Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170126

    Validate with FDA (Live)

    Date Cleared
    2017-03-09

    (55 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a pedicle screw fixation system, the NuVasive® 4.5-5.0 System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment 3. Fracture 4. Dislocation 5. Scoliosis 6. Kyphosis 7. Spinal tumor and/or 8. Failed previous fusion (pseudoarthrosis) The NuVasive Reline 4.5-5.0 System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion. When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline 4.5-5.0 System is also intended for the following indications: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2. Spinal stenosis 3. Spondylolisthesis 4. Spinal deformities 5. Fracture 6. Pseudoarthosis 7. Tumor resection and/or 8. Failed previous fusion When used for posterior non-cervical screw fixation in pediatric patients, NuVasive Reline 4.5-5.0 System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline 4.5-5.0 System is intended to treat pediatic patients diagnosed with the following conditions: spondylolysis, and fracture caused by turnor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft. In order to achieve additional levels of fixation, the NuVasive Reline 4.5-5.0 System rods may be connected to the Reline System.

    Device Description

    The NuVasive Reline 4.5-5.0 System is a pedicle screw system that consists of a varietv screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, iliac connectors and associated general instruments. Implant components are available in a variety sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient.

    AI/ML Overview

    This document is a 510(k) premarket notification for the NuVasive Reline 4.5-5.0 System, a pedicle screw system. It states that the device is substantially equivalent to legally marketed predicate devices. The information provided focuses on demonstrating this equivalence through non-clinical performance testing rather than clinical studies or AI-driven performance.

    Here's an analysis of the provided text based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What was measured)Reported Device Performance (Result)
    Static Compression Bending (per ASTM F1717)Results demonstrate that the subject device is substantially equivalent to the predicate.
    Static Torsion (per ASTM F1717)Results demonstrate that the subject device is substantially equivalent to the predicate.
    Dynamic Compression Bending (per ASTM F1717)Results demonstrate that the subject device is substantially equivalent to the predicate.
    Static Tulip Pull-off (per ASTM F1798)Results demonstrate that the subject device is substantially equivalent to the predicate.
    Design, Labeling/Intended Use, Material Composition, FunctionThe subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison.

    Important Note: The document confirms that "demonstrate that the subject NuVasive Reline 4.5-5.0 System is substantially equivalent to the predicate device." However, it does not provide specific numerical values for the acceptance criteria (e.g., minimum bending strength in Nm) or the numerical performance results of the NuVasive Reline 4.5-5.0 System or its predicate for direct comparison. It only states that the results show substantial equivalence.

    2. Sample Size Used for the Test Set and the Data Provenance

    The study described is non-clinical performance testing of a physical medical device (pedicle screw system), not an AI/software device involving data sets, patients, or human readers. Therefore:

    • Sample Size for Test Set: This information is not provided as the tests are for mechanical properties of the device components. The ASTM standards typically specify the number of samples required for each test.
    • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable, as this involves mechanical testing of physical implants, not patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. Ground truth, in the context of expert consensus, is relevant for AI/diagnostic devices interpreting medical images or patient data. This is a physical device subject to mechanical engineering standards.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable. Adjudication methods are used to resolve disagreements among human experts in clinical studies, not for non-clinical mechanical testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This document describes the testing of a physical medical implant (pedicle screw system), not an AI device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This document focuses on the mechanical properties and substantial equivalence of a physical medical device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this type of device is defined by the ASTM standards (ASTM F1717 and ASTM F1798) themselves, which specify the test methodologies and, implicitly, the expected performance characteristics for pedicle screw systems. Substantial equivalence is determined by comparing the new device's performance to that of a legally marketed predicate device under these standardized test conditions.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of mechanical testing for substantial equivalence of a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there's no training set, there's no ground truth for it to be established.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1