510(k) Data Aggregation

    K Number
    K173074

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-04

    (187 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker ICONIX XBraid TT with Needles devices are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, wrist, elbow, and shoulder. See indications below.

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collament Reconstruction

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

    Hand/Wrist: Scaphulolunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joint for All Digits, Digital Tendon Repair

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-Foot Reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

    Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair

    The NanoTack TT Suture Anchor is intended for the fixation of soft tissue to bone in the hip, shoulder, foot/ankle, hand/ wrist, elbow, and knee in the following procedures:

    Hip: Hip capsule repair, acetabular labrum reattachment

    Shoulder: Capsular stabilization (Bankart repair, anterior shoulder instability, SLAP lesion repair, capsular shifts or capsulolabral reconstructions), acromioclavicular separation repair, deltoid repair, rotator cuff tear repair, biceps tenodesis

    Foot and Ankle: Hallux valgus repair. Medial or lateral instability repair/reconstruction, Achilles tendon repair/reconstruction, midfoot reconstruction, Metatarsal ligament/tendon repair/reconstructions, bunionectomy

    Elbow, Wrist, and Hand: Biceps tendon reattachment, ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair

    Knee: Extra-capsular repair (medial collateral ligament, lateral collateral ligament, posterior oblique ligament), patellar realignment and tendon repairs (vastus medialis obliquous advancement), Iliotibial band tenodesis

    Device Description

    The ICONIX XBraid TT with Needles devices are soft-tissue fixation devices consisting of an all suture anchor with a push-in design, preloaded on a disposable inserter. The anchors consist of a polyester sheath interwoven over non-absorbable working sutures. Fixation is achieved by the polyester sheath bunching when the anchor is deployed. The suture "tape" is composed of 100% ultra-high molecular weight polyethylene (UHMWPE). These devices include needles, which are housed in a compartment within the inserter handle. These needles are swaged to the free ends of all working sutures for use at the discretion of the surgeon. The devices are provided in a sterile state; the anchor and inserter assembly will be packaged in a single-use sterile barrier system (SBS) containing a disposable drill and drill guide to aid in bone hole creation and anchor implantation.

    There are two proposed configurations of The ICONIX XBraid TT with Needles device:

    • 1.4mm suture anchor with 1.2mm XBraid TT suture tape, and
    • . 2.3mm suture anchor with 2.0mm XBraid TT suture tape.

    The NanoTack TT Suture Anchor is a non-degradable implant device consisting of a suture anchor with attached non-degradable, flat suture tape preassembled on an insertion device. This device is provided sterile, and is intended for single use only. The anchor is manufactured from polyether ether ketone (PEEK), and the suture tape is composed of 100% ultra-high molecular weight polyethylene (UHMWPE). The anchor and inserter assembly will be packaged in a single-use sterile barrier system (SBS).

    There is one proposed configuration of the NanoTack TT Suture Anchor:

    • . 1.4mm anchor with 1.2mm XBraid TT suture tape.
    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA regarding two medical devices: the ICONIX XBraid TT with Needles and the NanoTack TT Suture Anchor. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a clinical study. Therefore, much of the requested information regarding acceptance criteria, study details, ground truth, and expert involvement is not applicable or present in this type of regulatory submission.

    However, I can extract information related to the device performance testing conducted to support substantial equivalence.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results in a standalone section. Instead, it describes various "non-clinical verification and validation testing" performed to assess the efficacy and substantial equivalence of the proposed devices to their predicate devices. The implicit acceptance criterion for these tests is that the new devices perform comparably to or better than the predicate devices, or that they meet established engineering or biological safety standards.

    Below is a summary of the types of tests conducted and the general conclusion, as specific numerical acceptance criteria and performance metrics are not detailed in this high-level summary.

    Acceptance Criterion (Implicit)Reported Device Performance (General Conclusion)
    ICONIX XBraid TT with Needles:"The results of all testing demonstrate that the proposed ICONIX XBraid TT with Needles and NanoTack TT Suture Anchor devices are substantially equivalent to the respective predicate devices."
    Inserter removal effort meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Needle removal force meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Extension during cyclic loading within acceptable rangeNo specific range or result provided, but implied to be acceptable for substantial equivalence.
    Ultimate tensile strength following cyclic loading meets specsNo specific spec or result provided, but implied to be acceptable for substantial equivalence.
    Ultimate tensile strength meets specified requirementsNo specific requirement or result provided, but implied to be acceptable for substantial equivalence.
    Insertion effort meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Suture sliding force meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Design validation successfulCompleted successfully.
    Packaging validation successfulCompleted successfully.
    Sterilization validation successfulCompleted successfully.
    Biocompatibility evaluation acceptablePerformed and determined to be acceptable. "All materials used in the construction of the proposed devices have been shown to be biologically safe."
    Pyrogen limits met"This testing demonstrated that both of the proposed devices meet the recommended pyrogen limits."
    NanoTack TT Suture Anchor:
    Deployment force meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Suture sliding force meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Ultimate tensile strength meets specified requirementsNo specific requirement or result provided, but implied to be acceptable for substantial equivalence.
    Insertion effort meets specified limitsNo specific limit or result provided, but implied to be acceptable for substantial equivalence.
    Suture snagging test successfulCompleted successfully.
    Design validation successfulCompleted successfully.
    Packaging validation successfulCompleted successfully.
    Sterilization validation successfulCompleted successfully.
    Biocompatibility evaluation acceptablePerformed and determined to be acceptable. "All materials used in the construction of the proposed devices have been shown to be biologically safe."
    Pyrogen limits met"This testing demonstrated that both of the proposed devices meet the recommended pyrogen limits."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes used for each of the non-clinical tests (e.g., how many devices were tested for tensile strength, how many were subjected to cyclic loading, etc.).
    • Data Provenance: The tests are described as "non-clinical verification and validation testing." This implies laboratory-based or benchtop testing, not data derived from patients or clinical studies. Therefore, it's not applicable to discuss country of origin or retrospective/prospective nature in the typical sense of clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The evaluations described are "non-clinical verification and validation testing," which refers to product performance and safety testing (e.g., mechanical testing, biocompatibility). These types of tests do not involve human experts establishing "ground truth" in the way a clinical study would (e.g., radiologists interpreting images). The "ground truth" for these tests would be the measured physical properties or biological safety outcomes as defined by engineering standards or biological assessments.

    4. Adjudication Method for the Test Set

    This is not applicable as there is no "test set" in the context of expert review or clinical evaluation that would require adjudication. The testing described is objective, non-clinical performance data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This is not applicable. The devices are surgical suture anchors used for soft-tissue to bone fixation. They are physical implants, not AI-powered diagnostic or assistive technologies. Therefore, an MRMC study related to AI assistance for human readers is irrelevant to these devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. As mentioned, these are physical medical devices, not algorithms or AI.

    7. The Type of Ground Truth Used

    The "ground truth" for these non-clinical tests is based on objective measurements against engineering specifications, material properties, and established biological safety standards. For example:

    • Tensile Strength: Measured force at failure, compared against a predefined standard or predicate device performance.
    • Biocompatibility: Results of biological assays (e.g., cytotoxicity, sensitization, irritation) compared against ISO 10993 standards.
    • Sterilization Validation: Confirmation of sterility assurance levels (SAL) according to recognized standards.
    • Pyrogen Testing: Measurement of pyrogen levels, compared against recommended limits.

    8. The Sample Size for the Training Set

    This is not applicable. There is no "training set" in the context of non-clinical device performance and safety testing for these physical implants. The testing performed is to validate the device's design and manufacturing, not to train an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for these devices.

    In summary, the document details non-clinical verification and validation activities demonstrating that modifications to existing suture anchors (ICONIX XBraid TT with Needles and NanoTack TT Suture Anchor) result in devices that are substantially equivalent in performance and safety to their predicate devices. It does not contain information about clinical studies, AI algorithms, or expert ground-truthing as typically found in submissions for diagnostic or AI-based devices.