510(k) Data Aggregation

    K Number
    K172787

    Validate with FDA (Live)

    Device Name
    Access hsTnl
    Date Cleared
    2018-06-14

    (272 days)

    Product Code
    Regulation Number
    862.1215
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Access hsTnl is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnl) levels in human serum and plasma using the Access 2 Immuoasay Systems to aid in the diagnosis of myocardial infarction (MI).

    Device Description

    The Access hsTnl and the Access 2 Immunoassay System compose the Access Immunoassay System for the quantitative determination of cardiac troponin I (cTnl) in human serum and plasma.

    The Access hsTnI reagent packs contain specific reagents for the in vitro diagnostic measurement of cTnl including:

    • R1a: Dynabeads* paramagnetic particles coated with mouse . monoclonal anti-human cTnl antibody suspended in TRIS buffered saline, with surfactant, bovine serum albumin (BSA). < 0.1% sodium azide, and 0.1% ProClin** 300.
    • R1b: 0.1 N NaOH .
    • R1c: TRIS buffered saline, surfactant, protein (mouse). < 0.1% sodium . azide, and 0.1% ProClin 300.
    • R1d: Sheep monoclonal anti-human cTnl alkaline phosphatase conjugate diluted in ACES buffered saline, with surfactant, BSA matrix, protein (bovine, sheep, mouse), < 0.1% sodium azide, and 0.25% ProClin 300.
    AI/ML Overview

    The provided text describes the Beckman Coulter Access hsTnI diagnostic device, an immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma to aid in the diagnosis of myocardial infarction (MI).

    Here's an analysis of the acceptance criteria and study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides a comparison of the new device (Access hsTnI) with its predicate device (Access AccuTnI+3 Reagent) and details multiple performance characteristics. The specific acceptance criteria themselves are not explicitly stated as pass/fail thresholds in a dedicated table, but rather implied through the performance data presented and comparison to the predicate. The "Summary of Studies" section details the studies conducted to establish the device's performance, which in effect serves as demonstration of meeting acceptance criteria for analytical and clinical performance.

    Implicit Acceptance Criteria and Reported Performance from the Text:

    Feature/CriterionPredicate Device (Access AccuTnl+3) PerformanceNew Device (Access hsTnl) Performance
    Intended UseAid in diagnosis of MIAid in diagnosis of MI
    Assay PrincipleChemiluminescent sandwich immunoassaySame
    Sample TypeSerum and heparinized plasmaSame
    Analytical Measuring Range0.02 ng/mL to 100 ng/mL (20 pg/mL to 100,000 pg/mL)2.0 pg/mL to 27,027 pg/mL
    Expected Results (99th Percentile URL)0.02 ng/mL with 95% CI of 0.01-0.05 ng/mLLithium Heparin Plasma: - Overall: 17.5 pg/mL (95% CI: 12.6–20.7 pg/mL) - Females: 11.6 pg/mL (95% CI: 8.4–18.3 pg/mL) - Males: 19.8 pg/mL (95% CI: 14.0–42.9 pg/mL) Serum: - Overall: 18.2 pg/mL (95% CI: 13.1–23.1 pg/mL) - Females: 11.8 pg/mL (95% CI: 8.7–18.7 pg/mL) - Males: 19.7 pg/mL (95% CI: 14.3–44.3 pg/mL)
    PrecisionTotal CV of ≤8% at >0.075 ng/mL; SD ≤0.006 at ≤0.075 ng/mL≤ 10% within-laboratory CV for concentrations ≥ 11.5 pg/mL; ≤ 1.15 pg/mL within-laboratory SD for concentrations < 11.5 pg/mL (Reported ranged from 2.6% to 7.4% for concentrations ≥ 11.5 pg/mL and 0.34 to 0.93 for concentrations < 11.5 pg/mL)
    Open Reagent Pack StabilityStable at 2 to 10°C for 56 daysStable at 2 to 10°C for 64 days
    Clinical Performance (Sensitivity, Specificity, PPV, NPV)Not detailed for predicateLithium Heparin Plasma (Overall, at Baseline): Sensitivity 86% (86/100), Specificity 90% (510/564), PPV 61% (86/140), NPV 97% (510/524) Serum (Overall, at Baseline): Sensitivity 84% (91/108), Specificity 91% (541/595), PPV 63% (91/145), NPV 97% (541/558) (Detailed results provided for different time intervals and by sex)
    High-dose Hook EffectNot detailed for predicateNo high-dose hook effect up to at least 2,000,000 pg/mL
    LinearityNot detailed for predicateLinear across the range of the assay (2.0 to approximately 27,027 pg/mL) in both serum and lithium heparin plasma.
    Dilution RecoveryNot detailed for predicateDilution recovery demonstrated across the range of the assay (2.0 to approximately 27,027 pg/mL) with 10-fold dilution yielding 100 ± 10% recovery for samples from 27,027 to 270,270 pg/mL.
    Limit of Blank (LoB)Not detailed for predicate0.8 pg/mL
    Limit of Detection (LoD)Not detailed for predicate2.0 pg/mL in both serum and lithium heparin plasma.
    Limit of Quantitation (LoQ)Not detailed for predicate4.1 pg/mL (CV ≤ 10%), 2.0 pg/mL (CV ≤ 20%).
    Analytical SpecificityNot detailed for predicateNo significant cross-reactivity found for tested compounds (defined as concentration shift > ± 2 SD; 10% for > 11.5 pg/mL, 2.30 pg/mL for ≤ 11.5 pg/mL).
    Interfering SubstancesNot detailed for predicateNo significant interference found for tested compounds at specified concentrations (defined as concentration shift > ± 2 SD; 10% for > 11.5 pg/mL, 2.30 pg/mL for ≤ 11.5 pg/mL).

    2. Sample size used for the test set and the data provenance

    • 99th percentile URL study:

      • Sample Size: 1089 subjects (595 females, 494 males for lithium heparin plasma; 595 females, 493 males for serum).
      • Provenance: Multicenter prospective study conducted at five geographically diverse locations throughout the United States.
    • Clinical performance study:

      • Sample Size: 1,851 evaluable subjects.
      • Provenance: Multicenter prospective study conducted for ED patients presenting with chest pain or equivalent ischemic symptoms suggestive of Acute Coronary Syndromes (ACS). Patients were enrolled from 14 geographically diverse, primary care hospital-associated emergency departments in the United States, reflecting regional, urban, suburban, and rural patient populations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Ground Truth Experts: An independent panel of expert physicians.
    • Qualifications of Experts: Not explicitly stated beyond "expert physicians." However, their role in adjudicating MI statuses using criteria consistent with the Universal Definition of Myocardial Infarction suggests expertise in cardiology and MI diagnosis. They were blinded to assay results and attending physicians' diagnoses.

    4. Adjudication method for the test set

    • Adjudication Method: "True MI statuses of all subjects were adjudicated by an independent panel of expert physicians using criteria consistent with the Universal Definition of Myocardial Infarction." The specific number of experts involved in the panel (e.g., 2+1, 3+1) is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This device is a laboratory immunoassay for measuring a biomarker (cardiac troponin I), not an imaging-based AI diagnostic tool requiring "human readers" or "AI assistance." Therefore, an MRMC comparative effectiveness study, in the context described, was not performed and is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is a standalone diagnostic assay (algorithm equivalent in this context), as it directly measures cTnI levels. Its performance (sensitivity, specificity, PPV, NPV) was evaluated without direct human intervention in the result generation or interpretation of the assay itself during the study. The study focuses on the diagnostic accuracy of the assay using the established 99th percentile URLs against an expert-adjudicated ground truth. However, the note explicitly states: "The Access hsTnl assay is not intended to be used in isolation; results should be interpreted in conjunction with other diagnostic tests and clinical information," indicating that clinical decision-making ultimately involves a human interpreting the result in context.

    7. The type of ground truth used

    • Ground Truth Type: Expert consensus. Specifically, "True MI statuses of all subjects were adjudicated by an independent panel of expert physicians using criteria consistent with the Universal Definition of Myocardial Infarction."

    8. The sample size for the training set

    • The document describes performance studies and reference range establishment. It does not mention a separate "training set" in the context of machine learning. The studies described are for validation and characterization of the assay and its clinical performance. The 99th percentile URL was established using 1089 healthy subjects, and the clinical performance was evaluated on 1,851 evaluable subjects. These represent the datasets used to demonstrate key performance metrics.

    9. How the ground truth for the training set was established

    • As noted above, the concept of a "training set" in the machine learning sense is not directly applicable here. For the clinical performance evaluation (which is the closest analog to a test set in ML terms for diagnostic accuracy), the ground truth for MI status was established by "an independent panel of expert physicians using criteria consistent with the Universal Definition of Myocardial Infarction."