Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163372

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-07

    (127 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RX Takeru PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion. This product (balloon models 2.0-5.0 mm) is also indicated for the post-delivery expansion of balloon expandable stents.

    Device Description

    Rx Takeru PTCA Balloon Dilatation Catheter (Rx Takeru) is a rapid exchange type of balloon dilation catheter, which consists of a distal tube, guidewire transition tube, balloon, radiopaque marker(s), mid tube, proximal tube, core wire, hub, and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the hub at the proximal end. Rx Takeru has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the guidewire transition tube to the guidewire port for the exit of a guidewire.

    The maximum compatible diameter of a guidewire used together with Rx Takeru in a PTCA procedure is 0.014 inches. Additionally, guiding catheters with a diameter of 5 or 6 Fr have been deemed to be compatible with Rx Takeru. The nominal inflated balloon diameters range from 1.5 mm to 5.0 mm with balloon working lengths of 6 mm to 30 mm. The catheter working length is 1450 mm.

    AI/ML Overview

    This document describes the performance testing and acceptance criteria for the RX Takeru PTCA Balloon Dilatation Catheter.

    1. A table of acceptance criteria and the reported device performance

    The document states that the RX Takeru "met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." Specific numerical acceptance criteria and reported performance values are not provided in the given text.

    The performance tests conducted include:

    Test TypeAcceptance Criteria Not Explicitly Stated but "Met"Reported Performance Not Explicitly Stated but "Adequate"
    Dimensional VerificationMetAdequate for intended use
    Balloon PreparationMetAdequate for intended use
    Deployment and RetractionMetAdequate for intended use
    Balloon Rated Burst PressureMetAdequate for intended use
    Balloon Fatigue (Repeat Inflations)MetAdequate for intended use
    Balloon ComplianceMetAdequate for intended use
    Balloon Inflation and Deflation TimeMetAdequate for intended use
    Catheter Bond StrengthMetAdequate for intended use
    Flexibility and Kink TestMetAdequate for intended use
    Torque StrengthMetAdequate for intended use
    RadiopacityMetAdequate for intended use
    Coating IntegrityMetAdequate for intended use
    Particulate EvaluationMetAdequate for intended use
    Balloon Rated Burst Pressure (in Stent)MetAdequate for intended use
    Balloon Fatigue (Repeat Inflations; in Stent)MetAdequate for intended use
    Transportation TestingMetAdequate for intended use
    Shelf Life TestingMetAdequate for intended use
    Biocompatibility Tests:
    CytotoxicityMetBiocompatible for intended use
    SensitizationMetBiocompatible for intended use
    Intracutaneous ReactivityMetBiocompatible for intended use
    Acute Systemic ToxicityMetBiocompatible for intended use
    PyrogenMetBiocompatible for intended use
    Hemocompatibility (in vivo thromboresistance, hemolysis, complement)MetBiocompatible for intended use
    Genotoxicity (mouse lymphoma, bacterial reverse mutation, in vivo cytogenetics assay)MetBiocompatible for intended use

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for each of the performance or biocompatibility tests. All the tests mentioned are "in vitro" (bench) tests and biocompatibility tests conducted according to international standards (ISO 10993-1). There is no mention of human subject data, therefore, data provenance in terms of country of origin or retrospective/prospective nature is not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The tests performed are laboratory-based physical and chemical tests, and biocompatibility studies. "Ground truth" in the context of expert review for diagnostic devices is not relevant here. The "truth" is established by the results of the specified test methodologies and their adherence to predetermined acceptance criteria.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. As these are not studies involving human interpretation or subjective assessments, adjudication methods are not relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, AI assistance, or human reader improvement. This device is a PTCA balloon dilatation catheter, not a diagnostic imaging device with AI components.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical medical device (catheter), not an algorithm or software. Therefore, standalone algorithm performance is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests relies on adherence to established engineering and material science specifications, validated test methods (e.g., those recommended by FDA guidance documents like "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters"), and international standards (e.g., ISO 10993-1 for biocompatibility). The "truth" is whether the device physically performs according to these criteria.

    8. The sample size for the training set

    This information is not applicable. This is a physical medical device, not a machine learning model. There is no concept of a "training set" for its development or evaluation as described in this document.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1