Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171791

    Validate with FDA (Live)

    Device Name
    MAGEC® System
    Date Cleared
    2017-08-31

    (76 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGEC® System is intended for skeletally immature patients less than 10 years of age with severe progressive spinal deformities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) associated with or at risk of Thoracic Insufficiency Syndrome (TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.

    Device Description

    The MAGEC System is comprised of a sterile single use spinal rod that can be surgically implanted using appropriate NuVasive® Reline® 4.5-5.0 (Reline Small Stature) or Armada® fixation components (i.e. pedicle screws, hooks and/or connectors). The implanted MAGEC Rod is used to brace the spine during growth to minimize the progression of scoliosis. The system includes a non-sterile hand held External Remote Controller (ERC) 1 or 2 that is used periodically after implantation to non-invasively distract the implanted spinal rod. The titanium MAGEC Rod includes an actuator portion that holds a small internal magnet. The magnet in the actuator can be turned non-invasively by use of the ERC, which is electrically powered. The hand held non-invasive ERC is placed over the patient's spine and then manually activated, which causes the implanted magnet to rotate and either distract or retract the rod. Once the physician determines that the implant has achieved its intended use and is no longer required, the implant is explanted. Additional accessories for the MAGEC System include the MAGEC Manual Distractor, MAGEC Rod Template, and the MAGEC Wand Magnet Locator. The MAGEC Manual Distractor is a sterile, single use device, which is used in the operating room to test the MAGEC Rod prior to implantation. The MAGEC Manual Distractor used with the MAGEC System is made of Radel and contains a rare-earth magnet. The MAGEC Rod Template is a sterile, single-use accessory used for intraoperative planning of the surgical procedure. The MAGEC Wand Magnet Locator is a non-sterile device which is used during the distraction procedure to locate the magnet within the MAGEC Rod previously implanted in a patient.

    AI/ML Overview

    The provided documents describe the MAGEC® System, a non-fusion growing rod system for pediatric spinal deformities. The information included focuses on the FDA 510(k) submission for this device, which aims to demonstrate substantial equivalence to previously cleared predicate devices.

    Based on the provided text, here's an analysis of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device PerformanceComments by AI
    Pyrogenicity Test (LAL Kinetic Turbidimetric Assay)"Pyrogen testing was performed on the subject device to ensure it meets the pyrogen limit specifications for sterile implant devices."The documentation explicitly states that the device meets pyrogen limit specifications. This suggests a pass/fail criterion for pyrogen levels, but the specific numerical limits are not provided.
    Static Compression Bending (ASTM F1717-15)"The results demonstrate that the subject MAGEC System is substantially equivalent to the predicate."While the general conclusion is given, the specific loads, displacements, or failure modes, and how they compare numerically to the predicate device or a defined threshold, are not detailed.
    Static Torsion (ASTM F1717-15)"The results demonstrate that the subject MAGEC System is substantially equivalent to the predicate."Similar to static compression bending, specific numerical performance metrics are not provided.
    Dynamic Compression Bending (ASTM F1717-15)"The results demonstrate that the subject MAGEC System is substantially equivalent to the predicate."Again, specific numerical performance metrics and their comparison to a predicate or standard are not included in this summary.
    Mechanical Performance (overall)"Nonclinical testing was performed to demonstrate that the subject MAGEC System is substantially equivalent to the predicate device."This is a general statement. Substantial equivalence for mechanical properties implies that the new device's performance is either similar to or better than the predicate's, and does not raise new questions of safety or effectiveness. The specific threshold for "similar" is not defined here.

    Important Note: The provided document is a 510(k) summary, which is typically a high-level overview. It often summarizes the conclusions of testing rather than providing detailed raw data or specific numerical acceptance limits for each test. For a full understanding of the acceptance criteria, one would need to review the complete testing protocols and reports submitted to the FDA, which are not included here.

    Study Information

    Based on the provided text, the following information can be extracted or inferred:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not specify the sample sizes (number of devices or tests performed) for the mechanical (static and dynamic) or pyrogenicity tests. It only lists the types of tests.
      • Data Provenance: The studies were nonclinical (bench testing) performed by the manufacturer, NuVasive Specialized Orthopedics, Incorporated. The country of origin for the data is not explicitly stated but would presumably be the US where NuVasive is located. These are not clinical studies with human subjects or retrospective/prospective human data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. This was a nonclinical, bench-testing study to evaluate mechanical and biological safety (pyrogenicity) properties, not a study involving human readers or expert evaluation of images/data for ground truth establishment.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. Adjudication methods are typically relevant for clinical studies where multiple experts assess cases and their agreement or disagreement needs to be resolved. This document describes nonclinical engineering and safety testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study was not done. This device is a surgical implant system, not an AI-powered diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable. This device is a physical medical device (implant and external controller), not an algorithm or AI system.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Engineering Standards and Safety Limits. The "ground truth" for the nonclinical tests were the specifications and criteria defined by the relevant ASTM and ANSI/AAMI standards (e.g., ASTM F1717-15 for spinal implant constructs, ANSI/AAMI ST72 for bacterial endotoxins). For "substantial equivalence," the performance of the device was compared directly to that of the predicate device, implying the predicate's performance served as a benchmark for acceptable "truth."
    7. The sample size for the training set:

      • Not Applicable. This is not a machine learning or AI device, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1