ti6Al7Nb) according to ISO 5832-11 Second Edition 2014-09-15: Implants for Surgery -Metallic Materials - Part 11: Wrought Titanium 6-Aluminium 7-Niobium Alloy. The surface treatment consists of Ti coating with a thickness of 700 um in the proximal 2/3 of the shaft to improve proximal fixation. The distal portion of the stems are uncoated with a satin finish obtained from glass bead blasting.

    The MasterLoc " Stem: Lateralized Plus implants are available in sizes 4 to 12, with a range of stem lengths from 126.5 mm to 147.5 mm. The stems have a Eurocone (12/14 taper with an angle of 5° 42' 30'') and the necks are polished.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "MasterLoc™ Stem: Lateralized Plus," which is a hip prosthesis. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria in the same way a new drug or novel medical device might.

    Therefore, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of diagnostic accuracy for an AI/ML device is not directly applicable to this submission. This 510(k) is for a prosthetic implant, and its "acceptance criteria" are related to mechanical performance and biocompatibility to demonstrate it functions similarly to existing, cleared devices.

    However, I can extract the information that is relevant to testing and performance data for this device, which serves a similar purpose of demonstrating safety and performance.

    Here's a breakdown of the available information based on your requested categories, tailored to what's present in this 510(k) document:


    Acceptance Criteria and Device Performance (Adapted for a Prosthetic Device):

    This submission doesn't define "acceptance criteria" in terms of diagnostic accuracy metrics (e.g., sensitivity, specificity, AUC) or a specific "reported device performance" in that sense. Instead, the acceptance criteria are inherent in the standards to which the device was tested, and the "reported device performance" is the successful completion of these tests.

    Acceptance Criteria Category (Implied by Standards)Reported Device Performance (Implied by Submission)
    Fatigue Resistance (Stem)Device successfully passed fatigue testing per ISO 7206-4.
    Fatigue Resistance (Head/Neck)Device successfully passed fatigue testing per ISO 7206-6.
    Range of Motion (ROM)Device successfully passed range of motion testing per EN ISO 21535.
    Pyrogenicity (Endotoxin Levels)Bacterial Endotoxin Test (LAL test) was conducted according to European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>). Pyrogen test conducted according to USP chapter <151>. (Implied successful, as no adverse findings reported preventing clearance).
    Material CompositionTitanium alloy substrate (Ti6Al7Nb) according to ISO 5832-11.
    Surface TreatmentTi coating with a thickness of 700 um in the proximal 2/3 of the shaft. Distal portion uncoated with satin finish (glass bead blasting).

    1. A table of acceptance criteria and the reported device performance:
    As detailed above. The "acceptance criteria" are the successful completion of tests adhering to the specified ISO and EN standards. The "reported device performance" is that these tests were conducted and the device met the requirements of these standards, allowing for a substantial equivalence determination.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not explicitly stated as a "test set" in the context of data for an algorithm. For mechanical testing, the sample size refers to the number of physical devices tested. This information is not provided in the document (number of stems tested for fatigue, ROM, pyrogenicity).
    • Data Provenance: The tests conducted are non-clinical (mechanical and pyrogenicity) and were performed to international standards (ISO, EN, European Pharmacopoeia, USP). The location of these tests is not specified, but the submitter (Medacta International SA) is based in Switzerland. These are prospective tests performed on new devices manufactured for this submission, not retrospective data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    Not applicable. This is not a study requiring expert readers or ground truth establishment in the diagnostic sense. The "ground truth" for mechanical properties is determined by physical measurements against established engineering standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. There is no human interpretation or adjudication method described for the mechanical and pyrogenicity tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is not a diagnostic device or an AI/ML algorithm. No human reader studies were conducted as part of this submission.
    The document explicitly states under "Clinical Studies": "no clinical studies were conducted."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    The "ground truth" for this device's performance is based on well-established engineering and material science standards (ISO, EN) and pharmacopoeia standards (USP, European Pharmacopoeia) for mechanical properties, material composition, and biocompatibility (pyrogenicity). These standards define acceptable limits and testing methodologies.

    8. The sample size for the training set:
    Not applicable. This is not an AI/ML device that uses a training set.

    9. How the ground truth for the training set was established:
    Not applicable. This is not an AI/ML device.