Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163359

    Validate with FDA (Live)

    Date Cleared
    2017-08-18

    (261 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of methotrexate in human serum or plasma on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of methotrexate to help ensure appropriate therapy.

    Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate Assay.

    Device Description

    The ARK Methotrexate Assay is a homogeneous immunoassay based on competition between drug in the specimen and Methotrexate labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.

    The ARK Methotrexate Assay consists of reagents R1 anti-Methotrexate polyclonal antibody with substrate and R2 Methotrexate labeled with bacterial G6PDH enzyme. The ARK Methotrexate Calibrator consists of a six-level set to calibrate the assay, and the ARK Methotrexate Control consists of a six-level set used for quality control of the assay (tri-level calibration range set and tri-level high range set). The ARK Methotrexate Dilution Buffer is equivalent to zero calibrator (Calibrator A).

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study information:

    Acceptance Criteria and Device Performance for ARK™ Methotrexate Assay

    The provided document describes the acceptance criteria and the study that proves the ARK™ Methotrexate Assay, when used on the Beckman Coulter AU680 analyzer, meets these criteria. The study aims to demonstrate substantial equivalence to the assay's performance on the predicate device, the Roche/Hitachi 917 analyzer.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states: "The pre-determined Pass/Fail acceptance criteria were met for all of the above performance studies." However, the specific numerical acceptance criteria or performance values for each characteristic are not provided in the given text. It only lists the types of performance characteristics evaluated.

    Performance CharacteristicAcceptance Criteria (Not Detailed)Reported Device Performance
    Limit of BlankUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Limit of DetectionUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Limit of QuantitationUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    RecoveryUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    LinearityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Accuracy (Method Comparison)Undisclosed Pass/Fail CriteriaMet Acceptance Criteria
    PrecisionUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    SpecificityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Cross-reactivityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Carry-overUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Dilution RecoveryUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    On-board StabilityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the test set for any of the performance studies.

    The document refers to "human serum or plasma" as the sample type. The country of origin of the data is not explicitly stated. The study is a "validation and verification" activity, implying it was conducted specifically for this submission, likely making it a prospective study for the purpose of device validation.

    3. Number of Experts and Qualifications for Ground Truth

    The concept of "experts" and their qualifications for establishing ground truth, as typically seen in image analysis or diagnostic interpretation, is not applicable to this type of in vitro diagnostic device (immunoassay) study. The ground truth for this device's performance characteristics is established through analytical methods and reference standards, not expert interpretation.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (e.g., 2+1, 3+1) is not applicable to this type of in vitro diagnostic device (immunoassay) study. Performance is assessed through quantitative measurements against established analytical criteria, not through expert consensus on diagnostic outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this type of in vitro diagnostic device (immunoassay). This study design is typically used for imaging or diagnostic interpretation tasks where human readers' performance is being evaluated, often with and without AI assistance. This device is an automated assay, not an AI for human interpretive tasks.

    6. Standalone (Algorithm Only) Performance Study

    The study described is a standalone performance study for the algorithm (assay system) on the Beckman Coulter AU680 analyzer. The validation and verification activities listed (Limit of Blank, Limit of Detection, Accuracy, Precision, etc.) directly evaluate the performance of the device itself, without human interpretation in the loop. The purpose is to demonstrate that the assay system alone performs comparably to its predicate.

    7. Type of Ground Truth Used

    For an immunoassay like the ARK™ Methotrexate Assay, the "ground truth" for evaluating its performance characteristics (analytical accuracy, precision, linearity, etc.) would be established through:

    • Reference materials/standards: Calibrators and controls with precisely known concentrations of methotrexate.
    • Reference methods: Comparison with established, validated analytical methods (e.g., LC-MS/MS, or the predicate device's performance which was previously validated) for samples with unknown concentrations.
    • Spiking studies: Adding known amounts of analyte to a sample and measuring recovery to assess accuracy.

    The document implicitly refers to these as part of the "analytical performance studies" and "validation and verification activities."

    8. Sample Size for the Training Set

    The document does not mention a training set or its sample size. This is expected as the ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay kit, not a machine learning or AI model that requires a "training set" in the conventional sense. The "training" for such a system would involve validating its components and establishing its operational parameters on the target analyzer.

    9. How Ground Truth for the Training Set Was Established

    Since there is no "training set" in the context of machine learning, this question is not applicable. The assay is a chemical and biological system where parameters are set based on chemical principles and reagent characteristics, and then validated through performance studies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1