Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163301

    Validate with FDA (Live)

    Date Cleared
    2016-12-20

    (28 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

    With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    Device Description

    The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples, and connecting components, as well as implant components from other Medtronic spinal systems which can be rigidly locked into a variety of configurations with each construct being tailor-made for the individual case.

    A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    The purpose of this submission is to modify Medtronic's CD HORIZON® Spinal System to add additional components to the system, specifically non-sterile 30° and 45° Transverse Hooks.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Medtronic CD HORIZON® Spinal System, specifically for the addition of new 30° and 45° Transverse Hooks. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than rigorously proving the device meets acceptance criteria through extensive clinical studies as one might find for novel device approvals.

    Therefore, the information you've requested regarding acceptance criteria, study details, sample sizes, ground truth establishment, and multi-reader multi-case studies is largely not applicable nor present in this document, as the submission method (510(k)) and the nature of the device (an modification to an existing spinal system component) do not typically require such detailed clinical performance data.

    However, I can extract the information that is present concerning the demonstration of substantial equivalence which serves the purpose of regulatory acceptance for this type of device.

    Here's the closest interpretation of your requests based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission, "acceptance criteria" primarily refers to demonstrating equivalent technological characteristics and performance compared to existing predicate devices, rather than predefined clinical performance metrics. The "reported device performance" is framed in terms of this equivalence.

    Acceptance Criteria (for Substantial Equivalence)Reported Device Performance (vs. Predicate)
    Intended UseThe subject hooks (30° and 45° Transverse Hooks) have the same intended use as the predicate devices. (Page 5, Section V; Page 7, Section VI)
    Indications for UseThe subject hooks have the same indications for use as the predicate devices: posterior, non-cervical fixation as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudarthrosis, and/or failed previous fusion. (Page 5, Section V; Page 7, Section VI)
    MaterialThe subject transverse hooks are manufactured from the same material (titanium alloy - Ti-6Al-4V ELI) as the predicates. This material has a long history of safe and effective use. (Page 7, Section VII)
    Fundamental Scientific TechnologyThe subject and predicate hooks are based on the same fundamental scientific technology of providing posterior, non-cervical fixation, connecting to 5.5mm/6.0mm rods in the T1-S1 region, and being secured with set screws. The only difference is a transition offset that "does not raise any issues of safety and effectiveness." (Page 7, Section VI)
    Sterilization MethodThe subject hooks use the same sterilization method as the predicate devices. (Page 5, Section V; Page 7, Section VI)
    Mechanical Performance (Safety/Effectiveness)Medtronic evaluated the subject transverse hooks through an engineering rationale and a risk analysis. Both evaluations demonstrated that the subject hooks do not introduce a new worst case to the CD HORIZON® Spinal System. Therefore, no new mechanical testing was required because existing tests on the predicate devices are considered sufficient to demonstrate equivalence for the modified components. (Page 7, Section VII) (No specific performance values are given, only a qualitative assessment of non-inferiority/equivalence through engineering rationale.)
    BiocompatibilityDue to the use of the same implant material (Ti-6Al-4V ELI) that has a long history of safe and effective use in similar spinal implants, no additional biocompatibility testing was required. (Page 7, Section VII)

    2. Sample size used for the test set and the data provenance

    • Sample size for the test set: Not applicable/Not provided in the context of this 510(k) submission. No clinical "test set" in the traditional sense was used. The evaluation was primarily based on engineering rationale, risk analysis, and comparison to existing mechanical testing data from predicate devices.
    • Data provenance: Not applicable/Not provided. The data is internal engineering and risk analysis documentation from Medtronic Sofamor Danek USA, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No "ground truth" was established by experts in the context of clinical data for this type of submission. The safety and effectiveness are established through demonstrating substantial equivalence to predicate devices, supported by internal engineering/risk assessments.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical "test set" requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical spinal implant component, not an AI-powered diagnostic or assistive tool. MRMC studies are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a mechanical spinal implant component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" equivalent in a 510(k) for a device component like this is the established safety and effectiveness of the predicate devices and the physical/mechanical properties of the materials and design, confirmed through engineering rationale and risk analysis. The regulatory acceptance depends on demonstrating that the new components are equally safe and effective as the predicates.

    8. The sample size for the training set

    Not applicable/Not provided. There was no "training set" as this is not a machine learning or AI device.

    9. How the ground truth for the training set was established

    Not applicable. There was no "training set" or "ground truth" in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1