    K Number
    K163355

    Validate with FDA (Live)

    Date Cleared
    2017-04-12

    (133 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping.

    Device Description

    The device under review is a family of Depth Electrodes. These electrodes provide sub-surface brain contact to support recording, monitoring and stimulation from user supplied equipment.

    AI/ML Overview

    This document describes the Ad-Tech Medical Instrument Corporation's Depth Electrodes (K163355) and their substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria in the context of an AI/ML medical device.

    Therefore, most of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth for an AI/ML device is not applicable to this 510(k) summary for a traditional medical device.

    However, I can extract the acceptance criteria (defined as "Summary of Requirement" in the document) and the reported performance ("Result") for the physical device testing.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criterion (Summary of Requirement)Reported Device Performance (Result)
    Measure electrical resistance from the electrode contact to connector.Pass
    Dimensional requirements are met by manufacturing.Pass
    Stimulation at 30 $\mu$ C/cm2 does not affect the electrodes.Pass
    The optional accessory Stay Flange supports adequate Depth Electrode retention.Pass

    The following information is not applicable to this document as it pertains to an AI/ML software device, and this 510(k) is for a set of physical medical electrodes:

    1. Sample size used for the test set and the data provenance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
    3. Adjudication method for the test set.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc).
    7. The sample size for the training set.
    8. How the ground truth for the training set was established.

    Additional relevant information from the document:

    • Intended Use: The Ad-Tech Depth Electrodes are intended for temporary (< 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports the definition of epileptogenic foci and brain mapping.
    • Predicate Devices: The device was deemed substantially equivalent to K053358, K964644, K990788, K041604, K802151, and K151790 (all previous Depth Electrodes from Ad-Tech Medical Instrument Corporation or PMT Corporation).
    • Study Type: The submission focused on demonstrating substantial equivalence based on technical characteristics and performance testing of the physical device, rather than clinical trials or AI performance evaluations.