Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170250

    Validate with FDA (Live)

    Device Name
    SmartTarget
    Manufacturer
    Date Cleared
    2017-06-07

    (131 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartTarget device is intended as an accessory for image-guided interventional and diagnostic procedures involving the prostate gland, and to be used by physicians for enhanced visualization of two-dimensional (2D) transrectal ultrasound TRUS images of the prostate in clinic and hospital settings. It allows the user to segment medical images and it performs three-dimensional (3D) reconstruction of digitized TRUS video images to form a 3D TRUS volume. The SmartTarget software provides 2D and 3D image visualization features, including the ability to review images, generate multi-planar views, annotate images, and identify and record the locations of instruments inserted during the procedure.

    The device is intended to be used in diagnostic and treatment procedures in a clinical setting in which a needle or other instrument is inserted into the prostate through the perineum or urethra, or instruments that are positiond externally to the prostate so that treatment can be delivered to prostate insue, or diagnostic information obtained from prostate tissue.

    Example procedures include, but are not limited to: needle biopsy in which tissue samples are removed from the prostate; in situ diagnostic techniques, such as those based on optical sensing; thermal tissue ablation using radiofrequency, microwave, laser, cryotherapy, or high-intensity focused ultrasound; photodynamic therapy; irreversible electroporation; radioactive source implantation (brachy-injected drug therapies

    Device Description

    The SmartTarget software is designed for the fusion/registration of transrectal ultrasound (TRUS) images of the prostate with three-dimensional (3D) images, such as a magnetic resonance image (MRI) or x-ray computed tomography (CT) image etcetera, during procedures for which TRUS is used to provide real-time imaging of the prostate, nearby anatomical structures, and instruments inserted into the prostate to guide instrument placement. The software supports the following tasks: computer-assisted surgical planning, allowing target regions to be defined within the prostate in a diagnostic/planning image and TRUS images; reconstruction of 3D TRUS images from multiple US video frames, captured at pre-set intervals as the TRUS probe is translated or rotated; and image fusion/registration wherein the spatial relationship between prostate in the diagnostic/planning image and TRUS images is calculated. The system software reproduces and supplements the visual information provided by real-time TRUS images, and superimposes a graphical representation of one or more target regions on to the reproduced TRUS images. The target region(s) may represent a tumor, another tissue structure visible in the diagnostic/planning image, or a location of clinical relevance (such as a region from which tissue samples are to be removed).Displayed target regions provide additional information to the TRUS image, enabling the operating clinician to direct instruments to these regions.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the SmartTarget device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Measured Performance)Reported Device Performance (SmartTarget)
    Image alignment error (linear distance between image-visible target after registering ultrasound and MRI volumes)2.0 mm
    Overall needle-tip placement error2.9 mm

    2. Sample Size and Data Provenance for Test Set

    The document does not explicitly state a specific "test set" in the context of clinical images or patient data. The performance claims are based on "careful laboratory experiments" using "Test phantoms incorporating synthetic prostates."

    • Sample Size: Not specified for the phantoms, but implied to be sufficient for "careful laboratory experiments."
    • Data Provenance: Synthetic prostates within test phantoms. No specific country of origin or retrospective/prospective nature is applicable as it's a bench test.

    3. Number of Experts and Qualifications for Ground Truth of Test Set

    Not applicable. The ground truth for the bench testing was established through precise measurements within the controlled environment of laboratory experiments using phantoms. No human experts were involved in establishing the ground truth for these phantom tests.

    4. Adjudication Method for Test Set

    Not applicable. As the ground truth was established through physical measurements on phantoms, there was no need for expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document describes non-clinical/bench testing. It does not mention any MRMC study comparing human readers with and without AI assistance. The SmartTarget device is an accessory for visualization and guidance, rather than an AI diagnostic tool that directly impacts human reader performance.

    6. Standalone (Algorithm Only) Performance

    Yes, in essence. The reported performance metrics (image alignment error and needle-tip placement error) are derived from the device's algorithmic and mechanical accuracy during bench testing. This represents the standalone performance of the system in a controlled environment, demonstrating its precision in rendering and guiding.

    7. Type of Ground Truth Used

    Physical measurements within phantoms/laboratory experiments. The document states: "Test phantoms incorporating synthetic prostates were used to verify the accuracy of both the image registration/fusion functionality and physical instrument placement error during guided needle insertion."

    8. Sample Size for Training Set

    Not specified. The document primarily focuses on the device's functionality, comparison to predicates, and non-clinical performance testing. It does not provide details about a training set for machine learning models, as the device's core functionality appears to be image processing, registration, and visualization rather than a machine learning-based diagnostic algorithm requiring extensive training data.

    9. How Ground Truth for Training Set Was Established

    Not applicable. As no training set information is provided, the method for establishing its ground truth is also not mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1