510(k) Data Aggregation

    K Number
    K171447

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-05

    (203 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OxyPure Color Spherical Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lens for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Eyecare practitioners may prescribe the lens for either single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The OxyPure Color Silicone Hydrogel Soft Contact Lenses is a daily wear lens in spherical design. It may be prescribed as daily wear. It is a copolymer of 1-vinyl-2-pyrrolidinone (NVP) and Siloxane macromer. The water content is 47%. A benzotriazole UV absorbing monomer is used to block UV radiation. OxyPure Color Silicone Hydrogel Soft Contact Lenses contain a pigmented area that will alter or enhance the appearance of the natural iris. The lens is colored with one or more of the following color additives: iron oxides, titanium dioxide, phythalocyanine green and carbazole violet. Lenses are supplied sterile in sealed blister packs containing isotonic buffered saline solution. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date.

    AI/ML Overview

    This document is a 510(k) premarket notification for the OxyPure Color Silicone Hydrogel Soft Contact Lenses. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with acceptance criteria for the new device. Therefore, much of the requested information regarding acceptance criteria, study sample sizes, ground truth establishment, expert involvement, and MRMC studies is not directly applicable or available in this specific document.

    Here's a breakdown of the available information related to acceptance criteria and performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Instead of clearly defined acceptance criteria and reported device performance for a new study, this document provides a comparison table between the new device (OxyPure Color Silicone Hydrogel Soft Contact Lenses) and its predicate device (VISCO SOFT CONTACT LENS K141348). The "Result of Comparison" column essentially indicates whether the new device meets the performance characteristics of the predicate (implied acceptance) or if there's a difference.

    CategoryOxyPure Color Silicone Hydrogel Soft Contact Lenses (New device)VISCO SOFT CONTACT LENS (K141348)Result of Comparison
    Physical/Material Properties
    USAN NameOlifilcon AOlifilcon ASame
    FDA Category (Group)Group 1 Non-ionic, Low waterGroup 1 Non-ionic, Low waterSame
    Water Content47%47%Same
    Light Transmittance>94%>94%Same
    Refractive Index1.410 (hydrated)1.410 (hydrated)Same
    Oxygen Permeability (DK, 35°C)150 (Fatt method)150 (Fatt method)Same
    Leachabilityno leachable monomers and addictive residuesno leachable monomers and addictive residuesSame
    Mechanical Properties
    Tensile strength (Mpa)0.78±0.20.5±0.2Different
    Modulus (Mpa)0.60 ± 0.10.60±0.1Same
    Elongation at break (%)195 ± 20%100± 20%Different
    Toughness (J/m3)0.87±0.050.25±0.05Different
    Design/Manufacturing
    Manufacturing MethodCast MoldedCast MoldedSame
    Tint processPad printingNo Pad PrintingDifferent
    Print ZoneCorresponds to the irisN/ADifferent
    Location of printingembedded between HEMA (Sandwich Process)embedded between HEMA (Sandwich Process)Same
    Lens DesignSphericalSphericalSame
    Diameter Range13 to 15 mm13 to 15 mmSame
    Power Range- 20.00D to +20.00D in 0.25 steps- 20.00D to +20.00D in 0.25 stepsSame
    Center Thickness0.08mm @ -3.00D (Varies with Power)0.08mm @ -3.00D (Varies with Power)Same
    Base Curve8.0 mm to 9.2 mm8.0 mm to 9.2 mmSame
    Blue handling tintNoYes, reactive Blue19Different
    Color Additives in dyeIron oxide, Titanium dioxide, Phthalocyanine green, Carbazole violetReactive Blue19Different
    Packaging/Sterilization
    Single UseNo, Frequent/ Planned ReplacementNo, Frequent/ Planned ReplacementSame
    PackagingBlister PackBlister PackSame
    Packaging Solutionsterile isotonic Borate buffered salinesterile isotonic phosphate buffered salineSame
    Sterilization methodSteamSteamSame
    Shelf Life5 years5 yearsSame
    Sterility of DeviceSAL= 10-6SAL= 10-6Same

    The document states: "The differences between those two devices are mechanical properties and color additives, but other reference devices have same technical characteristic as new devices to support the substantial equivalence." This implies that while the values might be numerically different for some mechanical properties and the color additives differ, these differences are not considered to raise new questions of safety or effectiveness because they align with established characteristics of other marketed devices. The acceptance criteria essentially appear to be "comparable to the predicate device or other legally marketed devices, with no new questions of safety or effectiveness raised by differences."

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply." This indicates that there was no specific "test set" of patients or data collected for a clinical study with this device. The data provenance would be from laboratory bench testing (non-clinical tests) and comparison to the predicate device's characteristics, not a clinical trial with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical study with a test set requiring expert ground truth was conducted.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study with a test set requiring adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a contact lens, not an AI-powered diagnostic tool, and no MRMC comparative effectiveness study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a contact lens, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests, the "ground truth" refers to established scientific and engineering standards and validated test methods. These include:

    • ISO18369-3: Ophthalmic optics - Contact lenses - Part 3: Measurement Methods
    • ISO18369-4: Ophthalmic optics Contact lenses Part 4: Physicochemical properties of contact lens materials
    • ISO18369-2: Ophthalmic optics - Contact lenses - Part 2: Tolerances
    • ISO11986: Ophthalmic optics -- Contact lenses and contact lens care products -- Guidelines for etermination of preservation uptake and release
    • ISO11987: Ophthalmic optics -- Contact lenses and contact lens care products --Determination of physical compatibility of contact lens care products with contact lenses
    • ISO17665-2: Sterilization Of Health Care Products - Moist Heat - Part 2: Guidance on the Application of ISO 17665-1. (Sterility)
    • ISO17665-1: Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
    • ISO11980: Ophthalmic optics -- Contact lenses and contact lens care products -- Guidance for clinical investigations (though a clinical study was not used for this submission, compliance with this standard for potential future studies is implied).
    • ASTM D792-13: Standard Test Methods For Density And Specific Gravity (Relative Density) Of Plastics By Displacement. (Materials)
    • ISO10993-1, -5, -10, -11: For Biocompatibility.

    The "ground truth" for the device's properties is established through these standardized measurement methods, with the results compared to the known and acceptable properties of the predicate device and other legally marketed devices.

    8. The sample size for the training set

    Not applicable, as no AI/algorithm training set was used.

    9. How the ground truth for the training set was established

    Not applicable, as no AI/algorithm training set was used.