Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170853

    Validate with FDA (Live)

    Device Name
    InnoSpire Go
    Date Cleared
    2017-11-02

    (225 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InnoSpire Go is a vibrating mesh nebulizer system designed to aerosolize liquid medications for inhalation by the patient. The device may be used with pediatric (2 years and older), defined by the prescribed medication, and adults patients in the home environment or in a hospital/clinic setting.

    Device Description

    The InnoSpire Go nebulizer is a small, handheld, internally powered nebulizer which utilizes vibrating mesh technology to generate aerosol.

    AI/ML Overview

    The provided document describes the InnoSpire Go nebulizer's acceptance criteria and the study that proves it meets those criteria, primarily through comparison to a predicate device, the Aerogen AeroNeb Go.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct pass/fail thresholds but are implied through the comparison to the predicate device's performance, aiming for "substantially equivalent" results. The study demonstrates that where differences exist, the InnoSpire Go either performs comparably or better, particularly in delivering a higher total respirable dose or having smaller particle sizes.

    Parameter (Metric)Acceptance Criteria (Implied: Substantially Equivalent to Predicate)InnoSpire Go Performance (Mean ± 95% CI)Predicate (Aeroneb Go) Performance (Mean ± 95% CI)Notes on Equivalence/Difference
    Adult Flow Rate - 30 Ipm
    Delivered dose (µg) - SalbutamolEquivalent or better2370 ± 1132206 ± 216Higher
    Delivered dose (µg) - Ipratropium bromideEquivalent or better234 ± 6229 ± 12Higher
    Delivered dose (µg) - Sodium cromoglicateEquivalent or better10146 ± 5079210 ± 500Higher
    MMAD (µm) - SalbutamolEquivalent or better (smaller is better)3.90 ± 1.044.06 ± 0.47Smaller particle size
    MMAD (µm) - Ipratropium bromideEquivalent or better (smaller is better)3.87 ± 0.93.79 ± 0.20Comparable
    MMAD (µm) - Sodium cromoglicateEquivalent or better (smaller is better)4.02 ± 0.914.15 ± 0.32Smaller particle size
    GSD - SalbutamolEquivalent or better (smaller is better)2.02 ± 0.112.00 ± 0.05Comparable
    GSD - Ipratropium bromideEquivalent or better (smaller is better)2.02 ± 0.182.06 ± 0.07Smaller particle size
    GSD - Sodium cromoglicateEquivalent or better (smaller is better)2.00 ± 0.182.02 ± 0.04Smaller particle size
    Fine particle fraction <5 µm (%) - SalbutamolEquivalent or better (higher is better)61.1 ± 14.458.9 ± 7.4Higher
    Fine particle fraction <5 µm (%) - Ipratropium bromideEquivalent or better (higher is better)61.9 ± 11.462.4 ± 2.6Comparable
    Fine particle fraction <5 µm (%) - Sodium cromoglicateEquivalent or better (higher is better)59.6 ± 11.656.9 ± 4.5Higher
    Fine particle dose <5 µm (µg) - SalbutamolEquivalent or better (higher is better)1450 ± 3891300 ± 198Higher
    Fine particle dose <5 µm (µg) - Ipratropium bromideEquivalent or better (higher is better)145 ± 30143 ± 13Higher
    Fine particle dose <5 µm (µg) - Sodium cromoglicateEquivalent or better (higher is better)6047 ± 14075240 ± 358Higher
    Coarse particle fraction >5 µm (%) - SalbutamolEquivalent or better (lower is better)38.9 ± 14.441.1 ± 7.4Lower
    Coarse particle fraction >5 µm (%) - Ipratropium bromideEquivalent or better (lower is better)38.1 ± 11.437.6 ± 2.6Comparable
    Coarse particle fraction >5 µm (%) - Sodium cromoglicateEquivalent or better (lower is better)40.4 ± 11.643.1 ± 4.5Lower
    Coarse particle dose >5 µm (µg) - SalbutamolEquivalent or better (lower is better)920 ± 315906 ± 188Comparable
    Coarse particle dose >5 µm (µg) - Ipratropium bromideEquivalent or better (lower is better)89 ± 2586 ± 4Comparable
    Coarse particle dose >5 µm (µg) - Sodium cromoglicateEquivalent or better (lower is better)4098 ± 10543970 ± 556Comparable
    Respirable fraction 1 - 5 µm (%) - SalbutamolEquivalent or better (higher is better)56.0 ± 11.953.8 ± 6.6Higher
    Respirable fraction 1 - 5 µm (%) - Ipratropium bromideEquivalent or better (higher is better)56.6 ± 956.4 ± 1.6Comparable
    Respirable fraction 1 - 5 µm (%) - Sodium cromoglicateEquivalent or better (higher is better)47.9 ± 6.443.7 ± 1.7Higher
    Respirable dose 1 - 5 µm (µg) - SalbutamolEquivalent or better (higher is better)1328 ± 3281187 ± 185Higher
    Respirable dose 1 - 5 µm (µg) - Ipratropium bromideEquivalent or better (higher is better)132 ± 24129 ± 10Higher
    Respirable dose 1 - 5 µm (µg) - Sodium cromoglicateEquivalent or better (higher is better)4859 ± 7904023 ± 83Higher
    Ultra-fine particle fraction <1 µm (%) - SalbutamolEquivalent or better5.2 ± 2.65.1 ± 0.9Comparable
    Ultra-fine particle fraction <1 µm (%) - Ipratropium bromideEquivalent or better5.4 ± 3.26.0 ± 1.0Comparable
    Ultra-fine particle fraction <1 µm (%) - Sodium cromoglicateEquivalent or better11.7 ± 5.813.2 ± 3.4Comparable
    Ultra-fine particle dose <1 µm (µg) - SalbutamolEquivalent or better123 ± 65112 ± 16Higher
    Ultra-fine particle dose <1 µm (µg) - Ipratropium bromideEquivalent or better12.6 ± 813.7 ± 3Comparable
    Ultra-fine particle dose <1 µm (µg) - Sodium cromoglicateEquivalent or better1189 ± 6431217 ± 304Comparable
    Pediatric Flow Rate - 15 Ipm
    Delivered dose (µg) - SalbutamolEquivalent or better2370 ± 1132206 ± 216Higher
    Delivered dose (µg) - Ipratropium bromideEquivalent or better234 ± 6229 ± 12Higher
    Delivered dose (µg) - Sodium cromoglicateEquivalent or better10146 ± 5079210 ± 500Higher
    MMAD (µm) - SalbutamolEquivalent or better (smaller is better)3.99 ± 0.734.85 ± 0.77Smaller particle size
    MMAD (µm) - Ipratropium bromideEquivalent or better (smaller is better)3.93 ± 0.745.08 ± 0.48Smaller particle size
    MMAD (µm) - Sodium cromoglicateEquivalent or better (smaller is better)4.27 ± 0.764.87 ± 0.54Smaller particle size
    GSD - SalbutamolEquivalent or better (smaller is better)1.82 ± 0.022.10 ± 0.09Smaller particle size
    GSD - Ipratropium bromideEquivalent or better (smaller is better)1.82 ± 0.032.09 ± 0.11Smaller particle size
    GSD - Sodium cromoglicateEquivalent or better (smaller is better)1.83 ± 0.052.00 ± 0.08Smaller particle size
    Fine particle fraction <5 µm (%) - SalbutamolEquivalent or better (higher is better)64.4 ± 12.251.1 ± 9.5Higher
    Fine particle fraction <5 µm (%) - Ipratropium bromideEquivalent or better (higher is better)65.3 ± 12.348.4 ± 5.1Higher
    Fine particle fraction <5 µm (%) - Sodium cromoglicateEquivalent or better (higher is better)59.8 ± 11.651.0 ± 6.7Higher
    Fine particle dose <5 µm (µg) - SalbutamolEquivalent or better (higher is better)1526 ± 2851129 ± 324Higher
    Fine particle dose <5 µm (µg) - Ipratropium bromideEquivalent or better (higher is better)153 ± 29111 ± 6Higher
    Fine particle dose <5 µm (µg) - Sodium cromoglicateEquivalent or better (higher is better)6070 ± 14604689 ± 375Higher
    Coarse particle fraction >5 µm (%) - SalbutamolEquivalent or better (lower is better)35.6 ± 12.248.9 ± 9.5Lower
    Coarse particle fraction >5 µm (%) - Ipratropium bromideEquivalent or better (lower is better)34.7 ± 12.351.6 ± 5.1Lower
    Coarse particle fraction >5 µm (%) - Sodium cromoglicateEquivalent or better (lower is better)40.2 ± 11.649.1 ± 6.7Lower
    Coarse particle dose >5 µm (µg) - SalbutamolEquivalent or better (lower is better)844 ± 3051076 ± 109Lower
    Coarse particle dose >5 µm (µg) - Ipratropium bromideEquivalent or better (lower is better)81 ± 29118 ± 18.1Lower
    Coarse particle dose >5 µm (µg) - Sodium cromoglicateEquivalent or better (lower is better)4076 ± 10004521 ± 852Lower
    Respirable fraction 1 - 5 µm (%) - SalbutamolEquivalent or better (higher is better)62.7 ± 11.647.2 ± 0.2Higher
    Respirable fraction 1 - 5 µm (%) - Ipratropium bromideEquivalent or better (higher is better)63.6 ± 11.864.3 ± 0.2Comparable
    Respirable fraction 1 - 5 µm (%) - Sodium cromoglicateEquivalent or better (higher is better)58.3 ± 11.259.3 ± 11.2Comparable
    Respirable dose 1 - 5 µm (µg) - SalbutamolEquivalent or better (higher is better)1485 ± 2701042 ± 100Higher
    Respirable dose 1 - 5 µm (µg) - Ipratropium bromideEquivalent or better (higher is better)149 ± 28147 ± 8.3Higher
    Respirable dose 1 - 5 µm (µg) - Sodium cromoglicateEquivalent or better (higher is better)5922 ± 14035458 ± 750Higher
    Ultra-fine particle fraction <1 µm (%) - SalbutamolEquivalent or better1.7 ± 0.61.0 ± 0.2Higher
    Ultra-fine particle fraction <1 µm (%) - Ipratropium bromideEquivalent or better1.8 ± 0.61.0 ± 0.2Higher
    Ultra-fine particle fraction <1 µm (%) - Sodium cromoglicateEquivalent or better1.5 ± 0.50.4 ± 0.4Higher
    Ultra-fine particle dose <1 µm (µg) - SalbutamolEquivalent or better41 ± 1523 ± 5Higher
    Ultra-fine particle dose <1 µm (µg) - Ipratropium bromideEquivalent or better4.1 ± 1.42.2 ± 0.4Higher
    Ultra-fine particle dose <1 µm (µg) - Sodium cromoglicateEquivalent or better148 ± 58.240 ± 36.1Higher

    The general acceptance criterion is "substantially equivalent" to the predicate device (Aerogen AeroNeb Go), meaning that any differences do not raise new questions of safety or effectiveness. The discussion sections confirm that where statistically significant differences were found, the InnoSpire Go either provided a higher Total Dose, smaller particle size (MMAD and GSD), or larger Total Respirable Dose, all of which are considered beneficial or at least not detrimental, thus supporting substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the number of samples (e.g., number of nebulizers tested for performance) for each test. However, the performance tables provide "mean and 95% confidence interval," which implies that multiple measurements were taken for each drug and parameter. The nature of these tests (aerosol performance) suggests they are conducted in a controlled lab environment.

    • Sample Size: Not explicitly stated as a single number of devices. The "mean and 95% confidence interval" for each parameter suggests statistical analysis from multiple runs/samples.
    • Data Provenance: The tests are "Bench Testing" (Non-clinical performance testing). The document does not specify country of origin for the performance test data itself, but the company is "Respironics Respiratory Drug Delivery (UK) Ltd." The tests are prospective in the sense that they are specifically conducted to compare the new device to the predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable. The study is a bench performance test comparing aerosol characteristics of device-generated output to a predicate device, not an evaluation requiring human expert interpretation or a "ground truth" derived from clinical experts (like radiologists for image analysis). The "ground truth" here is the physical measurement of aerosol particles and drug delivery.

    4. Adjudication Method for the Test Set

    This section is not applicable. There is no human adjudication process described, as the test involves objective physical measurements of aerosol properties by instruments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This document describes the 510(k) premarket notification for a nebulizer, which involves laboratory performance testing and comparison to a predicate device, not a clinical trial involving human readers or cases. Therefore, no effect size of human readers improving with or without AI assistance is reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is also not applicable. The device is not an AI/algorithm-driven system. It is a physical medical device (nebulizer). The "standalone" performance here refers to the device's technical specifications and aerosol output in a laboratory setting, which is precisely what the performance tables (Tables 3 & 4) describe.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests (Tables 3 & 4) is objective physical and chemical measurements of aerosol characteristics and drug concentration. This includes:

    • Delivered dose of drug (µg)
    • Mass median aerodynamic diameter (MMAD in µm)
    • Geometric standard deviation (GSD)
    • Particle size fractions (e.g., fine particle fraction <5 µm, coarse particle fraction >5 µm, respirable fraction 1-5 µm, ultra-fine particle fraction <1 µm)
    • Corresponding doses for these particle fractions.

    These are quantitative measurements, not subjective evaluations or clinical outcomes.

    8. The Sample Size for the Training Set

    This is not applicable. This is a medical device submission based on predicate device comparison and bench testing, not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set mentioned or required for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1