510(k) Data Aggregation

    K Number
    K172738

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-11-08

    (57 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended to be used in displaying digital images, including standard and multiframe digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.

    Device Description

    RadiForce RX560 is a color LCD monitor for viewing medical images including those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 2,048 x 2,560 pixels (5MP) with a pixel pitch of 0.165 mm.

    Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.

    There are two model variations, RX560 and RX560-AR . The difference of the two variations is the surface treatment of the display screens; the surface treatment of the RX560 is Anti-Glare (AG) treatment and that of the RX560-AR is Anti-Reflection (AR) coating.

    Two RX560 monitors mounted on a single stand configuration is available identified by with "MD" like RX560-MD and RX560-AR-MD.

    RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX560 based on the QC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX560.

    AI/ML Overview

    The provided text describes a 510(k) summary for the EIZO RadiForce RX560, a color LCD monitor for medical imaging, including mammography and breast tomosynthesis. The document focuses on demonstrating substantial equivalence to a predicate device (RadiForce GX550) rather than presenting a standalone study with acceptance criteria for an AI-powered diagnostic device.

    Therefore, many of the requested elements for an AI device's acceptance criteria and study are not applicable or cannot be extracted from this document, as it concerns a display monitor, not an AI algorithm.

    However, I can extract the performance testing performed and the general findings related to the display characteristics.

    Here's the breakdown of what can be inferred and what cannot, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document defines "pre-defined criteria" for the display characteristics, but the specific numerical values of these criteria are not explicitly stated. The reported performance is that the device "meet[s] the pre-defined criteria when criteria are set."

    Acceptance CriteriaReported Device Performance
    Conformance to DICOM GSDF (TG18 guideline)Meets criteria
    Angular dependency of luminance response (horizontal, vertical, diagonal)Meets criteria
    Luminance non-uniformity (TG18 guideline)Meets criteria
    Chromaticity non-uniformity (TG18 guideline)Meets criteria
    Chromaticity at center (5%, 50%, 95% max luminance)Meets criteria
    Display reflections (specular, diffuse, haze)Meets criteria
    Small-spot contrast ratioMeets criteria
    Spatial resolution (MTF)Meets criteria
    Noise (NPS)Meets criteria
    Pixel aperture ratioMeets criteria
    Absence of miscellaneous artifacts (TG18 guideline)Meets criteria
    Temporal responseMeets criteria
    Luminance stabilityMeets criteria
    Maximum allowed pixel defects/faultsMeets criteria
    Overall display characteristicsEquivalent to predicate device (RadiForce GX550)

    2. Sample size used for the test set and the data provenance

    As this is a display monitor and the tests are "bench tests" on the device itself, there isn't a "test set" of medical images in the same way an AI algorithm would have. The 'sample size' would relate to the number of monitors tested, which is not specified but implied to be sufficient for a product release. Data provenance is not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for display performance is based on established technical standards and guidelines (e.g., AAPM TG18), not expert interpretation of medical images.

    4. Adjudication method for the test set

    Not applicable. Testing is objective measurement against technical standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a display device, not an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a display device. The "bench tests" are a form of standalone performance evaluation for the monitor itself.

    7. The type of ground truth used

    The ground truth is based on established technical specifications and performance guidelines for medical displays, specifically those outlined in:

    • AAPM Task Group 18 (TG18 guideline) for display performance assessment.
    • "Guidance for Industry and FDA Staff: Display Accessories for Full-Field Digital Mammography Systems-Premarket Notification (510(k)) Submissions."

    8. The sample size for the training set

    Not applicable. This is a display device, not an AI algorithm. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable.