    Intended Use

    The CareView 1500P detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This product is not intended for mammography applications.

    Device Description

    CareView 1500P is a kind of portable wireless digital X-ray flat panel detectors which have 434mm×355mm imaging area. The device communicates by not only the wireless communication but also wired communication feature (Giga-bit Ethernet communication mode by backup network port) optionally.

    The device intercepts X-ray photons and then the scintillator emits visible spectrum photons that illuminate an array of photo detectors (a-Si) that create electrical signals. After the electrical signals are generated, it is converted to a digital value and an image will be displayed on the monitor.

    The detector should be integrated with an operating PC and an X-ray generator to utilize as digitalizing X-ray images and transfer for radiography diagnostic.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the CareView 1500P X-ray Flat Panel Detectors. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove meeting acceptance criteria in the context of an AI/algorithm-based device.

    Therefore, many of the requested categories (e.g., sample sizes for test/training sets, expert qualifications, MRMC studies, standalone performance, ground truth types) are not applicable or cannot be extracted from this document, as it pertains to an X-ray detector itself, not an AI algorithm.

    However, I can extract information related to the device's technical specifications and the comparison made for substantial equivalence.

    Here's a summary based on the provided text, indicating when information is not applicable (N/A) or not provided (NP):

    1. Table of Acceptance Criteria and Reported Device Performance

    The document directly compares the proposed device (CareView 1500P) with its predicate device (CareView 1500Cw) to demonstrate substantial equivalence, rather than listing specific "acceptance criteria" in the sense of performance thresholds for an AI/algorithm. The performance specifications are presented as direct comparisons.

    ItemAcceptance Criteria (Predicate: CareView 1500Cw)Reported Device Performance (Proposed: CareView 1500P)
    X-ray AbsorberCsl ScintillatorCsl Scintillator
    Installation TypeWireless, PortableWireless, Portable
    Readout MechanismThin Film TransistorThin Film Transistor
    Image Matrix Size2304 x 2816 pixels2304 x 2816 pixels
    Pixel Pitch154μm154μm
    Effective Imaging Area355 mm × 434 mm355 mm × 434 mm
    Grayscale16 bit, 65536 grayscale16 bit, 65536 grayscale
    Spatial ResolutionMin. 3.3 line pair/mmMin. 3.3 line pair/mm
    Rated Power SupplyDC +24 V, Max.1.5 A; Powered by power box or battery packDC +24 V, Max. 4 A; Wired Powered by switching power supply, Wireless Powered by battery pack
    Power ConsumptionMax. 36 WMax. 96 W
    CommunicationsGigabit Ethernet; IEEE 802.11a/b/g/n (2.4 / 5 GHz)Wired Gigabit Ethernet; Wireless IEEE 802.11a/b/g/n (2.4 / 5 GHz)
    Imaging PlateCarbon Fiber PlateCarbon Fiber Plate
    CoolingAir coolingAir cooling
    Dimensions384 mm x 460 mm x 15 mm470.4 mm × 510.4 mm × 18.2 mm
    Operation Temperature+5 ~ +35°C+5 ~ +35°C
    Operation Humidity30 ~ 75% (Non-Condensing)30 ~ 75% (Non-Condensing)
    Operation Atmospheric Pressure70 ~ 106 kPa70 ~ 106 kPa
    Storage/Transport Temperature-20 ~ +55°C-20 ~ +55°C
    Storage/Transport Humidity10 ~ 90% (Non-Condensing)10 ~ 90% (Non-Condensing)
    Storage/Transport Atmospheric Pressure70 ~ 106 kPa70 ~ 106 kPa
    MTF (@ 1lp/mm)~70%~70%
    MTF (@ 2lp/mm)~40%~40%
    MTF (@ 3lp/mm)~22%~22%
    DQE (@RQA5, 30µGy, 0lp/mm)~65%~65%
    DQE (@RQA5, 30µGy, 3lp/mm)~20%~20%
    Dynamic range~82 dB~80 dB
    Mechanical Structure384mm x 460mm x 15mm; built-in foldable handle; carbon fiber plate front cover; multifunctional I/O port470.4mm x 18.2mm; integral handle; protective film on carbon fiber plate front cover; power input port and backup network port

    Study Proving Device Meets Acceptance Criteria:

    The document states that Electrical safety and EMC testing according to IEC/ES 60601-1 and IEC/EN 60601-1-2 was performed, and "All test results are satisfactory."

    For performance, the document relies on a comparison to the predicate device, CareView 1500Cw (K150929). It argues that the proposed device, CareView 1500P, maintains similar technological characteristics and performance (e.g., MTF, DQE, imaging specifications) to the cleared predicate, and that the differences in mechanical structure and power consumption do not raise new questions of safety and effectiveness.

    2. Sample size used for the test set and the data provenance: Not applicable (N/A) – this is a hardware device clearance, not an AI algorithm. No clinical test set data is described in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A - no test set involving expert ground truth is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A - no test set described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A - this is a hardware device, not an AI product.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: N/A - this is a hardware device, not an AI product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A - no ground truth in the context of an AI algorithm is mentioned. The ground truth for the device's technical specifications would be engineering measurements and adherence to standards.

    8. The sample size for the training set: N/A - no AI training set is described.

    9. How the ground truth for the training set was established: N/A - no AI training set is described.