Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173264

    Validate with FDA (Live)

    Device Name
    EmbryoScope+
    Manufacturer
    Date Cleared
    2018-05-11

    (212 days)

    Product Code
    Regulation Number
    884.6120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EmbryoScope+ consists of the following devices with the following indications for use:

    The EmbryoScope+ incubator provides an environment with controlled temperature and gas concentrations (CO2 and O2) for the development of embryos at or near body temperature. Use of the EmbryoScope+ incubator is limited to five days (120 hr) covering the time from post insemination to day five of development.

    The EmbryoSlide+ culture dish is intended for preparing, storing, and transferring human embryos. The EmbryoSlide+ culture dish must be used together with the EmbryoScope+ incubator.

    The Embryo Viewer software is intended for displaying, storing, and transferring images generated by the EmbryoScope+ incubator. This software includes a user annotation for capturing information on embryo development parameters as well as a user-defined modeling function, which allows the user to combine annotated information on embryo development parameters to aid in embryo Viewer software does not control any hardware components in the EmbryoScope+ incubator.

    The ES Server software is intended to store, archive and transfer data. In addition, this software includes functions for managing models and performing calculations based on image data and embryo development parameters.

    The EmbryoScope+ incubator, EmbryoViewer software, and ES Server software must be used together to export embryo images from the EmbryoScope+ incubator. The EmbryoViewer software must be used together to analyze the embryo images.

    Device Description

    EmbryoScope+ consists of the following devices: EmbryoScope+ incubator, EmbryoSlide+ culture dish, EmbryoViewer software, and ES Server software.

    The EmbryoScope+ incubator is a bench top embryo incubator with a time-lapse imaging function. It provides temperature control, and time-lapse microscopy at multiple focal planes. This device can hold up to 15 culture dishes (EmbryoSlide+ culture dish) in the incubation chamber. The culture dishes are placed on the dish holder in the EmbryoScope+ incubator. The holder provides direct heat transfer to the EmbryoSlide+ culture dish. The built-in microscope consists of an LED illumination unit and an inverted microscope/camera unit. During image acquisition, each culture dish located on the culture dish holder is rotated to the microscopy system and individual image stacks are acquired from all individual embryos in each culture dish.

    The EmbryoSlide+ culture dish is a radiation-sterilized polystyrene culture dish containing two separate reservoirs. Each reservoir has eight culture well is used to culture one embryo. Therefore, a total of 16 embryos can be cultured on one dish. Each dish includes four special wells that are only used for rinsing and handling the embryos either before or after incubation. An adhesive barcode label printed from the EmbryoViewer software is used to mark each dish. The barcode label contains two different 2D data matrices that provide information on the patient ID, treatment ID, and insemination time). The EmbryoSlide+ culture dish has a sterility assurance level of 10° and a shelf-life of four years. This device is subject to mouse embryo assay (MEA) and endotoxin testing before lot release.

    The EmbryoViewer software is used for displaying, comparing, storing, and transferring images generated by the EmbryoScope+ incubator. The data that can be viewed using this software includes embryo images, incubation details, alarms, log files and other instrument parameters. This software also includes a user annotation for capturing information on embryo development parameters as well as a user-defined modeling function, which allows the user to combine annotated information on embryo development parameters to aid in embryo selection. The EmbryoViewer software neither controls any hardware components in the EmbryoScope+ incubator nor performs any diagnostics.

    The ES Server software allows users to update and view common data. The server acts as the central unit, which stores data and controls the data flow to and from the connected devices. The server can be connected to multiple EmbryoScope+ incubators and computers with the EmbryoViewer software installed.

    AI/ML Overview

    The provided text does not contain information about a "study that proves the device meets the acceptance criteria" in the format of a clinical trial or a specific comparative effectiveness study with human readers and AI assistance. Instead, it describes non-clinical performance testing conducted to support the substantial equivalence of the EmbryoScope+ device to its predicate devices.

    The acceptance criteria are generally implied by the design specifications and testing standards mentioned.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The text doesn't explicitly present a direct "acceptance criteria" vs. "reported performance" table for all aspects. However, it lists performance specifications for the EmbryoScope+ incubator and the results of various non-clinical tests.

    Characteristic / TestAcceptance Criteria (Implied by design specification/standard)Reported Device Performance
    EmbryoScope+ Incubator
    Culture dish capacityN/A (Comparison to predicate)15 dishes
    Temperature rangeN/A (Comparison to predicate)36-39°C
    Temperature accuracyN/A (Comparison to predicate)+/- 0.2°C
    CO₂ accuracyN/A (Comparison to predicate)+/- 0.3 %
    O₂ accuracyN/A (Comparison to predicate)+/- 0.5 %
    Recirculation rateN/A (Comparison to predicate)>100 L/h (full purification of gas volume every 6 min)
    CO₂ recovery (5% ± 0.3 %) after 30s load door open<5 min<5 min
    O₂ recovery (5% ± 0.5 %) after 30s load door open<3 min<3 min
    Illumination per imageN/A (Comparison to predicate)<0.02 seconds
    Total light exposure/embryo/dayN/A (Comparison to predicate)<40 seconds
    Total light dose during 5-day incubation (J/m²)N/A (Comparison to predicate)65.2
    Time-lapse system10 min cycle time for 11 focal planes for up to 5 days10 min cycle time for 11 focal planes for up to 5 days
    EmbryoSlide+ Culture Dish
    Endotoxin testing≤20 EU/device (per USP <85> and ANSI/AAMI ST72:2002)Met the specification of ≤20 EU/device
    Mouse embryo assay (MEA)≥80% embryos developed to blastocyst in 96 hours (for 1-cell MEA)"1-cell MEA ≥80% embryos developed to blastocyst in 96 hours."
    General Non-Clinical Testing
    Electrical safetyCompliance with IEC 60601-1:2005 + C1:2006 + C2:2007 + A1:2012Testing conducted
    Electromagnetic compatibilityCompliance with EN/IEC 60601-1-2:2007Testing conducted
    Software V&VMet FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005)Testing conducted
    Radiation sterilizationCompliance with ISO 11137-2:2013Testing conducted
    Transportation simulationMet ASTM D4169-16Testing conducted
    Package integrity (Bubble test)Per ASTM F2096-11Testing conducted
    Package integrity (Peel strength)Per ASTM F88/F88M-09Testing conducted
    Gas maintenance (levels, flow, recovery)Met design specifications (described in Section 7)Met design specifications
    Temperature controlWithin defined ranges and stable over timeMet design specifications
    Time-lapse function (image acquisition, quality, resolution)Met design specificationsMet design specifications
    Light intensityLower than predicate deviceShown to be lower than the predicate device

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes non-clinical bench testing. For the Mouse Embryo Assay (MEA), the sample size is implicitly "one-cell mouse embryos" without a specific number. The data provenance is not mentioned (e.g., country of origin). The testing seems to be experimental/prospective in nature, rather than retrospective use of human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The document describes non-clinical performance and engineering testing (electrical safety, EMC, software V&V, bench performance, sterilization, package integrity, endotoxin, mouse embryo assay). These do not involve human experts establishing ground truth for a diagnostic AI system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is non-clinical performance testing, not a clinical study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study is mentioned. The device, EmbryoScope+, includes software (EmbryoViewer) with "user-defined modeling function, which allows the user to combine annotated information on embryo development parameters to aid in embryo selection." However, the text details non-clinical testing for substantial equivalence, not a clinical study on human reader performance with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    While the EmbryoViewer software includes "user-defined modeling function" to "aid in embryo selection," the substantial equivalence argument relies on comparing the entire system (incubator, software, dish) to predicate devices through non-clinical performance metrics. There is no specific mention of a standalone algorithm performance study without human involvement, particularly relating to "embryo selection" capability. The software primarily displays, stores, transfers images, and allows for user annotation and user-defined modeling, implying human-in-the-loop use. It explicitly states, "The EmbryoViewer software neither controls any hardware components in the EmbryoScope+ incubator nor performs any diagnostics."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described:

    • EmbryoSlide+ Culture Dish (MEA): The "ground truth" for the MEA test was defined by the biological outcome: "percent of embryos developed to the expanded blastocyst stage within 96 hours." This is an objective biological endpoint.
    • Other non-clinical tests (electrical safety, EMC, metrology, etc.): Ground truth is established by adherence to engineering specifications, recognized standards (e.g., IEC, EN, ASTM, ISO, USP, ANSI/AAMI), and design specifications.

    8. The sample size for the training set

    Not applicable. The document does not describe a machine learning model that requires a training set in the conventional sense. The "user-defined modeling function" in the EmbryoViewer software suggests that users define their own criteria based on embryo development parameters, rather than the device itself being trained on a large dataset.

    9. How the ground truth for the training set was established

    Not applicable, as no machine learning training set is described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1