Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172444

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-28

    (226 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

    Device Description

    The ROSA BRAIN device is a robotized platform providing guidance of any neurosurqical instruments compatible with the diameter of the adaptors supplied by Medtech (for example, a biopsy needle). The device is composed of a compact robotic arm and a touch screen mounted on a robot stand. Different types of instruments may be attached to the robot arm and changed according to the requirements of the procedure to be completed. The touch screen ensures the communication between the device and its user by indicating the actions to be done as well as by offering various commands. ROSA BRAIN is an image-quided device that assists the surgeon in planning the position of instruments or implants on preoperative images. It provides a stable, accurate and reproducible mechanical guidance in accordance with the planning. An image acquisition of the patient's head (MRI / CT images) is performed prior to surgery and loaded into the device. In the preoperative phase, the surgeon carries out the surgical planning on the patient images using the device software. The desired surgical parameters for positioning of the surgical instruments are defined (for example: target point, entry point and instrument lenath). During surgery, the device provides accurate and rigid guidance of the required instrument according to the previously completed planning. The optical distance sensor used with the ROSA BRAIN is susceptible to Electrostatic Discharge (ESD). The ROSA BRAIN device is intended to be used with anti-static sterile drapes which are designed for the device and listed in the list of compatible devices. The anti-static sterile drapes are mitigations to ROSA Brain device ESD immunity.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called ROSA BRAIN (v3.0.0.5), a computer-assisted surgical device (stereotaxic instrument). This document describes the device, its intended use, and the testing conducted to demonstrate its substantial equivalence to a predicate device (ROSA BRAIN v3.0.0.0).

    Based on the information provided, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document focuses on demonstrating substantial equivalence to a predicate device rather than setting new, specific acceptance criteria for this version's performance at a detailed algorithmic level. The performance data provided are primarily for safety and general functional aspects, and a key 'performance' metric mentioned is system applicable accuracy.

    Test CategoryAcceptance Criteria (Implied / Stated)Reported Device Performance
    System Applicative Accuracy (In vitro)Performance demonstrated to be substantially equivalent to the predicate device. Specific numerical targets for accuracy are given for the predicate and are implicitly the acceptance criteria for the new device.- Robot arm positioning accuracy: < 0.75 mm RMS - Device applicative accuracy: < 2mm(These results are stated to be "substantially equivalent to the predicate device")
    Electrical Safety and EMCCompliance with recognized electrical safety standards: IEC 60601-1 and electromagnetic compatibility standard IEC 60601-1-2."Evaluation and testing were performed on the subject device and demonstrated to be substantially equivalent to the predicate device." Device complies with IEC 60601-1:2005/A1:2012 and IEC 60601-1-2:2014.
    Biocompatibility TestingCompliance with FDA Guidance Document: Use of International Standard ISO 10993-1. (Specifically, the predicate device had passed Cytotoxicity, Sensitization, Irritation, and Acute systemic toxicity tests, which implied the new device should not introduce new risks)."The biocompatibility evaluation for ROSA BRAIN device has been conducted in accordance with FDA Guidance Document... The evaluation reveals that biocompatibility requirements are met by the ROSA BRAIN device." "The subject devices were evaluated against the predicate testing and determined to be substantially equivalent."
    Software Verification and ValidationCompliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices and IEC 62304 Standard (Medical Device Software - Life Cycle Process). Software considered "major" level of concern."Evaluation and testing were performed on the subject device and demonstrated substantially equivalent performance to identified predicate device." Includes code walkthroughs, unit, integration, and system test reports, and verification/validation activities.
    Cleaning and Sterilization ValidationCompliance with FDA Guidance Document Reprocessing of Reusable Medical Devices and standards such as ISO 17665-1, ISO 17664, ANSI/AAMI ST79, and AAMI TIR 12."Evaluation was performed of the subject device and demonstrated to be substantially equivalent to the identified predicate devices." Automated cleaning validation and sterilization validation (using two cycles) were performed.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states: "Clinical data were not required to support the safety and effectiveness of ROSA Brain. All validation was performed based on non-clinical performance tests."

    Therefore:

    • Test Set Sample Size: Not applicable in the context of human patient data or studies. The "test set" refers to the non-clinical bench testing conducted to verify accuracy and other engineering specifications. For physical accuracy, the entry "System applicative accuracy In vitro testing" indicates "Testing on the subject device was performed." This would imply multiple runs or repetitions, but a specific numerical sample size (e.g., number of test cases or measurements) is not provided.
    • Data Provenance: Not applicable as no human data was used. The testing was in vitro (bench testing) and non-clinical. Given the manufacturer is Medtech S.A. in Montpellier, France, the testing would presumably have been conducted in a controlled lab environment at or by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable, as no human-in-the-loop clinical or reader studies were performed, and the evaluation was based on non-clinical performance tests against engineering specifications. The "ground truth" for the accuracy test would have been established by precise measurement tools and calibration standards (e.g., coordinate measuring machines, optical tracking systems) rather than human experts in an interpretive sense.

    4. Adjudication Method for the Test Set:

    Not applicable, as no human-in-the-loop or interpretive studies requiring adjudication were performed.

    5. MRMC Comparative Effectiveness Study:

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The submission explicitly states: "Clinical data were not required to support the safety and effectiveness of ROSA Brain." The device's clearance is based on substantial equivalence to a predicate, primarily through non-clinical performance testing.

    6. Standalone (Algorithm Only) Performance:

    The device is a robotized platform providing guidance, not an AI algorithm that generates diagnostic outputs or interpretations. Its primary "performance" is in terms of the physical accuracy and reliability of its robotic positioning and guidance, as well as its software's ability to plan and execute these movements. The "System applicative accuracy" testing (Robot arm positioning accuracy < 0.75 mm RMS, Device applicative accuracy < 2mm) represents the standalone mechanical and software performance accuracy.

    7. Type of Ground Truth Used:

    For the "System applicative accuracy" test, the ground truth was metrological/engineering ground truth, established through precise physical measurements using calibrated equipment (e.g., optical tracking systems, mechanical benchmarks) in an in-vitro testing environment, and compared against predefined engineering tolerances.

    8. Sample Size for the Training Set:

    Not applicable. This device is a robotic surgical assistance system, not a machine learning model that requires a "training set" in the conventional sense of AI/ML. The software itself undergoes verification and validation, but this refers to traditional software engineering practices, not AI model training.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no "training set" for an AI/ML model for this device. The software development and testing follow standard software verification and validation procedures (IEC 62304).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1