510(k) Data Aggregation

    K Number
    K173062

    Validate with FDA (Live)

    Date Cleared
    2018-01-25

    (119 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) is intended for single use by operating room personnel and other general healthcare workers and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.

    Device Description

    Non Woven Face Mask is a single use multi-layer mask with outer layer and inner layer (spunbond polypropylene) that sandwich a meltblown polypropylene filter material. There are two options for the surgical mask to be secured on users via earloops or ties. Earloops are of urethane elastic fiber and not made with natural rubber latex; and ties are of spunbond polypropylene and also not made with natural rubber latex. The nose piece is a pliable white aluminum strip, covered by polypropylene covering. All of the materials used in the construction of the surgical mask are being used in currently marketed devices. Non Woven Face Mask has two models which are VQN0185W and VQN0185B. They are basically the same, the only difference is VQN0185W adopts earloops and VQN0185B adopts ties to secure the mask on user.

    AI/ML Overview

    This is a 510(k) summary for a Non Woven Face Mask, indicating substantial equivalence to a predicate device. The information provided is for regulatory clearance of a medical device and does not involve AI or algorithms, human readers, or image analysis. Therefore, many of the requested fields are not applicable.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance

    Test ParameterPredicate Device PerformanceSubject Device PerformanceAcceptance Criteria (Implied by Predicate)
    Fluid ResistanceFluid resistantPass at 120mm HgFluid resistant (Pass at 120mm Hg)
    Particle Filtration Efficiency (PFE)Average 99.54% at 0.1 micronAverage 99.74% at 0.1 μm≥ 99.54% at 0.1 micron (or comparable)
    Bacterial Filtration Efficiency (BFE)>99.9%Average 99.4%>99.9% (or comparable)
    Flammability Class11Class 1
    Delta – P (Breathability)Average 3.38 mmH2O/cm²Average 2.7 mmH2O/cm²Lower Delta-P is preferable
    BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritatingNon-cytotoxic, non-sensitizing, non-irritatingNon-cytotoxic, non-sensitizing, non-irritating

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each performance test. It only states that "The performance tests of Non Woven Face Mask were conducted." The data provenance is not specified beyond indicating it's for the "subject device" (V&Q Manufacturing Corporation, Wuhan, Hubei, CHINA). It's implied to be prospective testing for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a medical device performance testing for a face mask, not an AI or image analysis device where expert ground truth is typically established. The performance tests are based on standardized methodologies (ASTM, ISO, MIL-M, CFR).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI or image analysis study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device. The device itself (the face mask) is tested for its standalone physical performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests are the established standards and methods outlined, such as ASTM F2299 for PFE, ASTM F1862 for fluid resistance, ASTM F 2101-14 for BFE, and ISO 10993 for biocompatibility. These are objective measurements against defined criteria, not expert consensus or pathology in the context of diagnostic AI.

    8. The sample size for the training set

    Not applicable. This is not an AI or machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.