Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173182

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-20

    (82 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as an interbody fusion device, the ProLift® Expandable System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

    Device Description

    The ProLift® Expandable System is available in a range of sizes and footprints and can expand to the desired height (8mm to 16mm) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. The implant allows packing autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.

    All implants are provided sterile and intended for SINGLE USE ONL Y and should not be reused under any circumstances. Do not use any of the ProLift® Expandable System components with components from any other system or manufacturer. The ProLift® Expandable System components should never be reused under any circumstances.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the ProLift® Expandable System, an intervertebral body fusion device. This type of regulatory filing focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than proving efficacy or safety through clinical trials with defined acceptance criteria for a device's performance.

    Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, device performance studies, sample sizes for test/training sets, expert qualifications, adjudication methods, or ground truth establishment relevant to AI/algorithm performance.

    Instead, the "Performance Data" section (page 4) refers to:

    • Static compression, dynamic compression, static and dynamic shear testing according to ASTM F2077. This is mechanical testing to demonstrate the structural integrity and equivalence of the device to predicate devices, not performance related to clinical outcomes or diagnostic accuracy for an AI system.

    The document explicitly states: "The ProLift® Expandable System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance." This underlines that the focus is on equivalence in physical and mechanical properties and intended use, not on a performance study comparing diagnostic accuracy or clinical effectiveness against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1