    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Libra Pedicle Screw System is intended for immobilization of the posterior thoracic, lumbar and sacral/iliac spine (T1-S1/Ilium) in skeletally mature patients as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

    Device Description

    The Libra Pedicle Screw System consists of longitudinal members (rods), anchors (screws) and fasteners in a variety of sizes to accommodate differing anatomic requirements.

    AI/ML Overview

    This document is a 510(k) premarket notification for the "Libra Pedicle Screw System," specifically for adding a cobalt chrome rod series. It does not contain information about a study proving the device meets acceptance criteria in the way a clinical trial or algorithm performance study would. The acceptance criteria and "study" are related to demonstrating substantial equivalence to a predicate device through non-clinical performance data (e.g., theoretical analysis).

    Here's the breakdown of the information requested, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document only discusses "performance data" in the context of demonstrating substantial equivalence, not specific quantitative acceptance criteria with reported numerical performance values like sensitivity/specificity for a diagnostic device. The acceptance criterion is essentially "does not create a new worst case" or "maintains existing performance."

    Acceptance CriteriaReported Device Performance
    Mechanical Performance: The cobalt chrome rod, when integrated into the Libra Pedicle Screw System, should not create a new worst-case scenario compared to the previously cleared titanium alloy rods. This implies that the mechanical properties and system integrity should be maintained or improved, and not compromised.Theoretical analysis of the proposed device modification was performed. The results demonstrate the cobalt chrome rod does not create a new worst case.
    Material Biocompatibility: The new material (cobalt chrome) should be medically acceptable and suitable for implantation.Rods are additionally manufactured from cobalt chrome (ASTM F1537). (This implies compliance with a recognized standard, which serves as an acceptance criterion for material safety/biocompatibility).
    Technological Characteristics: The overall system should retain the same intended use, basic design, dimensions, and fundamental scientific technology as the predicate device.The Libra Pedicle Screw System possesses the same technological characteristics as the predicate device, including intended use, basic design (rod and screw configuration), material (medical grade), and sizes. The fundamental scientific technology is the same as the previously cleared device.

    2. Sample size used for the test set and the data provenance

    Not applicable in the context of a clinical study. The "test set" here refers to the theoretical analysis performed on the design of the cobalt chrome rod and its integration into the system. It's a theoretical analysis, not a study involving human subjects or a large dataset. No data provenance in the sense of patient data is mentioned, as this is a materials/mechanical evaluation.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. "Ground truth" in this context would refer to established engineering principles, material science standards (e.g., ASTM F1537), and possibly expert engineering analysis to determine if the new component "does not create a new worst case." The document does not specify the number or qualifications of experts involved in this theoretical analysis, but it would typically be conducted by qualified engineers and material scientists within Spinal Balance, Inc. or their consultants.


    4. Adjudication method for the test set

    Not applicable. There's no mention of an adjudication method as would be used for image review or clinical assessments. The "adjudication" is inherent in the internal engineering review and FDA's assessment of the 510(k) submission.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a pedicle screw system for spinal fusion, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI-related metrics are irrelevant.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical implant, not an algorithm.


    7. The type of ground truth used

    The "ground truth" used for this type of submission would be established engineering principles, material science standards (e.g., ASTM F136 for titanium, ASTM F1537 for cobalt chrome), and biomechanical testing standards (though direct testing results are not detailed here, theoretical analysis relies on such principles). The "ground truth" for showing substantial equivalence is that the new material change does not negatively impact the safety and effectiveness profile of the already cleared device, as assessed through engineering analysis.


    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable. See point 8.