510(k) Data Aggregation

    K Number
    K162544

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-04

    (204 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PINNACLE Driver is intended for driving screws and for drilling into bone, in conjunction with dental, orthognathic, and mandibular surgical procedures.

    Device Description

    The PINNACLE Driver is a handheld, cordless, software driven, battery operated driver for high speed insertion of screws and drilling of pilot holes. The device operates in both the forward and reverse directions. A sterile, single use battery powers the device and is intended for use solely with the PINNACLE Driver. Accessories for the PINNACLE Driver include hex shank drills and driver stems. The optional driver stems are reusable while the drills are single use. The PINNACLE Driver may be sterilized individually wrapped in two layers of 1-ply polypropylene wrap or unwrapped in a sterilization organizational tray inside a rigid sterilization container, which is available from OsteoMed.

    The PINNACLE Driver is intended for use in a healthcare facility/hospital for use by a clinician. The PINNACLE is a prescription device.

    The PINNACLE Driver and lithium battery is made of medical grade stainless steel, aluminum, silicone, PEEK, gold plated brass, and medical plastic. The driver stems and drills are made of medical grade stainless steel. The sterilization organizational trays and rigid sterilization containers are made of aluminum.

    The PINNACLE Driver is a battery operated driver/drill that operates in the both in the forward and reverse directions. The PINNACLE Driver contains torque limiting software to prevent over-torquing and stripping of screws in drive mode.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "PINNACLE Driver." It describes the device, its intended use, comparison to predicate devices, and performance data to support substantial equivalence.

    Based on the information provided, this document does not describe a study that proves the device meets specific acceptance criteria in the context of an AI/algorithm-driven medical device for diagnostic or prognostic purposes, nor does it refer to a multi-reader multi-case (MRMC) comparative effectiveness study.

    The "PINNACLE Driver" is a surgical tool (battery-operated driver/drill for bone, primarily for dental, orthognathic, and mandibular procedures). The "software" mentioned in the document is for torque limiting ("senses torque to prevent over-torquing and stripping of screws") and "bit recognition," not for diagnostic interpretation or image analysis.

    Therefore, many of the requested elements (e.g., ground truth, sample size for test/training sets, number of experts, adjudication methods, MRMC studies, standalone performance of an algorithm) are not applicable to this type of device and its regulatory submission.

    However, I can extract the closest applicable information regarding "acceptance criteria" and "performance study" for this mechanical/software-controlled surgical device:


    Acceptance Criteria and Device Performance (for a Mechanical/Software-Controlled Surgical Device)

    Since this is a mechanical surgical device with integrated software for operational control (torque limiting), the "acceptance criteria" and "performance" are primarily addressed through engineering and safety standards, and bench testing, rather than diagnostic accuracy metrics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implied/Stated in Document)Reported Device Performance (Summary from Document)
    BiocompatibilityCompliance with FDA Blue Book Memorandum #G95-1 for tissue/bone/dentin contact (<24 hrs).Evaluation conducted on exposed material as worst-case (implant device proximity). Tests included Cytotoxicity, Sensitization, Irritation, Systemic toxicity. (Implied compliance).
    Electrical SafetyCompliance with IEC 60601-1 standards.The system complies with IEC 60601-1.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60086-4 and ES60601-1 standards.The system complies with IEC 60086-4 and ES60601-1.
    Software FunctionalitySoftware validation for a "moderate" level of concern, demonstrating proper function of torque sensing and bit recognition.Software verification and validation testing were conducted, as recommended by FDA's guidance for software in medical devices. (Implied successful validation).
    Mechanical PerformanceAbility to drive screws and drill into bone as intended; proper functioning of forward/reverse modes, speed ranges, and torque limiting."Simulated Use" bench testing was conducted. (Implied successful performance in simulated use).
    Sterilization & ReprocessingValidation of sterilization methods (e.g., pre-vacuum steam) and reprocessing for reusable components.Sterilization Validation and Reprocessing Validation were conducted. (Implied successful validation).
    Shipping PerformanceDevice integrity maintained during transport.Shipping Validation was conducted. (Implied successful validation).
    Battery PerformanceAbsence of leak, proper power delivery.Battery Leak Testing was conducted. (Implied successful testing).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a "sample size" in the context of diagnostic data. Performance was evaluated through bench testing of the device itself (e.g., simulated use, electrical safety tests, sterilization validation). The number of test units or iterations for these engineering tests is not specified in this summary.
    • Data Provenance: Not applicable as it's not a clinical data study for algorithm development. The "data" refers to the results of engineering tests conducted by the manufacturer, OsteoMed.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable. This pertains to establishing ground truth for diagnostic data. For this device, "ground truth" would be established by engineering specifications, validated test methods, and compliance with recognized standards (e.g., IEC 60601). The "experts" would be the engineers, quality control personnel, and regulatory affairs specialists involved in the device design, manufacturing, and testing process.

    4. Adjudication Method for the Test Set

    • Not applicable. This applies to human review of diagnostic cases. For this device, compliance is determined by successful completion of specified engineering tests and validation protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. This type of study is specifically for evaluating the impact of AI assistance on human readers in diagnostic tasks. The PINNACLE Driver is a surgical tool, not a diagnostic imaging device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable in the context of an AI algorithm for diagnostic purposes. The "software" in this device controls mechanical function (torque limiting, bit recognition). Its "standalone performance" is validated through the engineering bench tests (e.g., demonstrating it can limit torque as designed).

    7. The Type of Ground Truth Used

    • Engineering Specifications and Recognized Standards: The "ground truth" for this device's performance is adherence to its design specifications, established engineering tolerances, and compliance with relevant international and national safety and performance standards (e.g., IEC 60601-1 for electrical safety, ISO standards for biocompatibility). For the software component, the ground truth is its ability to correctly execute its programmed functions (e.g., accurately sense torque, correctly identify bits) as per its design requirements and software validation protocols.

    8. The Sample Size for the Training Set

    • Not applicable. This device's software is not based on machine learning that requires a "training set" of data in the typical sense (e.g., medical images). It's software for operational control.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As above, there's no "training set" in the context of AI/ML. The "ground truth" for its simple, rule-based software is defined by its functional specifications and verified through testing.