    K172701

    Validate with FDA (Live)

    Device Name
    BIO-C SEALER
    Date Cleared
    2018-01-04

    (119 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
      1. Sealing the root canal of permanent teeth;
      1. Internal reabsorption treatment.
    Device Description

    BIO-C SEALER is a ready-to-use injectable endodontic bioceramic sealer, suitable for obturation of root canals.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a dental device called BIO-C SEALER. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing clinical efficacy studies with specific acceptance criteria as might be found for a novel device or drug.

    Therefore, the supplied text does not contain the information requested regarding:

    • A table of acceptance criteria and reported device performance in a clinical context. The performance data is for physical-chemical properties to show equivalence to the predicate.
    • Sample size used for a test set (clinical data).
    • Data provenance for a test set (clinical data).
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for a test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone performance (algorithm only without human-in-the-loop performance).
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for clinical efficacy.
    • Sample size for a training set (clinical data).
    • How ground truth for the training set was established (clinical data).

    Instead, the document highlights non-clinical (bench) testing to support substantial equivalence.

    Here's what can be extracted from the document regarding "performance criteria" in the context of a 510(k) and the "study" (bench tests) that proves them:

    1. A table of acceptance criteria and the reported device performance

    The document states: "BIO-C SEALER has undergone extensive bench testing to provide evidence that its physical-chemical properties are substantially equivalent to iRoot SP. Bench tests performed are: flow, setting time, film thickness, solubility and radiopacity. Both devices have comparable flowability, setting time, film thickness, solubility and radiopacity."

    This implies that the acceptance criteria for BIO-C SEALER's physical-chemical properties were to demonstrate comparability or substantial equivalence to the predicate device (iRoot SP) for these specific metrics. The reported performance is that they are comparable.

    Acceptance Criteria (Implicit)Reported Device Performance (BIO-C SEALER)
    Flow: Comparable to iRoot SPAchieved comparability
    Setting Time: Comparable to iRoot SPAchieved comparability
    Film Thickness: Comparable to iRoot SPAchieved comparability
    Solubility: Comparable to iRoot SPAchieved comparability
    Radiopacity: Comparable to iRoot SPAchieved comparability
    Biocompatibility: Non-mutagenic, does not cause allergenic potential, good tolerance.Achieved (product is non-mutagenic, does not cause an allergenic potential after multiple uses and has a good tolerance)

    2. Sample sized used for the test set and the data provenance

    • Sample size: Not specified. The document only mentions "extensive bench testing." The sample size would refer to the number of samples of the material tested for each physical-chemical property.
    • Data provenance: Bench test results; conducted by Angelus Indústria de Produtos Odontológicos S/A. It's non-clinical, so country of origin for patients/data is not relevant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was bench testing of physical properties, not a clinical study requiring expert ground truth.

    4. Adjudication method for the test set

    Not applicable. This was bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a dental material, not an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a dental material, not an AI algorithm.

    7. The type of ground truth used

    For physical-chemical properties, the "ground truth" is measured by standard laboratory methods according to relevant ISO or ASTM standards, against which the test device's performance is compared to the predicate device's established performance or regulatory requirements. For biocompatibility, the ground truth would be established by validated tests for mutagenicity, allergenicity, and general tolerance.

    8. The sample size for the training set

    Not applicable. This device does not use machine learning or require a training set.

    9. How the ground truth for the training set was established

    Not applicable. This device does not use machine learning or require a training set.