    K171194

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-24

    (30 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AS-10 / CXDI-401RF is indicated for use in generating fluoroscopic and radiographic images of human anatomy for angiography, diagnostic, and interventional procedures. The device is intended to replace spot-film devices is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.

    Device Description

    The AS-10 / CXDI-401RF is a solid state x-ray imager. It intercepts x-ray photons and the scintillator of the AS-10 / CXDI-401RF emits visible spectrum photons that illuminate an array of photo-detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the AS-10 / CXDI-401RF, an x-ray imager. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving specific performance criteria through a detailed clinical study with acceptance metrics for a diagnostic device.

    Therefore, the requested information cannot be fully extracted as it pertains to a diagnostic AI/CADe device, which this submission does not explicitly detail. This submission focuses on the safety and effectiveness of new imaging hardware (a flat panel detector) and its comparability to existing technology.

    However, I can extract information related to the device's technical specifications and the non-clinical tests performed to demonstrate its equivalence.

    Here's an attempt to answer based on the provided text, recognizing the limitations of the document's content for your specific questions:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the context of a diagnostic performance study with specific metrics (like sensitivity, specificity, or AUC) as one might find for an AI/CADe device. Instead, it compares the technical specifications and non-clinical performance of the AS-10 / CXDI-401RF to its predicate devices. The "performance" is implicitly demonstrated by showing that the new device's technical parameters are comparable to or improved upon the predicates and that it is safe and effective.

    CharacteristicPredicate Devices (CXDI-50RF K092439 & CSX-30 K162909) (Implicit "Acceptance Criteria" for Equivalence)AS-10 / CXDI-401RF (Reported Device Performance)Comparison to Predicates
    ApplicationFluoroscopy and Spot RadiologyFluoroscopy and Spot RadiologyIdentical
    TechnologyFlat panel detector: Scintillator and a-SiFlat panel detector: Scintillator and a-SiIdentical
    ScintillatorCsI(TI)CsI(TI)Identical
    Pixel Pitch160 x 160 µm160 x 160 µmIdentical
    Pixelsa) 2,208 x 2,688 (≈ 5.9 million)b) 2,496 x 1,856 (≈ 4.6 million)2,688 x 2,688 (≈ 7.2 million)Modified (Increased)
    Image Sizea) 350 x 430 mmb) 399 x 297 mm430 x 430 mmModified (Increased)
    Overall Dimensionsa) 493 x 503 x 26 mmb) 470 x 363 x 82.5 mm469 x 468 x 58 mmModified
    Weighta) 5.7 kgb) 19.0 kg13 kgModified
    Acquisition Mode (Binning)a) Up to 15 fps (1x1)b) Up to 30 fps (2x2)Up to 15 fps (1x1)Up to 30 fps (2x2)Up to 30 fps (3x3)Identical (1x1, 2x2), Modified (3x3 added)
    DQE @ 1 µGy in 0 lp/mm, RQA5a) 0.6b) 0.790.75Modified
    Spatial Resolution [MTF@2cycle/mm, RQA5]a) 0.3b) 0.220.28Modified
    A/D Conversiona) 14-bitb) 16-bit16-bitModified (from 14-bit to 16-bit for predicate a), Identical (for predicate b)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "non-clinical image comparisons involving flat panel display static and dynamic images taken with the new device and the predicate device CXDI-50RF." However, it does not specify a "sample size" in terms of number of patients or specific images, nor does it provide details on data provenance (country, retrospective/prospective). The testing focused on technical performance rather than clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided. The non-clinical tests described focus on technical image quality and system performance rather than a diagnostic evaluation requiring expert readers to establish ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as no expert human readers or adjudication process for diagnostic interpretations are described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for new imaging hardware, not an AI/CADe device designed to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "non-clinical image comparisons," the "ground truth" would likely be objective measurements of image quality (e.g., DQE, MTF, noise characteristics) comparing the new device's output to the predicate device's output, rather than a clinical ground truth established by medical experts or pathology.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.