    Intended Use

    The MOR™ implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit restorations. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The MOR™ implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.

    The PUR Implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit, and multiple unit retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.

    The PUR implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.

    Device Description

    The MOR 3.0mm Implant System consists of 3.0mm endosseous dental implants. The MOR 3.0mm implant is a one-piece endosseous dental implant, manufactured of Titanium 6AL-4V ELI. The portion of the implant that is submerged in the bone is a treated roughened surface (blasted and etched) to facilitate osseointegration. The MOR 3.0mm implant is similar in design to the MOR 2.4mm implant previously cleared by FDA. The implant lengths are10mm, 13mm, 15mm, and 18mm. The implant has two prosthetic head design options - 0-ball identical to Sterngold MOR 2.4mm Implant (K153173), and a Tapered Abutment similar to the IMTEC MDI MII One Piece Implant 2.9mm (K081653). This implant has the same design threads as the MOR 2.4mm Implant. The MOR Implants will be provided "Sterile" using Gamma sterilization.

    The PUR 3.2mm Implant System consists of 3.2mm threaded endosseous dental implants, as well as prosthetic components. The prosthetic components include a Straight Narrow Platform Slim Abutment with a 1.5mm and 3.0mm cuffs and a 0.050' Hex Prosthetic Screw. The purpose of the Slim Abutments is to provide surgical and prosthetic options for smaller spaces. They attach directly to the implant with the aid of the prosthetic screw and provide the transitional link between the head of the implant and the restorative components. These abutments are straight and not intended for angulation. The PUR Straight Slim Abutments are similar to FDA cleared PUR Straight Abutments Narrow and Regular Platforms 1.5mm and 3.0mm cuffs (K151928). The difference is the body of the proposed Straight Narrow Platform Slim Abutment is more tapered. The abutments and prosthetic screw are made of Titanium 6AL-4V ELI. The ancillary component is a 3.2mm cover screw, which is flush with the top surface of the implant to prevent tissue and bone from growing inside the implant. It is screwed onto the immediately after implant placement. It is made of Titanium 6AL-4V ELI. The PUR 3.2mm implant is an endosseous dental implant made of Titanium 6AL-4V ELI with a treated roughened surface (blasted and etched). The implant body is tapered with double-lead threads and10 start micro threads at the collar. The implant lengths are 8mm, 10mm, 12mm and 14 mm. This implant The PUR 3.2mm implant features the same prosthetic platform as the 3.5mm and 4.3mm implants already cleared by FDA (K151928). The purpose of the 3.2mm PUR implant is to provide surgical and prosthetic options for smaller spaces. The PUR 3.2mm Implants will be provided "Sterile" using Gamma sterilization.

    AI/ML Overview

    This looks like a 510(k) summary for dental implants, which are hardware devices and do not utilize AI. As such, the requested information (acceptance criteria, study details, sample sizes, expert involvement, MRMC studies, standalone performance, ground truth, and training data) is not applicable to this document.

    The document describes non-clinical performance testing for mechanical properties (torque) and sterilization/biocompatibility, which are standard for medical devices. The core of the submission is to demonstrate substantial equivalence to previously cleared predicate devices through direct comparison of material, design, and intended use.