510(k) Data Aggregation

    K Number
    K163656

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-14

    (112 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of:

    • Open and Closed Fractures
    • Post-traumatic joint contracture which has resulted in loss of range of motion
    • Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
    • Pseudoarthrosis or non-union of long bones
    • Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction
    • Correction of bony or soft tissue deformity
    • Correction of segmental bony or soft tissue defects
    • Joint arthrodesis
    • Management of comminuted intra-articular fractures of the distal radius
    • Bone transport

    The Hoffmann LRF System is indicated in adults for:

    • Osteotomy
    • Revision procedure where other treatments or devices have been unsuccessful
    • Bone reconstruction procedures
    • Fusions and replantations of the foot
    • Charcot foot reconstruction
    • Lisfranc dislocations
    Device Description

    The Hoffmann LRF System is a modular, ring-based, external fixation system designed to address certain orthopedic conditions of the limbs. Through a series of pins and wires, the bone is connected to this system with the rings statically placed, or gradually manipulated, depending on the type of correction needed. The modular design allows the system to be customized according to the needs of the patient and utilizes stainless steel, aluminum, PEEK, and carbon fiber.

    AI/ML Overview

    The provided text describes a medical device submission (K163656) for the Hoffmann LRF System. It focuses on regulatory approval and comparison to predicate devices, rather than a study proving the device meets specific acceptance criteria in the context of an AI/algorithm-driven device.

    Therefore, many of the requested elements (acceptance criteria, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, training set details) are not applicable or cannot be extracted from this document, as it pertains to a mechanical external fixation system.

    However, I can extract information related to the performance data presented within the document:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical performance comparable to predicate devicesStatic Compression: Passed (Demonstrated substantial equivalence to predicate systems)
    Dynamic Compression: Passed (Demonstrated substantial equivalence to predicate systems)
    Device performs as intendedPerformed as intended, and at least as well as predicates.

    Note: The document implicitly defines "acceptance criteria" through the comparison to predicate devices, where the subject device must demonstrate equivalent or superior mechanical performance. Specific numerical acceptance values are not provided.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for the test set: Not applicable / Not provided. The testing described is non-clinical mechanical testing, not a test set of data points as in an AI study.
    • Data provenance: Not applicable. The data comes from "Comparative mechanical testing," implying laboratory tests, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This was mechanical testing, not a study requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was mechanical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/algorithm-driven device, and no MRMC study was mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a mechanical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for mechanical testing is established engineering standards and comparative performance against predicate devices.

    8. The sample size for the training set

    Not applicable. There is no training set for a mechanical device.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set mentioned.