510(k) Data Aggregation

    K Number
    K162468

    Validate with FDA (Live)

    Device Name
    ClearCheck
    Manufacturer
    Date Cleared
    2016-12-01

    (90 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearCheck is intended for quality assessment of radiotherapy treatment plans.

    Device Description

    The ClearCheck device (model RADCC) is a software intended to present treatment plans obtained from Eclipse Treatment Planning System (also referred to as Eclipse TPS) of Varian Medical Systems in a user friendly way (numerical form of data) for user approval of the treatment plan. ClearCheck runs as a dynamic link library (dll) plugin to Varian Eclipse. It is designed to run on the Windows Operating System and generated reports can be viewed on Internet Explorer. ClearCheck performs calculations on the plan obtained from Eclipse TPS (Version 12 (K131891) and Version 13.5 (K141283)) which is a software used by trained medical professionals to design and simulate radiation therapy treatments for malignant or benign diseases. ClearCheck has two components: A standalone Windows Operating System executable application that is used for 1. administrative operations to set specified default settings and user settings. 2. A plan evaluation application that is a dynamic link library (dll) file that is a plugin to the Varian Medical Systems Eclipse TPS. The plugin is designed to evaluate the quality of an Eclipse treatment plan. Plan quality is based on user specified Dose Constraints and Plan Check Parameters.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ClearCheck device, organized according to your request.

    Please note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It explicitly states that "no clinical trials were performed for ClearCheck" (Section 5.7). Therefore, a substantial portion of your requested information (e.g., MRMC studies, specific performance metrics against ground truth, expert qualifications, adjudication methods, sample sizes for test/training sets with ground truth derivation methods) is not present in this type of regulatory submission. The verification tests mentioned are likely internal software validation rather than clinical performance studies.


    1. Table of acceptance criteria and the reported device performance

    The document does not provide a specific table of quantitative acceptance criteria for device performance based on clinical outcomes or accuracy metrics. Instead, "pass/fail criteria were used to verify requirements" during internal verification tests. These requirements are implicit in the comparison to the predicate device and the claim of substantial equivalence.

    Acceptance Criteria CategoryReported Device Performance / Assessment
    Intended UseClearCheck is intended for quality assessment of radiotherapy treatment plans, equivalent to the predicate device.
    Pure Software DeviceYes, equivalent to the predicate device.
    Intended UsersMedical physicists, medical dosimetrists, and radiation oncologists, equivalent to the predicate device.
    OTC/RxPrescription use (Rx), equivalent to the predicate device.
    Operating SystemRuns on Windows 7, 8, 10, Server 2008, 2008 RS, 2012. Supported an additional OS (Windows 10) compared to the predicate, which does not raise new safety/effectiveness questions.
    CPU2.4+ GHz and Multi-core processors (2+ cores, 4+ threads), equivalent to the predicate device.
    Hard Drive SpaceRequires ~3.5MB for software (vs. 20MB for predicate), suggests 100GB for patient data (vs. 900GB for predicate). Difference acknowledged and deemed not to raise new safety/effectiveness questions because ClearCheck stores constraint templates, not large DICOM datasets like the predicate.
    Display Resolution & Color Depth1280 x 1024, 24- or 32-bit color depth (vs. 1920 x 1080 for predicate). Difference acknowledged and deemed not to raise new safety/effectiveness questions as it supports smaller monitors without affecting image quality.
    Software FunctionalityPerforms calculations on plans from Eclipse TPS based on user-specified Dose Constraints and Plan Check Parameters. Verification tests performed to ensure the software works as intended and passed requirements.
    Safety and EffectivenessDeemed as safe and effective as the predicate device through Verification and Validation testing and Hazard Analysis.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of clinical or performance data. It mentions "Verification tests" were performed for the software. These tests would involve internal generated data or existing clinical plans to validate the software's functionality, but no details on sample size, data provenance, or specific test cases are provided for external review.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. Since no clinical trials or external performance evaluations of this nature were conducted (as stated in Section 5.7), the concept of "ground truth" as derived by experts for a test set is not applicable to the submitted performance data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided for the same reasons as above.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The document explicitly states "no clinical trials were performed for ClearCheck." The device is a "quality assessment" tool for radiotherapy plans, not an AI for image interpretation or diagnosis that would typically involve human reader performance studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance assessment in the sense of the algorithm's internal calculations and functionality was performed as part of "Verification tests." However, specific quantitative metrics common for standalone AI algorithms (e.g., sensitivity, specificity, AUC against a clinical ground truth) are not provided in this regulatory summary. The device's "performance" is primarily assessed by its functional correctness and consistency with the predicate device's overall purpose of quality assessment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This is not explicitly stated. For the internal "Verification tests," the "ground truth" would likely be the expected outputs or calculated values based on established physics principles and treatment planning guidelines, which the software is designed to implement and report. This is not the same as clinical "ground truth" derived from patient outcomes or expert consensus on a diagnosis.

    8. The sample size for the training set

    This information is not applicable and not provided. ClearCheck is described as a software tool that performs calculations and presents data based on user-specified dose constraints and plan check parameters from an existing Eclipse TPS plan. It is not an AI/ML algorithm that learns from a "training set" of data to make predictions or classifications.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as the device does not employ a machine learning model that requires a training set with associated ground truth.