Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171880

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-07-13

    (385 days)

    Product Code
    Regulation Number
    882.5550
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The use of IRRAflow CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed for ≤ 24 hours.

    Device Description

    The IRRAflow® CNS System is an intracranial pressure (ICP) monitoring and drainage system. The IRRAflow CNS System consists of an IRRAflow Control Unit and two sterile disposable parts, the IRRAflow Tube Set and the IRRAflow Catheter.

    The drainage flow of cerebrospinal fluid (CSF) into the IRRAflow Catheter is uni-directional and gravity-driven; there is no recirculation of the CSF. A parallel line from the saline infusion bag is used in case clearance at the tip of the catheter is required. The IRRAflow Tube Set has a cassette that clicks on to the IRRAflow Control Unit and aligns the tubing against a peristaltic pump and pinch valve. An aspiration bag is attached to the Control Unit tape measure, defining the height of the bag relative to the catheter's tip position in the patient's head and thus controlling the speed of drainage. The tubing and catheter can be disconnected and connected by standard Luer-Lock connectors. Settings can be changed via the user interface on the Control Unit.

    The default mode provides drainage and measuring ICP, allowing single bolus injections when indicated. The bolus injections allow the catheter to be flushed when it becomes clogged. CSF or intracranial fluid samples can be taken from the aspiration port.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device Name: IRRAflow CNS System

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a direct table of acceptance criteria with numerical targets. Instead, it details various tests performed and reports the outcome as "PASS" if the acceptance criteria for that specific test were met. The acceptance criteria for each test were implied to be compliance with documented requirements or relevant standards (e.g., ISO 10993, IEC 60601-1-2:2014).

    Test CategorySpecific Test / RequirementTest Method SummaryReported Device Performance
    Biocompatibility TestingSystemic Toxicity, Mediated PyrogenDetermine if a saline extract causes a febrile response in rabbits.PASS (Non-pyrogenic)
    Acute Systemic Toxicity, InjectionScreen test article extracts for potential toxic effects from a single-dose systemic injection in mice.PASS (No abnormal clinical signs indicative of toxicity)
    Irritation/Intracutaneous ReactivityDetermine if chemicals that may leach or be extracted cause local irritation in dermal tissues of rabbits.PASS (No abnormal clinical signs)
    Indirect Hemolysis (Extract)Determine the hemolytic properties of a medical device/material.PASS (All acceptance criteria met)
    Cytotoxicity (MEM Elution)Determine the cytotoxicity of extractable substances.PASS (All acceptance criteria met)
    SensitizationEvaluate the allergenic potential or sensitizing capacity.PASS (No abnormal clinical signs)
    Bench and Electrical TestingIRRAflow CNS System VerificationDocument results of system verification and system regression verification testing.PASS (Complied with documented requirements)
    IRRAflow CNS System ValidationEnsure the device complies with established requirements through validation test procedures.PASS (Complied with documented requirements)
    IRRAflow CNS System Static AnalysisDescribe the process and results from the static analysis of the IRRAflow CNS system software.PASS (No errors encountered)
    Basic Safety testInput Test; Heating Test; Leakage Current Test; Dielectric Voltage Test.PASS (Acceptance criteria met)
    Electromagnetic Compatibility testDetermine compliance with IEC 60601-1-2:2014 Class B.PASS (Acceptance criteria met)
    IRRAflow Catheter Performance TestDescribe the tensile test results for the IRRAS catheter.PASS (Acceptance criteria met)
    IRRAflow Catheter Torsion and ShearDescribe the torsion and shear testing results for the IRRAS catheter.PASS (Acceptance criteria met)
    IRRAflow Catheter Drainage FlowDescribe the Drainage Flow testing methods for the IRRAS catheter.PASS (Acceptance criteria met)
    Shelf Life / Package IntegrityValidation of sterile barrierDescribe procedures for validating sterile barrier meets design and standards requirements.PASS (Acceptance criteria met)
    Packaging Peel TestDescribe roll packing seal testing results for the IRRAS catheter and tube set.PASS (Acceptance criteria met)
    Aging TestDescribe objective evidence that the IRRAS catheter and tube set meet appropriate requirements for transportation and aging.PASS (Acceptance criteria met)
    Sterilization TestingSterilization processValidate the effectiveness of electron beam radiation of IRRAflow Catheter and Ethylene Oxide sterilization of IRRAflow Tube Set.PASS (Acceptance criteria met)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample sizes used for the various tests described (e.g., number of animals for biocompatibility, number of units for bench testing).

    The data provenance is not explicitly stated as country of origin or retrospective/prospective. The description suggests bench testing and laboratory studies (e.g., animal tests for biocompatibility, electrical tests), rather than studies involving human clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    This information is not applicable as the studies described are primarily laboratory-based verification and validation tests, not studies requiring expert interpretation of clinical data to establish a ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reason as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes bench testing and laboratory validation to demonstrate substantial equivalence to predicate devices, not studies involving human readers or comparative effectiveness in a clinical context. Therefore, there's no effect size reported for human readers improving with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    The IRRAflow CNS System is a physical medical device (drainage and monitoring system), not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance is not applicable. The verification and validation described represent the standalone performance of the device itself.

    7. The Type of Ground Truth Used:

    The "ground truth" for the tests performed is based on:

    • Standardized Biological Responses: For biocompatibility tests, the ground truth is the expected biological reaction in animal models (e.g., absence of fever, absence of toxic effects, normal cell growth, non-hemolytic properties, no irritation/sensitization) according to established international standards (like ISO 10993).
    • Engineering Specifications and Requirements: For bench and electrical testing, the ground truth is defined by the device's design specifications, documented requirements, and adherence to international safety and performance standards (e.g., IEC 60601-1-2:2014) and the manufacturer's own internal validated requirements.
    • Physical and Mechanical Properties: For catheter performance tests (tensile, torsion, shear, drainage flow), the ground truth is the specified physical and mechanical properties and expected performance parameters.
    • Sterility Assurance Levels: For sterilization and package integrity, the ground truth is the demonstration of sterility and maintenance of a sterile barrier according to validated protocols and standards.

    8. The Sample Size for the Training Set:

    This information is not applicable. The IRRAflow CNS System is a hardware medical device with integrated software for control and monitoring, not a machine learning or AI model that requires a "training set" in the conventional sense. The "training" for the device would be its design and manufacturing processes, guided by engineering principles and standards.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1