Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172372

    Validate with FDA (Live)

    Date Cleared
    2017-09-06

    (30 days)

    Product Code
    Regulation Number
    870.3300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QuadraSphere Microspheres are indicated for embolization of hypervascularized tumors including hepatoma, and peripheral arteriovenous malformations.

    Device Description

    QuadraSphere Microspheres are sterile, biocompatible, hydrophilic, nonresorbable, expandable, acrylic copolymer microspheres. QuadraSphere Microspheres are packaged in a vial containing 25 mg of dry microspheres. The expansion rate is dependent on ionic concentration. When in contact with blood, non-ionic contrast media, or normal saline (NaCl 0.9%), QuadraSphere Microspheres expand to approximately 4 times their dry state diameter; equal to 64x volume. They are available in a range of sizes.

    AI/ML Overview

    This document describes the QuadraSphere Microspheres, a vascular embolization device. The main change in this submission is the extension of the product line to include a 20-40 µm size, in addition to existing sizes.

    Here's the information regarding the acceptance criteria and study as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a precise "acceptance criteria" table with numerical targets, but it lists performance parameters tested and states that the device met the pre-determined acceptance criteria for each.

    Parameter TestedReported Device Performance
    Macroscopic appearance of the powderMet pre-determined acceptance criteria
    Microscopic examination of the microspheresMet pre-determined acceptance criteria
    Swelling in Hexabrix of the microspheresMet pre-determined acceptance criteria
    Humidity of the microspheresMet pre-determined acceptance criteria
    Bacterial endotoxinsMet pre-determined acceptance criteria
    Catheter compatibilityMet pre-determined acceptance criteria
    Animal testingMet pre-determined acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for each specific test (e.g., number of microsphere batches for appearance, number of animals for in-vivo testing).

    • Provenance: The studies were conducted by Biosphere Medical, S.A. No specific country of origin for the data is mentioned beyond the company's location in France. The document describes "design verification and validation testing," implying this was prospective testing directly supporting the device's clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This is not a study requiring expert readers to establish ground truth for image interpretation or diagnosis. The tests are focused on physical and biological properties of the microspheres themselves.

    4. Adjudication Method for the Test Set:

    Not applicable. This is a technical performance study, not one requiring adjudication of diagnostic interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No. This document does not describe an MRMC study. The device is a physical product (microspheres), not an AI algorithm for diagnostic assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    Not applicable. This device is not an AI algorithm.

    7. The Type of Ground Truth Used:

    The ground truth for the performance tests was based on established engineering and biological specifications for medical device manufacturing and safety. For example:

    • Physical properties (appearance, microscopic features, swelling) are typically compared against pre-defined specifications for the material and design.
    • Biocompatibility is assessed against ISO 10993-1 standards.
    • Bacterial endotoxin levels are measured against established safety limits.
    • Catheter compatibility is tested against functional requirements for delivery.
    • Animal testing would have pre-defined endpoints to evaluate safety and function in a biological system.

    8. The Sample Size for the Training Set:

    Not applicable. This device is not an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. This device is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1