510(k) Data Aggregation

    K Number
    K170541

    Validate with FDA (Live)

    Date Cleared
    2017-07-07

    (134 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Anorganic Bone Mineral in Delivery Applicator is intended for use in dental surgery. The products may be used in surgical procedures such as:

    • Augmentation or reconstructive treatment of alveolar ridge
    • Filling of infrabony periodontal defects
    • Filling of defects after root resection, apicoectomy, and cystectomy
    • Filling of extraction sockets to enhance preservation of the alveolar ridge
    • Elevation of maxillary sinus floor
    • Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
    Device Description

    Anorganic Bone Mineral in Delivery Applicator are porous bone mineral matrices consisting of calcium phosphate derived from bovine bone without the addition of bovine type I collagen, preloaded into a delivery applicator for ease of placement in the defect site. The anorganic bone mineral component is produced by removal of the organic components from bovine bone. Anorganic Bone Mineral in Delivery Applicator is sterilized by gamma irradiation. The products are non-pyrogenic and for single use only.

    AI/ML Overview

    The document describes a 510(k) premarket notification for a medical device called "Anorganic Bone Mineral in Delivery Applicator". It does not describe a study involving an algorithm or AI.

    Therefore, many of the requested categories (sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth type for test set, training set sample size, ground truth for training set) are not applicable to this document.

    However, I can provide the acceptance criteria and demonstrated performance based on the non-clinical testing described.

    1. Table of acceptance criteria and the reported device performance

    FeatureAcceptance Criteria / StandardReported Device Performance
    Material/SourceBovine Bone, Anorganic Bone MineralBovine Bone (anorganic bone mineral)
    Composition100% Anorganic Bone Mineral Granules100% Anorganic Bone Mineral Granules
    Particle Size0.25 - 1.0 mm0.25 - 1.0 mm
    BiocompatibilityBiocompatible (based on ISO standards)Biocompatible (demonstrated by testing)
    SterilityGamma irradiation, SAL 10^-6Gamma irradiation, SAL 10^-6 (validated based on ISO 11137-1)
    PyrogenicityNon-pyrogenicNon-pyrogenic
    Single UseSingle Use OnlySingle Use Only
    Plunger ForceNot explicitly defined as "acceptance criteria", but a performance measure for the delivery applicator8.18 ± 3.8 N (measured for the subject device)

    Additional Non-Clinical Performance Testing:

    • Biocompatibility of Sterile Finished Device: Cytotoxicity, irritation, and sensitization tests were performed and demonstrated biocompatibility. (Specific acceptance criteria for these individual tests are not detailed, but the conclusion states they were met.)
    • Biocompatibility of Delivery Applicator: Cytotoxicity, physicochemical attributes of a polymeric material, and USP <661> physicochemical tests were performed. (Specific acceptance criteria not detailed, but the conclusion states they were met.)
    • Biocompatibility of Anorganic Bone Mineral: Cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, pyrogenicity, implantation, and subacute/subchronic toxicity tests were performed. (Previously submitted and deemed applicable, conclusion states they were met.)
    • Sterilization Validation: Performed in accordance with ISO 11137-1. (Implies meeting the SAL 10^-6 requirement.)
    • Bench Testing of Delivery Applicator Functionality and Customer Assessment: Performed. (Implies proper functioning of the applicator.)
    • Animal Performance Testing:
      • Anorganic Bone Mineral (derived from porcine bone tissue): Tested in an intraoral defect in a canine model.
      • Anorganic Bone Mineral with and without Collagen: Tested in a rabbit femoral condyle defect model.
        (These studies supported the performance of the bone mineral component, which is part of the subject device.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as this document describes non-clinical performance testing for a medical device (bone graft material and applicator), not an AI/algorithm study. The relevant "tests" involved laboratory-based biocompatibility, sterilization validation, and bench testing, as well as animal studies. The sample sizes for these specific non-clinical tests (e.g., number of animals, number of test samples for material characterization) are not detailed in this summary.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this document describes non-clinical performance testing for a medical device (bone graft material and applicator), not an AI/algorithm study requiring expert-established ground truth.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as this document describes non-clinical performance testing not requiring a clinical "ground truth" as typically defined for diagnostic AI. The "ground truth" for these tests would be the established scientific standards and methods for assessing material properties, biocompatibility, and sterility.


    8. The sample size for the training set

    Not applicable.


    9. How the ground truth for the training set was established

    Not applicable.